Quantization Noise
 
spacer
Searchterm 'Digitization' was also found in the following services: 
spacer
Radiology  (3) Open this link in a new windowUltrasound  (3) Open this link in a new window
ReceiverForum -
related threads
 
The portion of the MRI equipment that detects and amplifies the RF signals picked up by the receiver coil. Includes a preamplifier, MR signal amplifier, and demodulator. Phase sensitive detectors (electronic device in which the output is dependent on the instantaneous difference in phase between two input signals) are used to down-convert the MR signal to audio-frequencies prior to digitization.

See also MRI Equipment.
spacer

• View the DATABASE results for 'Receiver' (45).Open this link in a new window

MRI Resources 
Abdominal Imaging - MR Myelography - Movies - Used and Refurbished MRI Equipment - Guidance - Safety pool
 
Truncation ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
DESCRIPTION
Edge ringing, syrinx-like stripe
REASON
Sharp changes in intensity (incomplete digitization of the echo)
HELP
Take more samples
A data truncation artifact may occur when the interface between high and low signal intensities is encountered in one imaging plane. The 2D-FT techniques transform the MR signal to spatial intensity image data with frequency and phase information encoding each axis in the plane of the scan. This artifact is found in both frequency and phase axes. Artifactual ripples adjacent to edges in an image or sharp features in a spectrum, caused by omission of higher frequency terms in Fourier transformation, particularly with the use of zero filling to replace unsampled higher frequencies.
Complex shapes are specified by series of sine and cosine waves of various frequencies, phase and amplitude. Some shapes are more difficult to encode than others. The most difficult shapes to represent with Fourier series of terms are waveforms with instantaneous transitions, tissue discontinuities or edges. The low-frequency components of the series describe the overall shape of the step function. Higher frequency components are needed to describe the corners if the step function more accurately. If not enough samples are taken, these areas cannot be accurately represented. The truncation of the infinite data series results in a ringing artifact because of the inability to accurately approximate this tissue discontinuity with a shorter truncated data set. Therefore, the ringing that occurs at all tissue boundaries on MR is called truncation artifact.
mri safety guidance
Image Guidance
This problem can be easily resolved by taking more samples - a higher acquisition matrix and/or a smaller FOV. See Gibbs Artifact and Gibbs Phenomenon.
spacer

• View the DATABASE results for 'Truncation Artifact' (2).Open this link in a new window

 
Further Reading:
  News & More:
Magnetic Resonance Imaging (MRI)
2003   by www.hull.ac.uk    
MRI Resources 
Mass Spectrometry - Contrast Enhanced MRI - Guidance - Journals - MRI Reimbursement - MRA
 
Wrap Around ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
A wrap around artifact (also called backfolding artifact or aliasing artifact) is produced by inadequate sampling or digitization. Wrap around artifacts result from digitizing fewer than two samples per period in a periodic function. Aliasing can occur in MR imaging whenever the scanned area extends beyond the field of view. These areas extending beyond the field of view boundaries are aliased back into the image to appear at artifactual locations.
 
Images, Movies, Sliders:
 Breast MRI Transverse T1 Pre Contrast 001  Open this link in a new window
    
 
spacer

• View the DATABASE results for 'Wrap Around Artifact' (3).Open this link in a new window

 
Further Reading:
  Basics:
Aliasing or wrap around artifacts
Thursday, 31 March 2011   by de.slideshare.net    
Searchterm 'Digitization' was also found in the following services: 
spacer
Radiology  (3) Open this link in a new windowUltrasound  (3) Open this link in a new window
previous      6 - 9 (of 9)     
Result Pages : [1]  [2]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 5 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]