Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Point Resolved Spectroscopy' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Point Resolved Spectroscopy' found in 1 term [] and 2 definitions []
1 - 3 (of 3)     
Result Pages : [1]
MRI Resources 
Corporations - Service and Support - Open Directory Project - Health - MRI Technician and Technologist Career - Sequences
 
Point Resolved SpectroscopyInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(PRESS) Point resolved spectroscopy is a multi echo single shot technique to obtain spectral data. PRESS is a 90°-180°-180° (slice selective pulses) sequence. The 90° radio frequency pulse rotates the spins in the yx-plane, followed by the first 180° pulse (spin rotation in the xz-plane) and the second 180° pulse (spin rotation in the xy-plane), which gives the signal.
With the long echo times used in PRESS, there is a better visualization of metabolites with longer relaxation times. Many of the metabolites depicted by stimulated echo technique are not seen on point resolved spectroscopy, but PRESS is less susceptible to motion, diffusion, and quantum effects and has a better SNR than stimulated echo acquisition mode (STEAM).
spacer
 
• Share the entry 'Point Resolved Spectroscopy':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Point Spread Function
    • Point Scanning
    • Special Imaging
    • Spectra
    • Echo Time
 
Further Reading:
  Basics:
The Basics of MRI
   by www.cis.rit.edu    
  News & More:
MRI evaluation of fatty liver in day to day practice: Quantitative and qualitative methods
Wednesday, 3 September 2014   by www.sciencedirect.com    
MRI Resources 
Directories - Portals - Services and Supplies - Supplies - Distributors - Software
 
Stimulated Echo Acquisition Mode
 
(STEAM) A single volume spectroscopy method. The stimulated echo acquisition mode techniques is similar to Point Resolved Spectroscopy (PRESS). The STEAM sequence use a 90°-90°-90° pulse sequence instead of 90°-180°-180° pulses.
spacer

• View the DATABASE results for 'Stimulated Echo Acquisition Mode' (2).Open this link in a new window

 
Further Reading:
  News & More:
MRI evaluation of fatty liver in day to day practice: Quantitative and qualitative methods
Wednesday, 3 September 2014   by www.sciencedirect.com    
MRI Resources 
IR - MR Guided Interventions - Nerve Stimulator - Online Books - Veterinary MRI - Safety pool
 
Volume Excitation Stimulated Echoes
 
(VEST) The 'Volume Excitation Stimulated Echoes' sequence is a method similar to point resolved spectroscopy with stimulated echoes and volume selective excitation.
spacer
MRI Resources 
Corporations - Breast Implant - Safety Training - Brain MRI - Used and Refurbished MRI Equipment - Implant and Prosthesis
 
     1 - 3 (of 3)     
Result Pages : [1]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 26 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]