Net Magnetization Vector
 
Hydrogen nuclei magnetic moments are randomly oriented in the absence of an external magnetic field and are considered to have a net magnetization of zero. Once hydrogen protons are placed in the presence of an external magnetic field, they align themselves in one of two directions, parallel or anti parallel to the net magnetic field, which is commonly referred to as the vector B0. The parallel and anti parallel protons cancel each other out, only the small number of low energy protons left aligned with the magnetic field create the overall net magnetization, this difference is all that counts. The magnetic moments of these protons are added together and are referred to as net magnetization vector (NMV) or the symbol 'M'.

See also Magnetization Transfer Contrast.
spacer
 
• Share the entry 'Net Magnetization Vector':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Flip Angle
    • Magnetization Transfer
    • Magnetization Transfer Contrast
    • Free Induction Decay
    • Resonance Frequency
 
Further Reading:
  Basics:
Mapping of low flip angles in magnetic resonance(.pdf)
Saturday, 1 January 2011   by www.hal.inserm.fr    
MRI Resources 
Education pool - Developers - MRI Physics - Hospitals - DICOM - Safety Training
 
Longitudinal Magnetization
 
(Mz) The component of the net magnetization vector in the direction of the static magnetic field (z). After RF excitation, this vector returns to its equilibrium value at a rate characterized by the time constant T1.
spacer

• View the DATABASE results for 'Longitudinal Magnetization' (25).Open this link in a new window

MRI Resources 
MRI Technician and Technologist Schools - Stimulator pool - Intraoperative MRI - Cardiovascular Imaging - Coils - Implant and Prosthesis pool
 
MagnitudeForum -
related threads
 
In physics, the magnitude (length) of a vector is a scalar in the physical sense, i.e. a physical quantity independent of the coordinate system, expressed as the product of a numerical value and a physical unit, not just a number.

See also Longitudinal Magnetization and Net Magnetization Vector.
spacer

• View the DATABASE results for 'Magnitude' (26).Open this link in a new window

 
Further Reading:
  Basics:
Less fit marathoners prone to heart damage for 3 months after race
Monday, 25 October 2010   by sify.com    
  News & More:
Medical Magnetic Resonance Imaging (MRI): Technologies and Global Markets
Wednesday, 9 October 2013   by www.digitaljournal.com    
MRI Resources 
Bioinformatics - Used and Refurbished MRI Equipment - Databases - Services and Supplies - Devices - MRI Centers
 
Tip Angle
 
Angle between the net magnetization vector before and after a RF excitation pulse. Small tip angles allow a decrease in TR, which is used to decrease scan time in Field Echo pulse sequences. See Flip Angle.
spacer

• View the DATABASE results for 'Tip Angle' (5).Open this link in a new window

 
Further Reading:
  Basics:
Mapping of low flip angles in magnetic resonance(.pdf)
Saturday, 1 January 2011   by www.hal.inserm.fr    
MRI Resources 
Pathology - IR - Equipment - Supplies - Musculoskeletal and Joint MRI - Guidance
 
Transverse Magnetization
 
The xy component of the net magnetization vector at right angles to the main magnetic field. The precession of the transverse magnetization at the Larmor frequency is responsible for the detectable MRI signal. In the absence of externally applied RF energy, the transverse magnetization will decay to zero with a characteristic time constant of T2, or more strictly T2*.
spacer

• View the DATABASE results for 'Transverse Magnetization' (43).Open this link in a new window

 
Further Reading:
  Basics:
Measuring T1 and T2 Relaxation - Introductory NMR & MRI from Magritek
   by www.azom.com    
MRI Resources 
Anatomy - MRI Technician and Technologist Career - DICOM - MR Guided Interventions - Colonography - Pediatric and Fetal MRI
 
     1 - 5 (of 20)     next
Result Pages : [1]  [2 3 4]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 19 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]