Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Raw Data' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Raw Data' found in 1 term [] and 22 definitions []
previous     11 - 15 (of 23)     next
Result Pages : [1]  [2 3 4 5]
Searchterm 'Raw Data' was also found in the following services: 
spacer
Resources  (4)  Forum  (3)  
 
Interleaved K-Space Coverage
 
The sequential collection of raw data from multiple excitations such that each excitation samples multiple lines or curvilinear paths in k-space.
spacer
Searchterm 'Raw Data' was also found in the following services: 
spacer
Radiology  (9) Open this link in a new windowUltrasound  (7) Open this link in a new window
K-SpaceForum -
related threads
 
The k-space is an extension of the concept of Fourier space that is well known in imaging. In MR imaging the k-space is a temporary memory of the spatial frequency information in two or three dimensions of an object; the k-space is defined by the space covered by the phase and frequency encoding data.
The relation between K-space data and image data is the Fourier Transformation. The data acquisition matrix contains raw image data before the image processing. In 2 dimensional Fourier transformation imaging, a line of data corresponds to the digitized MRI signal at a particular phase encoding level. The position in k-space is directly related to the gradient across the object being imaged. By changing the gradient over time, the k-space data are sampled in a trajectory through Fourier space at each point until it is filled.

See also Spatial Frequency and Raw Data.
spacer

• View the DATABASE results for 'K-Space' (65).Open this link in a new window

 
Further Reading:
  Basics:
K-space formulation of MRI
Tuesday, 22 March 2005   by www.ebyte.it    
The Basics of MRI
   by www.cis.rit.edu    
  News & More:
Optimal k-Space Sampling for Dynamic Contrast-Enhanced MRI with an Application to MR Renography
Thursday, 5 November 2009   by www.ncbi.nlm.nih.gov    
MRI Resources 
Coils - Safety Training - Process Analysis - Veterinary MRI - MR Myelography - Anatomy
 
Lorentzian Line Shape
 
A mathematical function used for signal fitting (filtering) of raw data. Other commonly used line shapes are the Gaussian line shape or the super-Lorentzian line shape.
spacer
Searchterm 'Raw Data' was also found in the following services: 
spacer
Resources  (4)  Forum  (3)  
 
Matrix Size
 
The number of data points collected in one, two or all three directions. Normally used for the 2D in plane sampling. The display matrix may be different from the acquisition matrix, although the latter determines the resolution. Measurement time may be saved by not acquiring raw data lines corresponding to high resolution. Not measured rows are filled with zeroes prior to the image calculation. A square image is the result of an interpolation in phase encoding direction. See also Zero Filling.
mri safety guidance
Image Guidance
The chosen matrix size effects scan time, resolution and SNR. Reduced measurement matrixes decrease the scan time and the resolution by increased SNR.
spacer

• View the DATABASE results for 'Matrix Size' (4).Open this link in a new window

Searchterm 'Raw Data' was also found in the following services: 
spacer
Radiology  (9) Open this link in a new windowUltrasound  (7) Open this link in a new window
Multi Shot Technique
 
When a multi shot technique is applied, each shot will have its own effect on the prepulse, with a scan time increase. Multiple shots allow a shorter IR delay but at the cost of increased scan time.
In multi shot technique (also called mosaic imaging), a group of samples, which are contiguous in k space are acquired in the same sequence repetition. The phase encoding steps or profiles are split into 'shots' (sub-acquisitions). The shot interval is the time between the shots. Usually kept as short as possible. Because the acquisitions are divided into different shots, each shot will have less T1 variation, thereby increasing T1 contrast. Two excitations, each requiring the data for one half of k-space, are the simplest variation of multi shot techniques (e.g. positive versus negative phase encoding). The alternative to this mosaic strategy for multi shot EPI is interleaving. In interleaved sequences, each repetition acquires every nth (n is the number of shots) line in k-space and for the complete raw data set the various repetition data are interlaced.

See also Single Shot Technique.
spacer
MRI Resources 
DICOM - Nerve Stimulator - Veterinary MRI - Cochlear Implant - PACS - Open Directory Project
 
previous      11 - 15 (of 23)     next
Result Pages : [1]  [2 3 4 5]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 3 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]