Resistive Magnet
 
A type of magnet that utilizes the principles of electromagnetism to generate the magnetic field. Typically large current values and significant cooling of the magnet coils is required. The resistive magnet does not require cryogens, but needs a constant power supply to maintain a homogenous magnetic field, and can be quite expensive to maintain.
Resistive magnets fall into two general categories - iron-core and air-core.
Iron-core electromagnets provide the advantages of a vertically oriented magnetic field, and a limited fringe field with little, if any, missile effects due to the closed iron-flux return path.
Air-core electromagnets exhibit horizontally oriented fields, which have large fringe fields (unless magnetically shielded) and are prone to missile effects. Resistive magnets are typically limited to maximum field strengths of approximately 0.6T.
spacer
 
• Share the entry 'Resistive Magnet':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Open MRI
    • Magnetism
    • Radio Frequency Pulse
    • Field Strength
    • Magnetic Field
MRI Resources 
Quality Advice - Shielding - Distributors - Brain MRI - Bioinformatics - MRA
 
DeviceForum -
related threadsInfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.
 
Magnetic resonance imaging (MRI) is based on the magnetic resonance phenomenon, and is used for medical diagnostic imaging since ca. 1977 (see also MRI History).
The first developed MRI devices were constructed as long narrow tunnels. In the meantime the magnets became shorter and wider. In addition to this short bore magnet design, open MRI machines were created. MRI machines with open design have commonly either horizontal or vertical opposite installed magnets and obtain more space and air around the patient during the MRI test.
The basic hardware components of all MRI systems are the magnet, producing a stable and very intense magnetic field, the gradient coils, creating a variable field and radio frequency (RF) coils which are used to transmit energy and to encode spatial positioning. A computer controls the MRI scanning operation and processes the information.
The range of used field strengths for medical imaging is from 0.15 to 3 T. The open MRI magnets have usually field strength in the range 0.2 Tesla to 0.35 Tesla. The higher field MRI devices are commonly solenoid with short bore superconducting magnets, which provide homogeneous fields of high stability.
There are this different types of magnets:
Resistive Magnet
Permanent Magnet
Superconducting Magnet
The majority of superconductive magnets are based on niobium-titanium (NbTi) alloys, which are very reliable and require extremely uniform fields and extreme stability over time, but require a liquid helium cryogenic system to keep the conductors at approximately 4.2 Kelvin (-268.8 Celsius). To maintain this temperature the magnet is enclosed and cooled by a cryogen containing liquid helium (sometimes also nitrogen).
The gradient coils are required to produce a linear variation in field along one direction, and to have high efficiency, low inductance and low resistance, in order to minimize the current requirements and heat deposition. A Maxwell coil usually produces linear variation in field along the z-axis; in the other two axes it is best done using a saddle coil, such as the Golay coil.
The radio frequency coils used to excite the nuclei fall into two main categories; surface coils and volume coils. The essential element for spatial encoding, the gradient coil sub-system of the MRI scanner is responsible for the encoding of specialized contrast such as flow information, diffusion information, and modulation of magnetization for spatial tagging.
An analog to digital converter turns the nuclear magnetic resonance signal to a digital signal. The digital signal is then sent to an image processor for Fourier transformation and the image of the MRI scan is displayed on a monitor.

For Ultrasound Imaging (USI) see Ultrasound Machine at US-TIP.com.

See also the related poll results: 'In 2010 your scanner will probably work with a field strength of' and 'Most outages of your scanning system are caused by failure of'
Radiology-tip.comGamma Camera,  Linear Accelerator
spacer
Radiology-tip.comUltrasound Machine,  Real-Time Scanner
spacer

• View the DATABASE results for 'Device' (141).Open this link in a new window


• View the NEWS results for 'Device' (29).Open this link in a new window.
 
Further Reading:
  News & More:
Kyoto University and Canon reduce cost of MRI scanner to one tenth
Monday, 11 January 2016   by www.electronicsweekly.com    
A transportable MRI machine to speed up the diagnosis and treatment of stroke patients
Wednesday, 22 April 2015   by medicalxpress.com    
Portable 'battlefield MRI' comes out of the lab
Thursday, 30 April 2015   by physicsworld.com    
Chemists develop MRI technique for peeking inside battery-like devices
Friday, 1 August 2014   by www.eurekalert.org    
New devices doubles down to detect and map brain signals
Monday, 23 July 2012   by scienceblog.com    
MRI Resources 
Claustrophobia - MRCP - Pacemaker - Pediatric and Fetal MRI - Safety pool - Abdominal Imaging
 
Electromagnet
 
A type of magnet that utilizes coils of wire, typically wound on an iron core, so that as current flows through the coil it becomes magnetized. The direction of the magnetic field is parallel to the long axis of the coil. Whole body electromagnets, used in medical imaging (also called resistive) are limited to their field strength, because the weight becomes prohibitively large at high field MRI. The magnetic field shuts down, if the current is switched of. Because this type of magnet generates heat, a good cooling system is essential.
For a stronger magnetic field, the wires must be manufactured of superconducting materials to reduce the power needed to produce the field.
See also Resistive Magnet, Superconducting Magnet and Upright™ MRI

spacer

• View the DATABASE results for 'Electromagnet' (24).Open this link in a new window


• View the NEWS results for 'Electromagnet' (8).Open this link in a new window.
 
