Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Inhomogeneity' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Inhomogeneity' found in 2 terms [] and 20 definitions []
previous     6 - 10 (of 22)     next
Result Pages : [1]  [2 3 4 5]
Searchterm 'Inhomogeneity' was also found in the following services: 
spacer
Forum  (2)  
 
DixonInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
The Dixon technique is a MRI method used for fat suppression and/or fat quantification. The difference in magnetic resonance frequencies between fat and water-bound protons allows the separation of water and fat images based on the chemical shift effect.
This imaging technique is named after Dixon, who published in 1984 the basic idea to use phase differences to calculate water and fat components in postprocessing. Dixon's method relies on acquiring an image when fat and water are 'in phase', and another in 'opposed phase' (out of phase). These images are then added together to get water-only images, and subtracted to get fat-only images. Therefore, this sequence type can deliver up to 4 contrasts in one measurement: in phase, opposed phase, water and fat images. An additional benefit of Dixon imaging is that source images and fat images are also available to the diagnosing physician.
The original two point Dixon sequence (number of points means the number of images acquired at different TE) had limited possibilities to optimize the echo time, spatial resolution, slice thickness, and scan time; but Dixon based fat suppression can be very effective in areas of high magnetic susceptibility, where other techniques fail. This insensitivity to magnetic field inhomogeneity and the possibility of direct image-based water and fat quantification have currently generated high research interests and improvements to the basic method (three point Dixon).
The combination of Dixon with gradient echo sequences allows for example liver imaging with 4 image types in one breath hold. With Dixon TSE/FSE an excellent fat suppression with high resolution can be achieved, particularly useful in imaging of the extremities.
For low bandwidth imaging, chemical shift correction of fat images can be made before recombination with water images to produce images free of chemical shift displacement artifacts. The need to acquire more echoes lengthens the minimum scan time, but the lack of fat saturation pulses extends the maximum slice coverage resulting in comparable scan time.
spacer
 
Further Reading:
  Basics:
Separation of fat and water signal in magnetic resonanace imaging
2011   by www.diva-portal.org    
Direct Water and Fat Determination in Two-Point Dixon Imaging
April 2013   by scholarship.rice.edu    
MRI evaluation of fatty liver in day to day practice: Quantitative and qualitative methods
Wednesday, 3 September 2014   by www.sciencedirect.com    
Measurement of Fat/Water Ratios in Rat Liver Using 3DThree-Point Dixon MRI
2004   by www.civm.duhs.duke.edu    
  News & More:
The utility of texture analysis of kidney MRI for evaluating renal dysfunction with multiclass classification model
Tuesday, 30 August 2022   by www.nature.com    
Liver Imaging Today
Friday, 1 February 2013   by www.healthcare.siemens.it    
mDIXON being developed to simplify and accelerate liver MRI
September 2010   by incenter.medical.philips.com    
Searchterm 'Inhomogeneity' was also found in the following service: 
spacer
Ultrasound  (1) Open this link in a new window
Echo Offset
 
Echo offset is the time setting of spin echo and gradient echo to be not coincident and to generate phase differences between different spectral line signals (e.g., water and fat). The echo offset is the product of the frequency line difference and the time difference (TD) in the echo times and is equal to the magnitude of the result of the phase difference between two spectral lines. Phases may not change linearly with echo offset time in the presence of a large field inhomogeneity. An echo offset excitation pulse sequence can be used in the magnetic field mapping method, to generate maps from which the standard deviation of the phase difference can be calculated.
spacer
 
Further Reading:
  News & More:
T1rho-prepared balanced gradient echo for rapid 3D T1rho MRI
Monday, 1 September 2008   by www.ncbi.nlm.nih.gov    
A short-TR single-echo spin-echo breath-hold method for assessing liver T2
Sunday, 10 December 2023   by link.springer.com    
MRI Resources 
NMR - MRI Accidents - Software - Corporations - Collections - Knee MRI
 
