M Shultz
Mon. 7 Nov.05, 03:47
[Start of: 'Head Coil' 1 Reply]
Category:
Funktional MRI |
Head Coil |
I have a colleague who wants to use our 3T for an fMRI experiment and is looking for a good head coil for this. Any good suggestions?
Thanks
|
| View the whole thread | Reply to this thread (login or register first) | |
Susan Angelico
Thu. 15 Sep.05, 20:07
[Reply (1 of 3) to: '3T imaging parameters' started by: 'Nancy Dudek' on Fri. 26 Aug.05]
Category:
Sequences and Imaging Parameters |
3T imaging parameters |
At 3T the T1 time is longer (so the TR should also be longer) and the T2 time is shorter (TE can be shorter). However, when we got a 3T machine we tested different TR and TE's and found out that in practice the useful range don't has a big difference from that of a 1.5T MR system.
The contrast looks a bit different - that's normal.
In addition, ask the manufacturer.
There are a lot of other parameters that affect the contrast in 3T.
Hope this will help
|
| View the whole thread | | |
Donna Ponticello
Sat. 13 Aug.05, 16:29
[Start of: 'Protection Against Noise' 4 Replies]
Category:
Patient Care and Comfort |
Protection Against Noise |
In the future, we will scan also newborn children at our 3T machine. I see a problem in the loudness of the MRI system because I know that babies are more sensitive to noise. Adults get earplugs and headphones and there is no problem. The headphones will not fit and I am worried about the noise pollution. Has anybody a good idea?
|
| View the whole thread | Reply to this thread (login or register first) | |
Sue Johanson
Wed. 27 Apr.05, 03:36
[Reply (1 of 4) to: 'Contrast MRA COW.' started by: 'Marilia White' on Thu. 3 Mar.05]
Category:
Contrast Agents |
Contrast MRA COW. |
Hi,
seems crazy, you should have many undesirable veins! We run a 3D TOF MRA of the circle of willis without contrast with best results (1.5T and 3T).
Contrast enhanced 3D MRA of the carotids are done with machine injection (never with a butterfly!), automated bolus detection and a flow of 3ml/sec and 25 ml Gad. If no bolus detection program is available, look for the delay with a test bolus of 2ml. With a high resolution scan, both carotids and circle of willis are assessable. The use of a set like Smartset for the injection is helpful if no power injector is available.
|
| View the whole thread | | |
Brad Johnson
Wed. 26 Jan.05, 03:57
[Reply (1 of 2) to: '3T Pulsoxymeter' started by: 'Lynn Dunkan' on Wed. 3 Nov.04]
Category:
Equipment |
3T Pulsoxymeter |
Hi Lynn,
we use the pulse oximeter Nonin 8600 for our 1.5 T scanner as well as for the 3T system. I'm not sure whether it is tested for 3 T. At the 3 T system, we run the cable through filter boxes in the shielded scanner room, so there is no safety problem and we do not have artifacts.
It works well also for children.
|