Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Gauss' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Gauss' found in 6 terms [] and 68 definitions []
previous     6 - 10 (of 74)     next
Result Pages : [1 2]  [3 4 5 6 7 8 9 10 11 12 13 14 15]
Searchterm 'Gauss' was also found in the following services: 
spacer
News  (1)  Resources  (1)  Forum  (3)  
 
5 Gauss Fringe Field
 
This line specifies the perimeter around a MR scanner within which the static magnetic fields are higher than five gauss. Five gauss and below are considered 'safe' levels of static magnetic field exposure for the general public.
Due to the physical properties of magnetic fields, the magnetic flux, which penetrates the useful volume of the magnet will return through the surroundings of the magnet to form closed field lines. Depending on the magnet construction, the returning flux will penetrate large open spaces (unshielded magnets) or will be confined largely to iron yokes or through secondary coils (shielded magnets).
spacer
 
• Related Searches:
    • Tesla
    • 5 Gauss Limit
    • Low Field MRI
    • MRI Safety
    • Gauss
 
Further Reading:
  Basics:
How strong are magnets?
   by my.execpc.com    
  News & More:
Magnetic Field
   by hyperphysics.phy-astr.gsu.edu    
Searchterm 'Gauss' was also found in the following service: 
spacer
Radiology  (4) Open this link in a new window
MRI SafetyMRI Resource Directory:
 - Safety -
 
There are different types of contraindications that would prevent a person from being examined with an MRI scanner. MRI systems use strong magnetic fields that attract any ferromagnetic objects with enormous force. Caused by the potential risk of heating, produced from the radio frequency pulses during the MRI procedure, metallic objects like wires, foreign bodies and other implants needs to be checked for compatibility. High field MRI requires particular safety precautions. In addition, any device or MRI equipment that enters the magnet room has to be MR compatible. MRI examinations are safe and harmless, if these MRI risks are observed and regulations are followed.

Safety concerns in magnetic resonance imaging include:
•
the magnetic field strength;
•
possible 'missile effects' caused by magnetic forces;
•
the potential for heating of body tissue due to the application of the radio frequency energy;
•
the effects on implanted active devices such as cardiac pacemakers or insulin pumps;
•
magnetic torque effects on indwelling metal (clips, etc.);
•
the audible acoustic noise;
•
danger due to cryogenic liquids;
•
the application of contrast medium;
mri safety guidance
MRI Safety Guidance
It is important to remember when working around a superconducting magnet that the magnetic field is always on. Under usual working conditions the field is never turned off. Attention must be paid to keep all ferromagnetic items at an adequate distance from the magnet. Ferromagnetic objects which came accidentally under the influence of these strong magnets can injure or kill individuals in or nearby the magnet, or can seriously damage every hardware, the magnet itself, the cooling system, etc.. See MRI resources Accidents.
The doors leading to a magnet room should be closed at all times except when entering or exiting the room. Every person working in or entering the magnet room or adjacent rooms with a magnetic field has to be instructed about the dangers. This should include the patient, intensive-care staff, and maintenance-, service- and cleaning personnel, etc..
The 5 Gauss limit defines the 'safe' level of static magnetic field exposure. The value of the absorbed dose is fixed by the authorities to avoid heating of the patient's tissue and is defined by the specific absorption rate. Leads or wires that are used in the magnet bore during imaging procedures, should not form large-radius wire loops. Leg-to-leg and leg-to-arm skin contact should be prevented in order to avoid the risk of burning due to the generation of high current loops if the legs or arms are allowed to touch. The patient's skin should not be in contact with the inner bore of the magnet.
The outflow from cryogens like liquid helium is improbable during normal operation and not a real danger for patients.
The safety of MRI contrast agents is tested in drug trials and they have a high compatibility with very few side effects. The variations of the side effects and possible contraindications are similar to X-ray contrast medium, but very rare. In general, an adverse reaction increases with the quantity of the MRI contrast medium and also with the osmolarity of the compound.

See also 5 Gauss Fringe Field, 5 Gauss Line, Cardiac Risks, Cardiac Stent, dB/dt, Legal Requirements, Low Field MRI, Magnetohydrodynamic Effect, MR Compatibility, MR Guided Interventions, Claustrophobia, MRI Risks and Shielding.
Radiology-tip.comradRadiation Safety,  Ionizing Radiation
spacer
Medical-Ultrasound-Imaging.comUltrasound Safety,  Absorbed Dose
spacer

• View the DATABASE results for 'MRI Safety' (42).Open this link in a new window


• View the NEWS results for 'MRI Safety' (13).Open this link in a new window.
 
