Steady State Free Precession SequenceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(SFP or SSFP) Steady state free precession is any field or gradient echo sequence where the TR is shorter than the T1 and T2 times of the tissue.
The flip angle and the TR maintain the steady state. The flip angle should be 60-90° if the TR is 100 ms, if the TR is less than 100 ms, than the choice of the flip angle for steady state is 45-60°. The T1 weighting is controlled by TR and flip, the T2 weighting increases with the TE. Common TR is between 20 - 50 msec.
spacer
 
Further Reading:
  News & More:
Generic Eddy Current Compensation for Rapid Magnetic Resonance Imaging(.pdf)
   by www.switt.ch    
Cutting Edge Imaging of THE Spine
February 2007   by www.pubmedcentral.nih.gov    
MRI Resources 
Non-English - Libraries - Colonography - MRI Accidents - IR - MR Myelography
 
Double Mode Steady State Free PrecessionInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
spacer
MRI Resources 
Hospitals - Bioinformatics - General - Distributors - Universities - Stimulator pool
 
Fast Imaging with Steady State PrecessionInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(FISP) A fast imaging sequence, which attempts to combine the signals observed separately in the FADE sequence, generally sensitive about magnetic susceptibility artifacts and imperfections in the gradient waveforms. Confusingly now often used to refer to a refocused FLASH type sequence.
This sequence is very similar to FLASH, except that the spoiler pulse is eliminated. As a result, any transverse magnetization still present at the time of the next RF pulse is incorporated into the steady state. FISP uses a RF pulse that alternates in sign. Because there is still some remaining transverse magnetization at the time of the RF pulse, a RF pulse of a degree flips the spins less than a degree from the longitudinal axis. With small flip angles, very little longitudinal magnetization is lost and the image contrast becomes almost independent of T1. Using a very short TE (with TR 20-50 ms, flip angle 30-45°) eliminates T2* effects, so that the images become proton density weighted. As the flip angle is increased, the contrast becomes increasingly dependent on T1 and T2*. It is in the domain of large flip angles and short TR that FISP exhibits vastly different contrast to FLASH type sequences. Used for T1 orthopedic imaging, 3D MPR, cardiography and angiography.
spacer

• View the DATABASE results for 'Fast Imaging with Steady State Precession' (5).Open this link in a new window

 
Further Reading:
  Basics:
MRI techniques improve pulmonary embolism detection
Monday, 19 March 2012   by medicalxpress.com    
Searchterm 'Precess' was also found in the following services: 
spacer
News  (3)  Forum  (5)  
 
Reverse Fast Imaging with Steady State PrecessionInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(PSIF) A heavily T2* weighted contrast enhanced gradient echo (mirrored FISP) technique. Because TE is relatively long, there are much flow artifacts and less signal to noise. In normal gradient echo techniques a FID-signal results after the RF pulses. This FID is rephased very fast and just before the next FID follows a spin echo signal. The SE is spoiled in FLASH sequences, but with PSIF sequences, only the SE is measured, not the FID.
spacer

• View the DATABASE results for 'Reverse Fast Imaging with Steady State Precession' (2).Open this link in a new window

 
Further Reading:
  News & More:
Fast T2 weighted imaging by PSIF at 0.2T for interventional MRI.(.pdf)
   by cds.ismrm.org    
MRI Resources 
Shoulder MRI - Bioinformatics - MRI Centers - Process Analysis - Case Studies - Blood Flow Imaging
 
Larmor Frequency
 
The Larmor precession frequency is the rate of precession of a spin packet under the influence of a magnetic field. The frequency of an RF signal, which will cause a change in the nucleus spin energy level, is given by the Larmor equation. The frequency is determined by the gyro magnetic ratio of atoms and the strength of the magnetic field. The gyromagnetic ratio is different for each nucleus of different atoms.
The stronger the magnetic field, the higher the precessional frequency. If an RF pulse at the Larmor frequency is applied to the nucleus of an atom, the protons will alter their alignment from the direction of the main magnetic field to the direction opposite the main magnetic field. As the proton tries to realign with the main magnetic field, it will emit energy at the Larmor frequency. By varying the magnetic field across the body with a magnetic field gradient, the corresponding variation of the Larmor frequency can be used to encode the position. For protons (hydrogen nuclei), the Larmor frequency is 42.58 MHz/Tesla.

See also Larmor Equation.
spacer

• View the DATABASE results for 'Larmor Frequency' (27).Open this link in a new window

 
Further Reading:
  Basics:
Magnetic resonance imaging
   by www.scholarpedia.org    
  News & More:
Magnetic resonance-guided motorized transcranial ultrasound system for blood-brain barrier permeabilization along arbitrary trajectories in rodents
Thursday, 24 December 2015   by www.ncbi.nlm.nih.gov    
MRI Resources 
MR Guided Interventions - Breast Implant - RIS - Artifacts - Sequences - Universities
 
previous      6 - 10 (of 52)     next
Result Pages : [1 2]  [3 4 5 6 7 8 9 10 11]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 4 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]