Fri. 17 Feb.12, 20:25
[Start of: 'MRI Noise' 0 Reply]

Category:
General |
MRI Noise |
How many effective the noise of mri machine on auditary canal? rn
|
| | Reply to this thread (login or register first) | |
Donna Ponticello
Sat. 13 Aug.05, 16:29
[Start of: 'Protection Against Noise' 4 Replies]

Category:
Patient Care and Comfort |
Protection Against Noise |
In the future, we will scan also newborn children at our 3T machine. I see a problem in the loudness of the MRI system because I know that babies are more sensitive to noise. Adults get earplugs and headphones and there is no problem. The headphones will not fit and I am worried about the noise pollution. Has anybody a good idea?
|
| View the whole thread | Reply to this thread (login or register first) | |
|
BILL KIMUTAI
Fri. 18 Apr.25, 02:59
[Start of: 'HI Everyone,I need help with signa profile 0.2T MRI.' 0 Reply]

Category:
General |
HI Everyone,I need help with signa profile 0.2T MRI. |
currently Im working with this machine and I have a a problem with image qulaity especially in Spine imaging.I would like to know the required temp in MRI room together with technical room.rnand also how to do away with noise.
Lumber spine.


i welcome all feedbacks through email or whatever forum
|
| | Reply to this thread (login or register first) | |
Renate S.
Sat. 6 Jun.15, 13:18
[Reply (1 of 2) to: 'is this kind of zebra artifact?' started by: 'luna xu' on Fri. 5 Jun.15]

Category:
Artifacts |
is this kind of zebra artifact? |
Looks more like a Zipper artifact. Can be caused by electronic noise. Switch off all light bulbs, remove all electronic equipment from the scanner room, make sure that the door is really closed (clean the contacts between door and jam if necessary) and try again. If the artifact consists control the penetration panel where the cables enter the room and the faraday cage, it is possible that there is a leak in the RF shield. If the artifact remains with all coils and sequences you need the service to repair the problem.
|
| View the whole thread | | |
Rena Miller
Fri. 15 Aug.14, 07:16
[Reply (1 of 2) to: 'door alarm/ MRI pauses if door opens' started by: 'Samuel Strode' on Thu. 7 Aug.14]

Category:
Devices, Scanner, Machines |
door alarm/ MRI pauses if door opens |
MRIs are enclosed by a Faraday cage to shield the MRI scanner against incoming radio waves which would disturb the sent and received signals. The reception or transmission of radio waves is heavily attenuated or blocked by the cage, which is usually integrated in the walls of the scanning room. The closed RF door completes the cage preventing RFI/EMI leakage, absorbs also much of the acoustic noise and prevents unauthorized persons to enter the room. Usually, all or the unreconstructed part of information is lost if the scan is stopped by door opening. Some MRI systems allow to pause the scan (at the keyboard, not by door opening) and to continue scanning while not loosing data.
|
| View the whole thread | |
| |