Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Electronic Noise' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Electronic Noise' found in 1 term [] and 1 definition [], (+ 2 Boolean[] results
1 - 4 (of 4)     
Result Pages : [1]
MRI Resources 
Mobile MRI Rental - Pacemaker - Distributors - Crystallography - Veterinary MRI - RIS
 
Electronic Noise
 
Any electrical resistance generates thermal noise called Johnson Noise. This noise is also generated by the human body as a consequence of the ionic nature of body tissue and fluids, all active electronic devices (e.g. transistors) generate noise and noise is transmitted through the ether (radio-signals, interference from electrical equipment, extraterrestrial etc.).
spacer
 
• Share the entry 'Electronic Noise':  Facebook  Twitter  LinkedIn  
Searchterm 'Electronic Noise' was also found in the following service: 
spacer
Radiology  (3) Open this link in a new window
Preamplifier
 
A device that amplifies very low-level signals. A preamplifier is generally placed close to its signal source and has a very low noise figure as it is the principal determinant of electronic noise within the system. Preamplifiers used in NMR systems usually have a 50 ohm input impedance, and require a matching network to interface to the RF coil, although preamplifiers with high input impedance may be used with surface coils. Such devices typically use a field effect transistor (FET) as their input stage.
spacer

• View the DATABASE results for 'Preamplifier' (5).Open this link in a new window

MRI Resources 
Sequences - - Image Quality - Nerve Stimulator - Homepages - Collections
 
NoiseForum -
related threadsInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.
 
An undesirable background interference or disturbance that affects image quality.
The Noise is commonly characterized by the standard deviation of signal intensity in the image of a uniform object (phantom) in the absence of artifacts. The measured noise may depend on the particular phantom used due to variable effects on the Q of the receiver coil.
Noisy images appear when the SNR-Rate is too low - this is induced by the operator. Image artifacts and RF noise can often be caused by the presence and/or operation of a medical device in the MR environment. There are various noise sources in any electronic system, including Johnson noise, shot noise, thermal noise. Materials produce their own characteristic static magnetic field that can perturb the relationship between position and frequency essential to accurate image reconstruction.
RF noise, which often appears as static on the image, can be caused by a medical device located anywhere in the MR procedure room. RF noise is a result of excessive electromagnetic emissions from the medical device that interfere with the proper operation of the MR scanner. Since the MR procedure room is shielded from extraneous RF fields entering the room (Faraday cage), operation of electromagnetically noisy equipment outside the room does not typically affect the MR scanner.

See Signal to Noise Ratio and Radio Frequency Noise Artifact.
spacer

• View the DATABASE results for 'Noise' (86).Open this link in a new window


• View the NEWS results for 'Noise' (2).Open this link in a new window.
 
Further Reading:
  Basics:
Image Characteristics and Quality
   by www.sprawls.org    
  News & More:
Noise from Magnetic Resonance Imaging Can Have Short-Term Impact on Hearing
Thursday, 22 February 2018   by www.diagnosticimaging.com    
MRI Noise in Utero Not Harmful for Baby's Ears
Tuesday, 28 September 2010   by www.medgadget.com    
A Neural Mosaic Of Tones
Tuesday, 20 June 2006   by www.sciencedaily.com    
MRI Resources 
Developers - Sequences - Resources - MR Myelography - Stent - Equipment
 
Dynamic Range
 
The range of signal intensities that may need to be distinguished in an image or spectrum or that can be distinguished by the electronic components. If the signal dynamic range is too great, the need to keep the highest intensities from overloading the digitizer may result in the weaker features being lost in the digitization noise. This can be dealt with by using an analog to digital converter with a larger range of sensitivity or by using techniques to reduce the dynamic range, e.g. suppressing the signal from water in order to detect the signal from less abundant compounds.
spacer

• View the DATABASE results for 'Dynamic Range' (7).Open this link in a new window

Searchterm 'Electronic Noise' was also found in the following service: 
spacer
Radiology  (3) Open this link in a new window
     1 - 4 (of 4)     
Result Pages : [1]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 26 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]