MRI Forum
'Spin Echo'
SEARCH FORUM FOR   
 
Result: Searchterm 'Spin Echo' found in 12 messages
Result Pages: 1 [2] 3 
More Results: Database  (96)  News Service  (1)  
Forum Overview
 bottom
David Brooks

Tue. 22 Dec.09,
02:35

[Start of:
'Scan equation I don't get, HELP'
2 Replies]


 
  Category: 
Basics and Physics

 
Scan equation I don't get, HELP
Here's the question:
The approximate scan time of a rapid spin echo sequence with a TR of 3500ms, a TE of 90ms, a 256x256 matrix, 1 exicitation, a 220mm FOV and an echo train length of 5 is______ minutes.

Formal they give is 3500msx1x256/5=?
a.3
b. 8
c. 11
d. 15

I get 179200. What am I missing in my studies and the math?
 View the whole threadReply to this thread
(login or register first)
George LoGuirato

Thu. 29 Oct.09,
19:40

[Reply (1 of 4) to:
'Math equation'
started by: 'crystal f'
on Sun. 18 Oct.09]


 
  Category: 
Basics and Physics

 
Math equation
Time in msec = TR * #phase_steps * NEX

Time in sec = (TR * #phase_steps * NEX) / 1000

Time in minutes = (-as above-) / 60,000

note: for FAST or TURBO spin echo divide the
#phase_steps by the ETL
 View the whole thread
Jenny Jordan

Sat. 17 Nov.07,
20:11

[Reply (2 of 4) to:
'Haste and Rare sequences'
started by: 'Elena sussi'
on Tue. 13 Nov.07]


 
  Category: 
Sequences and Imaging Parameters

 
Haste and Rare sequences
Hi Elena,

you can find the different names for the same sequences and options used by manufacturers at: http://www.mr-tip.com/serv1.php?type=cam .

The manufacturers have in principle similar sequences with small differences. So, as you mentioned you can take "sequences as HASTE" or "sequences type RARE".
However, you can lead back these sequences also to the fundamental sequence type FSE ( fast spin echo: http://www.mr-tip.com/serv1.php?type=seq&sub=12 ). Fast sequences as HASTE ( http://www.mr-tip.com/serv1.php?type=db1&dbs=haste ) are especially useful in cases of movement caused by their single shot technique: http://www.mr-tip.com/serv1.php?type=db1&dbs=Single%20Shot%20Technique .

Hope this helps
 View the whole thread
Reader Mail

Tue. 6 Mar.07,
22:48

[Start of:
'fluid PD vs T2 weighting'
1 Reply]


 
  Category: 
Sequences and Imaging Parameters

 
fluid PD vs T2 weighting
Can anyone explain me, why - considering SE sequences with the same TR - fluid in a PD weighting (short TE) should have less signal than in a T2 weighting (long TE)? eg:
Fig 10, p 1397; radiographics resident tutorial: contrast mechanisms in spin echo MR imaging; Radiographics 1994; 14; pp 1389-1404

I would expect that the signal intensity of any tissue (regardless of it Proton density of T2-relaxation time) should decrease with increasing echo time.

Thank you
 View the whole threadReply to this thread
(login or register first)
Jennifer Jackson

Fri. 21 Jul.06,
04:44

[Reply (1 of 2) to:
'Phase encoding'
started by: 'guillaume bierry'
on Fri. 7 Jul.06]


 
  Category: 
General

 
Phase encoding
Wow this is a long subject, First you have to under stand how the 3 main gradients are used. (slice select, Freq, then phase)
So you are use 1 of the gradients for your phase encoding. You need to fill the k-space which depends on you matrix. Lets assume you are using 256 (phase). In order to fill K-space you have to have 256 aquistions of the signal. Meaning for each picture (or slice) the signal must be acquire(sampled) 256 times each at a differnt amplitude to put it in a different Line of k-space.

Example- If you are performing a SE(spin echo) you have 1 TR period and 1 TE period (TE=is when the signal is sampled) so you will have to do the TR period 256 times to fill K-space with diffent amplitudes.

There is alot more that I left out. If this did not answer your question let me know
 View the whole thread

Result Pages : 1 [2] 3 
 top
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Look
      Ups






MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 31 October 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]