Flow CompensationInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.
 
Flow compensation is based on the principle of even echo rephasing and a function of specific pulse sequences, wherein the application of strategic gradient pulses can compensate for the objectionable spin phase effects of flow motion. Gradient moment nulling of the first order of flow is another adjustment for the reduction of flow artifacts.
Gradient field changes can be configured in such a way that during an echo the magnetization signal vectors for all pixels have zero phase angle independent of velocities, accelerations etc. of the measured tissue. The simplest velocity-compensated pulse sequence is the symmetrical second echo of a spin echo pulse sequence.
Strategic gradient pulses are integrated in special sequences (e.g. CRISP, Complex Rephasing Integrated with Surface Probes) and for the most sequences flow compensation is an optional parameter.
spacer
 
• Related Searches:
    • Gradient Moment Nulling
    • Cardiac Motion Artifact
    • Velocity
    • Motion Artifact
    • Gradient Motion Rephasing
 
Further Reading:
  Basics:
Motion Compensation in MR Imaging
   by ccn.ucla.edu    
Flow comp off: An easy technique to confirm CSF flow within syrinx and aqueduct
Wednesday, 2 January 2013   by medind.nic.in    
Searchterm 'Vector' was also found in the following services: 
spacer
Radiology  (4) Open this link in a new windowUltrasound  (2) Open this link in a new window
Gradient Motion Rephasing
 
(GMR) The application of strategic gradient pulses can compensate the objectionable spin phase effects of flow motion. That means the reducing of flow effects, e.g. gradient moment nulling of the first order of flow. The simplest velocity-compensated pulse sequence is the symmetrical second echo of a spin echo pulse sequence.
Gradient field changes can be configured in such a way that during an echo the magnetization signal vectors for all pixels have zero phase angle independent of velocities, accelerations etc. of the measured tissue. E.g. the adjustment to zero at the time TE of the net moments of the amplitude of the waveform of the magnetic field gradients with time. The zeroth moment is the area under the curve, the first moment is the 'center of gravity' etc. The aim is to minimize the phase shifts acquired by the transverse magnetization of excited nuclei moving along the gradients (including the effect of refocusing RF pulses), particularly for the reduction of image artifacts due to motion.
Also called Flow Compensation (FC), Motion Artifact Suppression Technique (MAST), Flow motion compression (STILL), Gradient Rephasing (GR), Shimadzu Motion Artifact Reduction Technique (SMART).
spacer

• View the DATABASE results for 'Gradient Motion Rephasing' (2).Open this link in a new window

 
Further Reading:
  Basics:
Motion Compensation in MR Imaging
   by ccn.ucla.edu    
MRI Resources 
IR - MRI Technician and Technologist Career - Claustrophobia - Abdominal Imaging - Blood Flow Imaging -
 
InversionForum -
related threads
 
A nonequilibrium state in which the macroscopic magnetization vector is oriented opposite to the magnetic field; usually produced by adiabatic fast passage or 180° RF pulses.
spacer

• View the DATABASE results for 'Inversion' (61).Open this link in a new window

Searchterm 'Vector' was also found in the following services: 
spacer
Resources  (1)  
 
Isochromat
 
A microscopic group of spins, which resonate at the same frequency. The spin precesses around a circle in the xy-plane and the net magnetization is the length, amplitude or magnitude of the magnetization vector Mxy: the quantity normally represented on a pixel-by-pixel basis in a MR image (thus amplitude or magnitude image).
spacer

• View the DATABASE results for 'Isochromat' (4).Open this link in a new window

Searchterm 'Vector' was also found in the following services: 
spacer
Radiology  (4) Open this link in a new windowUltrasound  (2) Open this link in a new window
Macroscopic Magnetic Moment
 
spacer
MRI Resources 
Equipment - Directories - Shielding - Jobs - Process Analysis - NMR
 
previous      21 - 25 (of 38)     next
Result Pages : [1]  [2 3 4 5 6 7 8]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 8 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]