Displacement Encoding with Stimulated EchoesInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(DENSE) Displacement Encoding with Stimulated Echoes is a functional cardiac MRI pulse sequence, used to create maps of myocardial displacement with high resolution.
The DENSE magnitude images produce black blood images to show better myocard-blood contrast and to reduce motion artifacts.

See also Myocardial Late Enhancement, Spin Tagging, Coronary Angiography with D-Tagging, Cardiovascular Imaging, and Black Blood MRA.
spacer
 
Further Reading:
  Basics:
Latest Pulse Sequence for Displacement-encoded MR Imaging Incorporates Essential Technical Improvements for Multiphase Measurement of Intramyocardial Strain
March 2004   by radiology.rsna.org    
MRI Resources 
DICOM - Patient Information - Bioinformatics - Education pool - Crystallography - Blood Flow Imaging
 
Periodically Rotated Overlapping Parallel Lines with Enhanced ReconstructionInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(PROPELLER) The PROPELLER MRI technique reduces the sensitivity to various sources of image artifacts (e.g., motion artifact, field inhomogeneity artifact, eddy current artifact). PROPELLER can be used with gradient echo and turbo spin echo sequences in a wide range of applications to improve the image quality, for example cardiac MRI, brain MRI, and pediatric examinations.
spacer
 
Further Reading:
  Basics:
Periodically Rotated Overlapping ParallEL Lines with Enhanced Reconstruction(PROPELLER) MRI; Application to Motion Correction
1999   by cds.ismrm.org    
MR Field Notes
   by www.gehealthcare.com    
Advances in Magnetic Resonance Neuroimaging
Friday, 27 February 2009   by www.ncbi.nlm.nih.gov    
  News & More:
Patient movement during MRI: Additional points to ponder
Tuesday, 5 January 2016   by www.healthimaging.com    
New MR sequence helps radiologists more accurately evaluate abnormalities of the uterus and ovaries
Thursday, 23 April 2009   by www.eurekalert.org    
MRI Resources 
Education - Databases - MRI Technician and Technologist Career - Software - Breast MRI - Bioinformatics
 
Cerebro Spinal Fluid Pulsation ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Cerebro spinal fluid pulsation
DESCRIPTION
Ghosting
REASON
Inconsistencies in phase and amplitude
Pulsatile cerebro spinal fluid flow produces ghost artifacts that are superimposed in the image.
mri safety guidance
Image Guidance
Flow compensation should be used to reduce these artifacts. This applies an additional gradient to eliminate phase differences for both stationary and moving spins at the echo time. At TE no phase differences is measured. If flow compensation is applied and there are still flow artifacts, cardiac triggering is an additional option to reduce these artifacts.
See also Motion Artifact.
spacer

• View the DATABASE results for 'Cerebro Spinal Fluid Pulsation Artifact' (3).Open this link in a new window

MRI Resources 
Diffusion Weighted Imaging - Safety Training - Spectroscopy - Guidance - Cardiovascular Imaging - Journals
 
Phase Encoded Motion ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
Phase encoded motion, motion, phase effect
DESCRIPTION
Blurring and ghosting
REASON
Movement of the imaged object
HELP
Compensation techniques, more averages, anti spasmodic, presaturation
This artifact is caused by movements of the patient or organic processes taking place in the body of the patient. The artifact appears as bright noise, repeating densities or ghosting in the phase encoding direction.
mri safety guidance
Image Guidance
There are different solutions for reduction of phase encoded motion artifacts.
Cardiac and respiratory gating, breath holding, sedation of the patient, presaturation pulses for flow artifacts (e.g. arterial pulsation, breathing), fast imaging sequences, etc.

See also Motion Artifact, Ghosting Artifact, Motion Compensation Pulse Sequences and Artifact Reduction - Motion.
spacer

• View the DATABASE results for 'Phase Encoded Motion Artifact' (5).Open this link in a new window

MRI Resources 
Musculoskeletal and Joint MRI - Chemistry - Image Quality - Non-English - Service and Support - MRA
 
Motion ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
Motion, phase encoded motion, instability, smearing
DESCRIPTION
Blurring and ghosting
REASON
Movement of the imaged object
HELP
Compensation techniques, more averages, anti spasmodic
Patient motion is the largest physiological effect that causes artifacts, often resulting from involuntary movements (e.g. respiration, cardiac motion and blood flow, eye movements and swallowing) and minor subject movements.
Movement of the object being imaged during the sequence results in inconsistencies in phase and amplitude, which lead to blurring and ghosting. The nature of the artifact depends on the timing of the motion with respect to the acquisition. Causes of motion artifacts can also be mechanical vibrations, cryogen boiling, large iron objects moving in the fringe field (e.g. an elevator), loose connections anywhere, pulse timing variations, as well as sample motion. These artifacts appear in the phase encoding direction, independent of the direction of the motion.
mri safety guidance
Image Guidance
Motion artifacts can be flipped 90° by swapping the phase//frequency encoding directions.
The artifacts can be reduced by using breath holding, cardiac synchronization or respiratory compensation techniques: triggering, gating, retrospective triggering or phase encoding artifact reduction. Flow effects can be reduced by using gradient moment nulling of the first order of flow, gradient moment rephasing or flow compensation, depending of the MRI system.
Peristaltic motion can be reduced with the intravenous injection of an anti-spasmodic (e.g. Buscopan).
By using multiple averages, respiratory motion can be reduced in the same way that multiple averages increase the signal to noise ratio. Noticeable motion averaging is seen when four averages are obtained, six averages are often as good as respiratory compensation techniques and higher averages will continue to improve image quality.
In some cases will help a presaturation of the anatomy that was generating the motion.

See also Phase Encoded Motion Artifact.
spacer

• View the DATABASE results for 'Motion Artifact' (24).Open this link in a new window

 
Further Reading:
  Basics:
The Effects of Breathing Motion on DCE-MRI Images: Phantom Studies Simulating Respiratory Motion to Compare CAIPIRINHA-VIBE, Radial-VIBE, and Conventional VIBE
Tuesday, 7 February 2017   by www.kjronline.org    
  News & More:
Patient movement during MRI: Additional points to ponder
Tuesday, 5 January 2016   by www.healthimaging.com    
Motion-compensation of Cardiac Perfusion MRI using a Statistical Texture Ensemble(.pdf)
June 2003   by www.imm.dtu.dk    
MRI Resources 
Stent - Diffusion Weighted Imaging - Implant and Prosthesis - Functional MRI - Brain MRI - Online Books
 
previous      6 - 10 (of 12)     next
Result Pages : [1]  [2 3]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 2 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]