FID Signal ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
FID (with spin echo) signal
DESCRIPTION
Line across the center of the image
REASON
Combination of problems
HELP
Call the service
A combination of B1 inhomogeneity, poor slice profile, and insufficient spoiler gradients between the refocusing pulse and the readout interval of a spin echo sequence results in a FID signal being detected along with the echo.
Since the FID is not phase encoded (normally the phase encoding occurs before the refocusing pulse), it is not dispersed along the phase encoding axis, but appears as a line across the center of the image.
mri safety guidance
Image Guidance
If the problem persists, it must be addressed by a service representative.
spacer

• View the DATABASE results for 'FID Signal Artifact' (2).Open this link in a new window

 
Further Reading:
  News & More:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
Searchterm 'Inhomogeneity' was also found in the following services: 
spacer
Forum  (2)  
 
Field of View
 
(FOV) Defined as the size of the two or three dimensional spatial encoding area of the image. Usually defined in units of mm². The FOV is the square image area that contains the object of interest to be measured. The smaller the FOV, the higher the resolution and the smaller the voxel size but the lower the measured signal. Useful for decreasing the scantime is a field of view different in the frequency and phase encoding directions (rectangular field of view - RFOV).
The magnetic field homogeneity decreases as more tissue is imaged (greater FOV). As a result the precessional frequencies change across the imaging volume. That can be a problem for fat suppression imaging. This fat is precessing at the expected frequency only in the center of the imaging volume. E.g. frequency specific fat saturation pulses become less effective when the field of view is increased. It is best to use smaller field of views when applying fat saturation pulses.
mri safety guidance
Image Guidance
Smaller FOV required higher gradient strength and concludes low signal. Therefore you have to find a compromise between these factors. The right choice of the field of view is important for MR image quality. When utilizing small field of views and scanning at a distance from the isocenter (more problems with artifacts) it is obviously important to ensure that the region of interest is within the scanning volume.
A smaller FOV in one direction is available with the function rectangular field of view (RFOV).

See also Field Inhomogeneity Artifact.
 
Images, Movies, Sliders:
 MRI - Anatomic Imaging of the Foot  Open this link in a new window
    
SlidersSliders Overview

 MRI - Anatomic Imaging of the Ankle 1  Open this link in a new window
    
SlidersSliders Overview

 
spacer

• View the DATABASE results for 'Field of View' (27).Open this link in a new window

 
Further Reading:
  Basics:
Image Characteristics and Quality
   by www.sprawls.org    
  News & More:
Optimizing Musculoskeletal MR
   by rad.usuhs.mil    
Path Found to a Combined MRI and CT Scanner
Wednesday, 20 March 2013   by spectrum.ieee.org    
Searchterm 'Inhomogeneity' was also found in the following service: 
spacer
Ultrasound  (1) Open this link in a new window
Gradient EchoForum -
related threads
 
(GE) An echo signal generated from a free induction decay by means of a bipolar switched magnetic gradient. The echo is produced by reversing the direction of a magnetic field gradient or by applying balanced pulses of magnetic field gradient before and after a refocusing RF pulse so as to cancel out the position dependent phase shifts that have accumulated due to the gradient.
In the latter case, the gradient echo is generally adjusted to be coincident with the RF spin echo. When the RF and gradient echoes are not coincident, the time of the gradient echo is denoted echo time (TE) and the difference in time between the echoes is denoted time difference (TD).
Gradient echo does not refocus the effects of main field inhomogeneity and therefore is generally used with a short echo time. Disadvantages of gradient echo imaging are compromised anatomic details and artifacts in regions with varying susceptibility e.g. between the air-containing sinuses and brain and especially between haemorrhages and normal tissue.

See also Susceptibility Artifact.
spacer

• View the DATABASE results for 'Gradient Echo' (121).Open this link in a new window

 
Further Reading:
  Basics:
Mapping of low flip angles in magnetic resonance(.pdf)
Saturday, 1 January 2011   by www.hal.inserm.fr    
A LANTHANIDE LANTHOLOGY(.pdf)
   by www.phy.davidson.edu    
Enhanced Fast GRadient Echo 3-Dimensional (efgre3D) or THRIVE
   by www.mri.tju.edu    
MRI Resources 
Shoulder MRI - Open Directory Project - Guidance - Universities - Sequences - Stent
 
previous      6 - 10 (of 22)     next
Result Pages : [1]  [2 3 4 5]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 26 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]