Further Reading:
  Basics:
MRI Safety
2001   by www.fda.gov    
What MRI Sequences Produce the Highest Specific Absorption Rate (SAR), and Is There Something We Should Be Doing to Reduce the SAR During Standard Examinations?
Thursday, 16 April 2015   by www.ajronline.org    
Contrast Agents: Safety Profile
   by www.clinical-mri.com    
  News & More:
How safe is 7T MRI for patients with neurosurgical implants?
Thursday, 17 November 2022   by healthimaging.com    
Newer Heart Devices Safe During MRI
Monday, 23 August 2004   by www.hospimedica.com    
Study: Face Masks Unsafe in MRI Machines
Wednesday, 13 July 2022   by www.laboratoryequipment.com    
COVID-19: Attention shifts to MRI infection control
Thursday, 9 July 2020   by https://www.auntminnieeurope.com/index.aspx?sec=ser§sub=def§pag=dis§ItemID=619012    
FDA Releases New Guidance On Establishing Safety, Compatibility Of Passive Implants In MR Environments
Tuesday, 16 December 2014   by www.meddeviceonline.com    
Modern Implantable Heart Devices Safe For Use In MRI Scans
Wednesday, 16 March 2005   by www.sciencedaily.com    
MRI Resources 
NMR - Safety Training - Bioinformatics - PACS - Developers - Patient Information
 
Faraday Shield
 
In electromagnetism, the Faraday cage or shield is an application of Gauss's law, one of Maxwell's equations. Gauss's law describes the distribution of electrical charge on a conducting form, such as a sphere, a plane, a torus, etc. Intuitively, since like charges repel each other, charge will "migrate" to the surface of the conducting form, as described below. The application is named after physicist Michael Faraday, who built the first Faraday cage in 1836, to demonstrate his finding. A Faraday shield is used generally for any kind of electrostatic shielding.
In MRI, one use of the Faraday shield is the shielding of the scanning room, to block incoming radio frequency (RF) signals which would contaminate the send and received signals of the MRI scanner, and it suppresses RF signals, which would else pollute the environment around.
spacer

• View the DATABASE results for 'Faraday Shield' (5).Open this link in a new window

 
Further Reading:
  Basics:
Faraday's Law
   by hyperphysics.phy-astr.gsu.edu    
Faraday_shield
   by en.wikipedia.org    
  News & More:
Electrical eddy currents in the human body: MRI scans and medical implants
   by www.phy.olemiss.edu    
Searchterm 'Gauss' was also found in the following services: 
spacer
News  (1)  Resources  (1)  Forum  (3)  
 
Magnetism
 
Magnetic forces are fundamental forces that arise due to the movement of electrical charge. Maxwell's equations describe the origin and behavior of the fields that govern these forces. Thus, magnetism is seen whenever electrically charged particles are in motion. This can arise either from movement of electrons in an electric current, resulting in 'electromagnetism', or from the quantum-mechanical orbital motion (there is no orbital motion of electrons around the nucleus like planets around the sun, but there is an 'effective electron velocity') and spin of electrons, resulting in what are known as 'permanent magnets'.
The physical cause of the magnetism of objects, as distinct from electrical currents, is the atomic magnetic dipole. Magnetic dipoles, or magnetic moments, result on the atomic scale from the two kinds of movement of electrons. The first is the orbital motion of the electron around the nucleus this motion can be considered as a current loop, resulting in an orbital dipole magnetic moment along the axis of the nucleus. The second, much stronger, source of electronic magnetic moment is due to a quantum mechanical property called the spin dipole magnetic moment.
Gauss (G) and tesla (T) are units to define the intensity of magnetic fields. One tesla is equivalent to 10 000 gauss.
Typically, the field strength of MRI scanners is between 0.15 T and 3 T.

See also Diamagnetism, Paramagnetism, Superparamagnetism, and Ferromagnetism.
spacer

• View the DATABASE results for 'Magnetism' (18).Open this link in a new window


• View the NEWS results for 'Magnetism' (1).Open this link in a new window.
 
Further Reading:
  Basics:
Magnet basics
   by my.execpc.com    
  News & More:
What affects the strength of a magnet?
   by my.execpc.com    
Searchterm 'Gauss' was also found in the following service: 
spacer
Radiology  (4) Open this link in a new window
Rayleigh Noise
 
The distribution associated with the magnitude of the noise amplitude following a Gaussian distribution. The mean value of this distribution is roughly 1.25 s0, where s0 is the standard deviation of the original Gaussian distribution.
spacer
MRI Resources 
IR - Implant and Prosthesis pool - Safety Products - Crystallography - Quality Advice - Knee MRI
 
previous      6 - 10 (of 74)     next
Result Pages : [1 2]  [3 4 5 6 7 8 9 10 11 12 13 14 15]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 28 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]