Magnetic
Resonance -
Technology
Information
Portal |
Welcome to MRI Technology • |
|
|
 | Info Sheets |
| | | | | | | | | | | | | | | | | | | | | | | | |
 | Out- side |
| | | | |
|
| | | | | |  | Searchterm 'SPIR' was also found in the following services: | | | | |
|  |  |
| |
|
The resorting of data collected over several acquisitions to create an image where all phase encoding lines are acquired with the object at the same spatial location of the re spiratory cycle. | |  | | | |
|  | |  |  |  |
| |
|
Re spiratory synchronization that acquires image data at regular times independent of the re spiratory cycle, but chooses the sequence of phase encoding data acquisition so as to minimize the re spiratory motion-induced artifacts in the resulting image. For example, choosing the sequence of phase encoding such that adjacent samples in the final full data set have minimal differences in re spiratory phase will minimize the spacing of ghosting artifacts in the final image. | |  | | • View the DATABASE results for 'Respiratory Ordered Phase Encoding' (5).
| | | | |
|  | |  |  | |  |  | Searchterm 'SPIR' was also found in the following services: | | | | |
|  |  |
| |
|
The navigator technique measures with an additional quick MR prepulse the position, of e.g. the diaphragm before data collecting. Similar re spiratory conditions of the patient can be identified and used to synchronize image data acquisition so that re spiration induced image blurring is minimized by either respiratory ordered phase encoding or respiratory gating.
The prepulse sequence images a small area perpendicular to the structure, which is moving. The contrast of the interface between the diaphragm and the lung should be high to permit easy automatic detection.
After data acquisition, the position of the interface is automatically recorded and imaging data are only accepted when the position of the interface falls within a range of prespecified values.
This technique has the advantage of greater accuracy than other respiratory gating (therefore used for coronary angiography) and has no need for additional sensing MRI equipment, as the MR system itself provides it. | |  | | • View the DATABASE results for 'Navigator Technique' (5).
| | | | |
|  | |  |  |  |
| |
|
Quick Overview
Please note that there are different common names for this artifact.
NAME
Motion, phase encoded motion, instability, smearing
REASON
Movement of the imaged object
HELP
Compensation techniques, more averages, anti spasmodic
Patient motion is the largest physiological effect that causes artifacts, often resulting from involuntary movements (e.g. re spiration, cardiac motion and blood flow, eye movements and swallowing) and minor subject movements.
Movement of the object being imaged during the sequence results in inconsistencies in phase and amplitude, which lead to blurring and ghosting. The nature of the artifact depends on the timing of the motion with respect to the acquisition. Causes of motion artifacts can also be mechanical vibrations, cryogen boiling, large iron objects moving in the fringe field (e.g. an elevator), loose connections anywhere, pulse timing variations, as well as sample motion. These artifacts appear in the phase encoding direction, independent of the direction of the motion.
Image Guidance
| |  | | • View the DATABASE results for 'Motion Artifact' (24).
| | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
|  | |  |  |
|  | | |
|
| |
 | Look Ups |
| |
|
MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies. | | [last update: 2025-05-13 01:28:00] |
|
|