Further Reading:
  Basics:
What types of magnets are there?
   by my.execpc.com    
  News & More:
Continued access to MRI for patients in Europe
Wednesday, 12 June 2013   by www.european-hospital.com    
Concentration And Visuospatial Awareness Affected By MRI Scanners
Thursday, 30 August 2012   by www.redorbit.com    
'Electromagnetic Wormhole' Possible with Invisibility Technology
Monday, 15 October 2007   by www.brightsurf.com    
Superconductors Face the Future
Wednesday, 15 September 2010   by www.nanotech-now.com    
MRI Resources 
Lung Imaging - Mass Spectrometry - Online Books - Veterinary MRI - Sequences - Case Studies
 
Quenching
 
A quench is the rapid helium evaporation and the loss of superconductivity of the current-carrying coil that may occur unexpectedly, or from pressing the emergency button in a superconducting magnet. As the superconductive magnet becomes resistive, heat will be released that can result in boiling of liquid helium in the cryostat. This may present a hazard if not properly planned for.
The evaporated coolant requires emergency venting systems to protect patients and operators. Quenching can cause total magnet failure and cannot be stopped. MRI systems are designed such that all of the escaping cryogenic gas is directed out of the building (quench pipe through the roof or the wall). In the event of a burst of the tank (possible in the case of an accident) or a blockage of the pipes, the helium gas will be forced into the scanner room, giving rise to a large white cloud of chilled gas. Under such circumstances it is essential that the scanner room is evacuated, also caused by the displacement of oxygen, which under extreme conditions could lead to asphyxiation. The force of quenching can be strong enough to destroy the walls of the scanner room or the MRI equipment.
spacer

• View the DATABASE results for 'Quenching' (5).Open this link in a new window

MRI Resources 
Software - Breast MRI - Societies - PACS - Mobile MRI - Artifacts
 
Panorama 0.23T™InfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
From Philips Medical Systems;
www.medical.philips.com/main/products/mri/products/panoramafamily/panorama0.23t_rt/features/ the Panorama 0.23 T, providing a new design optimized for patient comfort, faster reconstruction time than before (300 images/second) and new gradient specifications. Philips' Panorama 0.23 T I/T supports MR-guided interventions, resulting in minimally invasive procedures, more targeted surgery, reduced recovery time and shorter hospital stays. Optional OptoGuide functionality enables real-time needle tracking. Philips' Panorama 0.23 T™ R/T is the first and only open MRI system to enable radiation therapy planning using MR data sets. The Panorama also features the new and consistent Philips User Interface, an essential element of the Vequion clinical IT family of products and services.

Device Information and Specification
CLINICAL APPLICATION Whole body
CONFIGURATION Open MRI/C-arm
SURFACE COILS Head, head-neck, extremity M-L, neck, body/spine S-XL, shoulder, bilateral breast, wrist, TMJ, flex XS-S-M-L-XL-XXL
SPECTROSCOPY No
SYNCHRONIZATION ECG/peripheral: Optional/optional, respiratory gating
PULSE SEQUENCES SE, FE, IR, FFE, DEFFE, DESE, TSE, DETSE, Single shot SE, DRIVE, Balanced FFE, MRCP, Fluid Attenuated Inversion Recovery, Turbo FLAIR, IR-TSE, T1-STIR TSE, T2-STIR TSE, Diffusion Imaging, 3D SE, 3D FFE, MTC;; Angiography: CE-ANGIO, MRA 2D, 3D TOF
IMAGING MODES Single, multislice, volume study, dynamic, SIMEX, multi chunk 3D, multiple stacks
TR Min. 6.2 msec
TE Min. 2.8 msec
SINGLE/MULTI SLICE 50 slices/sec
FOV 0.4 cm - 40 cm
SLICE THICKNESS 0.4 mm
DISPLAY MATRIX 1280 X 1024
MEASURING MATRIX Up to 512 x 512
PIXEL INTENSITY 256 gray scale
SPATIAL RESOLUTION 0.78 mm
MAGNET TYPE Resistive/iron core
BORE DIAMETER
or W x H
Open x 46 cm x infinite (side-first patient entry)
MAGNET WEIGHT 13110 kg
H*W*D 196 x 121 x 176 cm
POWER REQUIREMENTS 400/480 V
COOLING SYSTEM TYPE Closed loop chilled water (chiller included)
CRYOGEN USE N/A
FIELD STRENGTH 0.23 T
STRENGTH 19 mT/m
5-GAUSS FRINGE FIELD 2.4 m / 3.7 m
SHIMMING Passive/active
spacer

• View the DATABASE results for 'Panorama 0.23T™' (2).Open this link in a new window

 
Further Reading:
  News & More:
Magnetic resonance imaging guided musculoskeletal interventions at 0.23T: Chapter 4. Materials and methods
2002
MRI Resources 
Education pool - Cardiovascular Imaging - Online Books - MRI Centers - Hospitals - Intraoperative MRI
 
     1 - 5 (of 5)     
Result Pages : [1]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?  



In 2075 (after about 100 years of ...) the MRI scan will be :
obsolete 
done with handheld probe 
done at home (app, ...) 
a 3 second walk through 
daily done 
replaced by something much ... 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • US-TIP • 
Copyright © 2003 - 2018 SoftWays. All rights reserved. [ 14 December 2018]
Terms of Use | Privacy Policy | Advertising
 [last update: 2018-03-08 05:11:00]