Passive Shimming
 
Shimming by adjusting the position of suitable pieces of ferromagnetic metal within or around the main magnet of a MRI system. Typically, the iron is in the form of small plates which are held in removable trays. The magnetic field is plotted, the position and quality of the iron required is then calculated and the appropriate 'shims' put in place. Several repetitions of this process will normally be required in order to achieve the optimum arrangement.
Passive shimming does not replace adjustable shimming, but is installed during the commissioning of a unit to improve the homogeneity of the bare magnet.
spacer
MRI Resources 
Pregnancy - Claustrophobia - MRI Accidents - Open Directory Project - NMR - Shoulder MRI
 
Periodically Rotated Overlapping Parallel Lines with Enhanced ReconstructionInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(PROPELLER) The PROPELLER MRI technique reduces the sensitivity to various sources of image artifacts (e.g., motion artifact, field inhomogeneity artifact, eddy current artifact). PROPELLER can be used with gradient echo and turbo spin echo sequences in a wide range of applications to improve the image quality, for example cardiac MRI, brain MRI, and pediatric examinations.
spacer
 
Further Reading:
  Basics:
Periodically Rotated Overlapping ParallEL Lines with Enhanced Reconstruction(PROPELLER) MRI; Application to Motion Correction
1999   by cds.ismrm.org    
MR Field Notes
   by www.gehealthcare.com    
Advances in Magnetic Resonance Neuroimaging
Friday, 27 February 2009   by www.ncbi.nlm.nih.gov    
  News & More:
Patient movement during MRI: Additional points to ponder
Tuesday, 5 January 2016   by www.healthimaging.com    
New MR sequence helps radiologists more accurately evaluate abnormalities of the uterus and ovaries
Thursday, 23 April 2009   by www.eurekalert.org    
MRI Resources 
Pacemaker - Portals - Calculation - Spectroscopy - Patient Information - Shoulder MRI
 
Picture Archiving and Communication SystemMRI Resource Directory:
 - PACS -
 
(PACS) A system used to communicate and archive medical imaging data, mostly images and associated textural data generated in a radiology department, and disseminated throughout the hospital. A PACS is usually based on the DICOM (Digital Imaging and Communications in Medicine) standard.
The main components in the PACS are:
acquisition devices where the images are acquired,
short and longer term archives for storage of digital and textural data,
a database and database management,
diagnostic and review workstations,
software to run the system,
a communication network linking the system components,
interfaces with other networks (hospital and radiological information systems).

Acquisition devices, which acquire their data in direct digital format, like a MRI system, are most easily integrated into a PACS.
Short term archives need to have rapid access, such as provided by a RAID (redundant array of independent disks), whereas long term archives need not have such rapid access and can be consigned, e.g. to optical disks or a magnetic.
High speed networks are necessary for rapid transmission of imaging data from the short term archive to the diagnostic workstations. Optical fiber, ATM (asynchronous transfer mode), fast or switched Ethernet, are examples of high speed transmission networks, whereas demographic textural data may be transmitted along conventional Ethernet.
Sophisticated software is a major element in any hospital-wide PACS. The software concepts include: preloading or prefetching of historical images pertinent to current examinations, worklists and folders to subdivide the vast mass of data acquired in a PACS in a form, which is easy and practical to access, default display protocols whereby images are automatically displayed on workstation monitors in a prearranged clinically logical order and format, and protocols radiologists can rapidly report worklists of undictated examinations, using a minimum of computer manipulation.
spacer

• View the DATABASE results for 'Picture Archiving and Communication System' (5).Open this link in a new window


• View the NEWS results for 'Picture Archiving and Communication System' (1).Open this link in a new window.
 
Further Reading:
  Basics:
Healthcare IT Yellow Pages PACS / Image Management Directory
   by www.health-infosys-dir.com    
Searchterm 'Range' was also found in the following services: 
spacer
News  (25)  Resources  (10)  Forum  (21)  
 
Pulse Length
 
Time duration of a pulse. For an RF pulse near the Larmor frequency, the longer the pulse length, the greater the angle of rotation of the macroscopic magnetization vector will be (greater than 180° can bring it back toward its original orientation). For an RF pulse of a given shape as a function of time, the longer the pulse length, the narrower the equivalent range of frequencies in the pulse will be.
spacer

• View the DATABASE results for 'Pulse Length' (2).Open this link in a new window

 
Further Reading:
  News & More:
Factors influencing flip angle mapping in MRI: RF pulse shape, slice-select gradients, off-resonance excitation, and B0 inhomogeneities.
Tuesday, 1 August 2006   by www.ncbi.nlm.nih.gov    
MRI Resources 
Mobile MRI - Health - Colonography - Safety Training - Resources - Most Wanted
 
Quadrature Detector
 
The quadrature detector is a part of the receiver that converts the high-frequency MRI signal to a lower frequency. This phase sensitive detector or demodulator detects the components of the signal in phase with a reference signal and 90° out of phase with the reference signal. The modulated signal contains i.e. the frequency range across the field of view encoded by the frequency encoding gradient. This may be performed by either analog or digital means.
spacer

• View the DATABASE results for 'Quadrature Detector' (7).Open this link in a new window

MRI Resources 
Process Analysis - Manufacturers - Intraoperative MRI - Research Labs - Collections - Examinations
 
previous      56 - 60 (of 79)     next
Result Pages : [1]  [2 3 4 5 6 7 8 9 10 11 12 13 14 15 16]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 6 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]