Digitization Noise ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
Digitization noise, digitizer quantization, low dynamic range
DESCRIPTION
Noise
REASON
Finite voltage resolution of the digitizer
HELP
Larger range of sensitivity
Noise introduced into digitized signals by the finite voltage resolution of the digitizer.
You can see the effects of quantization if the noise level is smaller than the digitizer quantum. If the signal dynamic range is too great, the highest intensities from overloading the digitizer may result in the weaker features being lost in the digitization noise.
mri safety guidance
Image Guidance
This can be resolved by using an analog to digital converter with a larger range of sensitivity or by using techniques to reduce the dynamic range, e.g. suppressing the signal from water in order to detect the signal from less abundant compounds.
spacer
 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
Searchterm 'View' was also found in the following services: 
spacer
Radiology  (40) Open this link in a new windowUltrasound  (50) Open this link in a new window
Drug Development and Approval Process (USA)
 
Different stages of the drug development and approval process in the USA, lead from preclinical trials (testing in animals), first application in humans through limited and broad clinical tests, to postmarketing studies.
Drug Development and Approval Process (USA)
By Dale E. Wierenga, Ph.D. and C. Robert Eaton
Office of Research and Development
Pharmaceutical Manufacturers Association

'In reviewing this report, it is important to keep in mind the realities of the drug discovery and development process. The U.S. system of new drug approvals is perhaps the most rigorous in the world. On average, it costs a company $359 million to get one new medicine from the laboratory to the pharmacist's shelf, according to a February 1993 report by the Congressional Office of Technology Assessment.'

See also Phase 1 2 3 4 Drug Trials, Food and Drug Administration, and European Medicines Agency.
spacer
MRI Resources 
Societies - Education - Cochlear Implant - MRI Training Courses - RIS - Case Studies
 
Echelon™ 1.5TInfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.
 
www.hitachimed.com/contentindex.asp?ID=971 From Hitachi Medical Systems America Inc.;
Hitachi expanded its portfolio with the Echelon™ 1.5T. The MRI scanner combines a compact magnet and a scalable 8-channel RF system with high-performance gradients and slew rate to select short echo times, small field of views, high matrices and thin slices. Standard features of the Echelon MRI system include higher-order active shim, RAPID (parallel imaging for use on brain MRI, body, cardiovascular imaging, and orthopedic coils), multiple coil ports, and an advanced reconstruction engine.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Short bore
Head, body coil, spine, breast, knee, shoulder, vascular multiple array coils.
SYNCHRONIZATION
Cardiac gating, ECG/peripheral, respiratory gating
PULSE SEQUENCES
SE, IR, FSE, FIR, GE, SG, BASG, PBSG, PCIR, DWI, Radial, Angiography: TOF, FLUTE (Fluoro-triggered bolus MRA), Time-resolved MRA
IMAGING MODES
Single, multislice, volume study
PIXEL INTENSITY
Level Range: -2,000 to +4,000
Sub millimeter
POWER REQUIREMENTS
208/220/240 V, single phase
CRYOGEN USE
Low cryogen boil-off
STRENGTH
30 mT/m
150 T/m/sec
Higher-order active shim
spacer

• View the DATABASE results for 'Echelon™ 1.5T' (2).Open this link in a new window


• View the NEWS results for 'Echelon™ 1.5T' (3).Open this link in a new window.
 
Further Reading:
  Basics:
Echelon 1.5T
   by www.hitachimed.com    
Searchterm 'View' was also found in the following services: 
spacer
News  (56)  Resources  (52)  Forum  (50)  
 
Eddy Current ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.
 
Quick Overview
Artifact Information
NAME
Eddy current
DESCRIPTION
Image distortion
REASON
Induced by changes in a magnetic field
Eddy currents can cause artifacts in images and may seriously degrade overall magnet performance.
mri safety guidance
Image Guidance
Common uses to reduce the influence of eddy currents on gradient fields are eddy current compensation and shielded gradient coils (active or passive). If persistent, it must be addressed by a service representative to verify the magnet performance.

See also Eddy Current Compensation.
spacer

• View the DATABASE results for 'Eddy Current Artifact' (2).Open this link in a new window

 
Further Reading:
  Basics:
Generic Eddy Current Compensation for Rapid Magnetic Resonance Imaging(.pdf)
   by www.switt.ch    
Searchterm 'View' was also found in the following services: 
spacer
Radiology  (40) Open this link in a new windowUltrasound  (50) Open this link in a new window
FID Signal ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
FID (with spin echo) signal
DESCRIPTION
Line across the center of the image
REASON
Combination of problems
HELP
Call the service
A combination of B1 inhomogeneity, poor slice profile, and insufficient spoiler gradients between the refocusing pulse and the readout interval of a spin echo sequence results in a FID signal being detected along with the echo.
Since the FID is not phase encoded (normally the phase encoding occurs before the refocusing pulse), it is not dispersed along the phase encoding axis, but appears as a line across the center of the image.
mri safety guidance
Image Guidance
If the problem persists, it must be addressed by a service representative.
spacer

• View the DATABASE results for 'FID Signal Artifact' (2).Open this link in a new window

 
Further Reading:
  News & More:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
MRI Resources 
MRI Reimbursement - MRI Technician and Technologist Career - General - Breast Implant - Developers - Open Directory Project
 
previous      41 - 45 (of 104)     next
Result Pages : [1]  [2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Next big thing in MRI will be :
AI 
remote operator 
personalized protocols 
helium-free 
molecular MRI 
portable MRI 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 3 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]