Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Magnetization' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Magnetization' found in 11 terms [] and 116 definitions []
previous     16 - 20 (of 127)     next
Result Pages : [1 2 3]  [4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
Searchterm 'Magnetization' was also found in the following services: 
spacer
News  (1)  Forum  (3)  
 
Coherent Gradient EchoInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
Coherent gradient echo sequences can measure the free induction decay (FID), generated just after each excitation pulse or the echo formed prior to the next pulse. Coherent gradient echo sequences are very sensitive to magnetic field inhomogeneity. An alternative to spoiling is to incorporate residual transverse magnetization directly into the longitudinal steady state. These GRE sequences use a refocusing gradient in the phase encoding direction during the end module to maximize remaining transverse (xy) magnetization at the time when the next excitation is due, while the other two gradients are, in any case, balanced.
When the next excitation pulse is sent into the system with an opposed phase, it tilts the magnetization in the -a direction. As a result the z-magnetization is again partly tilted into the xy-plane, while the remaining xy-magnetization is tilted partly into the z-direction.
A fully refocused sequence with a properly selected and uniform f would yield higher signal, especially for tissues with long T2 relaxation times (high water content) so it is used in angiographic, myelographic or arthrographic examinations and is used for T2* weighting. The repetition time for this sequence has to be short. With short TR, coherent GE is also useable for breath hold and 3D technique. If the repetition time is about 200 msec there's no difference between spoiled or unspoiled GE. T1 weighting is better with spoiled techniques.
The common types include GRASS, FISP, FAST, and FFE.
The T2* component decreases with long TR and short TE. The T1 time is controlled by flip angle. The common TR is less than 50 ms and the common TE less than 15 ms
Other types have stronger T2 dependence but lower SNR. They include SSFP, CE-FAST, PSIF, and CE-FFE-T2.
Examples of fully refocused FID sequences are TrueFISP, bFFE and bTFE.
spacer
 
• Related Searches:
    • Coherent
    • Incoherent Gradient Echo (Gradient Spoiled)
    • Short Repetition Techniques
    • Incoherent Gradient Echo (RF Spoiled)
    • Spoiled Gradient Echo Sequence
Searchterm 'Magnetization' was also found in the following service: 
spacer
Radiology  (3) Open this link in a new window
Incoherent Gradient Echo (RF Spoiled)InfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
A gradient echo is generated by using a pair of bipolar gradient pulses. The gradient field is negatively pulsed, causing the spins of the xy-magnetization to dephase. A second gradient pulse is applied with the opposite polarity. During the pulsing, the spins that dephased begin to rephase and generate a gradient echo.
Spoiling can be accomplished by RF or a gradient. The incoherent RF spoiled type of a gradient echo sequence use a continuous shifting of the RF pulse to spoil the residual transverse magnetization. The phase of the RF excitation and receiver channel are varied pseudo randomly with each excitation cycle to prevent the xy magnetization from achieving steady state. T2* does not dominate image contrast, so T1 and PD weighting is practical. This method is effective and can be used to achieve a shorter TR, due to a lack of additional gradients. Spoiling eliminates the effect of the remaining xy-magnetization and leads to steady state longitudinal magnetization. These sequences can be used for breath hold, dynamic imaging and in cine and volume acquisitions.
spacer
MRI Resources 
Pregnancy - Musculoskeletal and Joint MRI - Implant and Prosthesis pool - Directories - Spectroscopy - General
 
Spoiler Gradient PulseInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
Magnetic field gradient pulse applied to effectively remove transverse magnetization by producing a rapid variation of its phase along the direction of the gradient. This is done after the echo so that transverse magnetization is destroyed prior to the next excitation pulse, to spoil any remaining xy-magnetization or to refocus the xy-magnetization.
For example, when used to remove the unwanted signal resulting from an imperfect 180° refocusing RF pulse, a corresponding compensating gradient pulse may be applied prior to the refocusing RF pulse in order to avoid spoiling the desired transverse magnetization resulting from the initial excitation. Also called homospoil pulse.
spacer

• View the DATABASE results for 'Spoiler Gradient Pulse' (6).Open this link in a new window

 
Further Reading:
  News & More:
Clinical evaluation of a speed optimized T2 weighted fast spin echo sequence at 3.0 T using variable flip angle refocusing, half-Fourier acquisition and parallel imaging
Wednesday, 25 October 2006
Searchterm 'Magnetization' was also found in the following services: 
spacer
News  (1)  Forum  (3)  
 
Ferromagnetism
 
Ferromagnetism is a phenomenon by which a material can exhibit a spontaneous magnetization: a net magnetic moment in the absence of an external magnetic field. More recently: a material is ferromagnetic, only if all of its magnetic ions add a positive contribution to the net magnetization (for differentiation to ferrimagnetic and antiferromagnetic materials). If some of the magnetic ions subtract from the net magnetization (if they are partially anti-aligned), then the material is ferrimagnetic. If the ions anti-align completely so as to have zero net magnetization, despite the magnetic ordering, then it is an antiferromagnet. All of these alignment effects only occur at temperatures below a certain critical temperature, called the Curie temperature (for ferromagnets and ferrimagnets) or the Néel temperature (for antiferromagnets). Typical ferromagnetic materials are iron, cobalt, and nickel.
In MRI ferromagnetic objects, even very small ones, as implants or incorporations distort the homogeneity of the main magnetic field and cause susceptibility artifacts.
spacer

• View the DATABASE results for 'Ferromagnetism' (7).Open this link in a new window

 
Further Reading:
  Basics:
Magnet basics
   by my.execpc.com    
Ferromagnetism
   by en.wikipedia.org    
Superconducting Magnets
   by hyperphysics.phy-astr.gsu.edu    
  News & More:
Physicists observe an exotic 'multiferroic' state in an atomically thin material
Wednesday, 23 February 2022   by www.sciencedaily.com    
Searchterm 'Magnetization' was also found in the following service: 
spacer
Radiology  (3) Open this link in a new window
Nuclear Magnetic Resonance Signal
 
The electromagnetic signal in the radio-frequency range produced by the precession of the transverse magnetization of the spins. The rotation of the transverse magnetization induces a voltage in a receiving antenna (coil), which is amplified and demodulated by the receiver circuits. Electromagnetic signal in the radio frequency range produced by the precession of the transverse magnetization of the spins. The rotation of the transverse magnetization induces a voltage in a coil, which is amplified and demodulated by the receiver;; the signal may refer only to this induced voltage.
spacer

• View the DATABASE results for 'Nuclear Magnetic Resonance Signal' (2).Open this link in a new window

 
Further Reading:
  Basics:
Spin echoes, CPMG and T2 relaxation - Introductory NMR & MRI from Magritek
2013   by www.azom.com    
NMR Spectroscopy - Theory
   by www.shu.ac.uk    
  News & More:
A powder to enhance NMR signals
Thursday, 12 December 2013   by phys.org    
MRI Resources 
Mobile MRI - Non-English - Implant and Prosthesis - MRI Technician and Technologist Schools - MRI Training Courses - Cochlear Implant
 
previous      16 - 20 (of 127)     next
Result Pages : [1 2 3]  [4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 9 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]