Transverse Magnetization
 
The xy component of the net magnetization vector at right angles to the main magnetic field. The precession of the transverse magnetization at the Larmor frequency is responsible for the detectable MRI signal. In the absence of externally applied RF energy, the transverse magnetization will decay to zero with a characteristic time constant of T2, or more strictly T2*.
spacer
 
• Related Searches:
    • Pulse, 90°
    • Steady State Free Precession
    • Hahn Echo
    • Free Induction Decay
    • Longitudinal Magnetization
 
Further Reading:
  Basics:
Measuring T1 and T2 Relaxation - Introductory NMR & MRI from Magritek
   by www.azom.com    
Searchterm 'Magnetization' was also found in the following service: 
spacer
Radiology  (3) Open this link in a new window
Longitudinal Magnetization
 
(Mz) The component of the net magnetization vector in the direction of the static magnetic field (z). After RF excitation, this vector returns to its equilibrium value at a rate characterized by the time constant T1.
spacer

• View the DATABASE results for 'Longitudinal Magnetization' (25).Open this link in a new window

MRI Resources 
Cardiovascular Imaging - Veterinary MRI - RIS - Corporations - Guidance - Directories
 
Magnetization Transfer Contrast
 
(MTC) This MRI method increases the contrast by removing a portion of the total signal in tissue. An off resonance radio frequency (RF) pulse saturates macromolecular protons to make them invisible (caused by their ultra-short T2* relaxation times). The MRI signal from semi-solid tissue like brain parenchyma is reduced, and the signal from a more fluid component like blood is retained.
E.g., saturation of broad spectral lines may produce decreases in intensity of lines not directly saturated, through exchange of magnetization between the corresponding states; more closely coupled states will show a greater resulting intensity change. Magnetization transfer techniques make demyelinated brain or spine lesions (as seen e.g. in multiple sclerosis) better visible on T2 weighted images as well as on gadolinium contrast enhanced T1 weighted images.
Off resonance makes use of a selection gradient during an off resonance MTC pulse. The gradient has a negative offset frequency on the arterial side of the imaging volume (caudally more off resonant and cranially less off resonant). The net effect of this type of pulse is that the arterial blood outside the imaging volume will retain more of its longitudinal magnetization, with more vascular signal when it enters the imaging volume. Off resonance MTC saturates the venous blood, leaving the arterial blood untouched.
On resonance has no effect on the free water pool but will saturate the bound water pool and is the difference in T2 between the pools. Special binomial pulses are transmitted causing the magnetization of the free protons to remain unchanged. The z-magnetization returns to its original value. The spins of the bound pool with a short T2 experience decay, resulting in a destroyed magnetization after the on resonance pulse.

See also Magnetization Transfer.
spacer

• View the DATABASE results for 'Magnetization Transfer Contrast' (5).Open this link in a new window

 
Further Reading:
  News & More:
MRI of the Human Eye Using Magnetization Transfer Contrast Enhancement
   by www.iovs.org    
Searchterm 'Magnetization' was also found in the following services: 
spacer
News  (1)  Forum  (3)  
 
Macroscopic Magnetization Vector
 
Net magnetic moment per unit volume (a vector quantity) of a sample in a given region, considered as the integrated effect of all the individual microscopic nuclear magnetic moments.
spacer

• View the DATABASE results for 'Macroscopic Magnetization Vector' (9).Open this link in a new window

Searchterm 'Magnetization' was also found in the following service: 
spacer
Radiology  (3) Open this link in a new window
Magnetization Prepared Rapid Gradient EchoInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(MP-GRE / MPRAGE / MP-RAGE) A fast 3D gradient echo pulse sequence using a magnetization preparation pulse like TurboFLASH. Only one segment or partition of a 3D data record is obtained per inversion preparation pulse. After the acquisition, for all rows a delay time (TD) is used to prevent saturation effects.
MPRAGE is designed for rapid acquisition with T1 weighted dominance. Fast gradient echoes are characterized by their rapid sampling time, high signal intensity and image contrast while approaching steady state (the echo is collected during the time when tissues are experiencing T1 relaxation). The rapid speed of the acquisition makes it an excellent alternative to breath-hold abdominal imaging, neuro, dynamic bolus, MR angiography and cardiac imaging.

See Gradient Echo Sequence.
 
Images, Movies, Sliders:
 Brain MRI Sagittal T1 001  Open this link in a new window
    
 
spacer

• View the DATABASE results for 'Magnetization Prepared Rapid Gradient Echo' (3).Open this link in a new window

MRI Resources 
Artifacts - Mobile MRI - Jobs pool - Supplies - Functional MRI - Process Analysis
 
previous      6 - 10 (of 127)     next
Result Pages : [1 2 3]  [4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Next big thing in MRI will be :
AI 
remote operator 
personalized protocols 
helium-free 
molecular MRI 
portable MRI 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 9 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]