BeamformingMRI Resource Directory:<br> - Artifacts -
 
The wider the ultrasound beam, the more severe the problem with volume averaging and the beam-width artifact, to avoid this, the ultrasound beam can be shaped with lenses.
Different possibilities to focus the beam:
point Mechanical focusing is performed by placing an acoustic lens on the surface of the transducer or using a transducer with a concave face.
point Electronic focusing uses multiple phased array (annular or linear) elements, sequentially fired to focus the beam.
Conventional multi-element transducers are electronically focused in order to minimize beam width. This transducer type can be focused electronically only along the long axis of the probe where there are multiple elements, along the short axis (elevation axis) are conventional transducers only one element wide. Electronic focusing in any axis requires multiple transducer elements arrayed along that axis. Short axis focusing of conventional multi-element transducers requires an acoustic lens which has a fixed focal length.
For operation at frequencies at or even above 10 MHz, quantization noise reduces contrast resolution. Digital beamforming gives better control over time delay quantization errors. In digital beamformers the delay accuracy is improved, thus allowing higher frequency operation. In analog beamformers, delay accuracy is in the order of 20 ns.
Phased beamformers are suitable to handle linear phased arrays and are used for sector formats such as required in cardiography to improve image quality. Beamforming in ultrasound instruments for medical imaging uses analog delay lines. The signal from each individual element is delayed in order to steer the beam in the desired direction and focuses the beam.
The receive beamformer tracks the depth and focuses the receive beam as the depth increases for each transmitted pulse. The receive aperture increase with depth. The lateral resolution is constant with depth, and decreases the sensitivity to aberrations in the imaged tissue. A requirement for dynamic control of the used elements is given. Since often a weighting function (apodization) is used for side lobe reduction, the element weights also have to be dynamically updated with depth.
See also Huygens Principle.
Radiology-tip.comBeam
spacer
• Related Searches:
    • Beam Steering
    • Beam Width Artifact
    • Partial Volume Artifact
    • Echo
    • Far Field

 Further Reading:
  Basics:
Ultrasound beamforming and image formation(.pdf)Open this link in a new window
2007   by dukemil.bme.duke.edu    
  News & More:
Ultrasonic Testing Using Phased ArraysOpen this link in a new window
   by www.ndt.net    
US Resources  
Patient Information - Universities - Image Quality - Ultrasound Gel - Manufacturers - Liver
 
128 BW (SONOACE 6000II)InfoSheet: - Devices -
Intro, 
TypesMRI Resource Directory:<br> - Devices Machines Scanners Systems -
 
www.medisonusa.com/main/sonoace6000.html

From Medison Co.,Ltd.;
'The Ultimate Black & White Ultrasound System
Redefining the standard of excellence in digital B/W ultrasound
Medison, the innovator in medical science, proudly introduces the ultimate evolution in black-and-white ultrasound system - the 128 BW. With its revolutionary responsive imaging technology, powered by Medison's harmonic imaging and digital 128 channel beamforming, 128 BW produces crystal-clear images in fraction of time. Designed and committed for excellent performance, 128 BW guarantees its scientific breakthrough results.'

Specifications for this system will be available soon.

spacer

US Resources  
History of UltraSound - Image Libraries - Renal - Rental - Obstetric - Online Books
 
HD11InfoSheet: - Devices -
Intro, 
TypesMRI Resource Directory:<br> - Devices Machines Scanners Systems -
 
www.medical.philips.com/main/products/ultrasound/general/hd11/features/

From Philips Medical Systems;
'Clinicians are demanding smaller, higher performing systems specifically designed to meet their clinical and operational challenges. The new Philips HD11 system provides an uncompromising platform, plus advanced options in a highly mobile and easy-to-use system.'

Device Information and Specification
APPLICATIONS Abdominal, cardiac (also for adults with TEE), musculoskeletal (also pediatric), OB/GYN, prostate, small parts, transcranial, vascular
CONFIGURATION 17" high resolution non-interlaced flat CRT, 4 active probe ports
RANGE OF PROBE TYPE Multi-frequency, 4D, convex - micro convex, phased array, linear, specialty
TRANSDUCERS LOGIQ 7 Probes
IMAGING MODES B-mode, M-mode, coded harmonic imaging, color flow mode (CFM), power Doppler imaging (PDI), color Doppler, pulsed wave Doppler, tissue harmonic imaging
IMAGING OPTIONS CrossXBeam spatial compounding, coded ultrasound acquisition),speckle reduction imaging (SRI), TruScan technology store raw data, CINE review with 4 speed types
OPTIONAL PACKAGE Transesophageal scanning, stress echo, tissue velocity imaging (TVI), tissue velocity Doppler (TVD), contrast harmonic imaging
STORAGE, CONNECTIVITY, OS Patient and image archive, HDD, DICOM 3.0, CD/DVD, MOD, Windows-based
DATA PROCESSING Digital beamformer with 1024 system processing channel technology
H*W*D m (inch.) 1.62 * 0.61 * 0.99 (64 * 24 * 39)
WEIGHT 246 kg (498 lbs.)
POWER CONSUMPTION less than 1.5 KVA
spacer

• View the news results for 'HD11' (3).



 Further Reading:
  Basics:
HD11 XE Ultrasound SystemOpen this link in a new window
   by www.medical.philips.com    
Searchterm 'Beamforming' was also found in the following service: 
spacer
News  (6)  
 
HD3MRI Resource Directory:<br> - Devices Machines Scanners Systems -
 
www.medical.philips.com/main/products/ultrasound/general/hd11/features/

From Philips Medical Systems;
Introduced in June 2005, 'one of the less expensive and more dedicated' ultrasound systems.


Device Information and Specification
APPLICATIONS Abdominal, cardiac, OB/GYN, pediatric, prostate, small parts, urology, vascular
CONFIGURATION LCD monitor
RANGE OF PROBE TYPE Broadband, convex, linear,
digital beamformer and focal tuning
IMAGING MODES 2-D, M-mode, Doppler, color power angio (CPA), tissue harmonic imaging, color flow mapping
IMAGING OPTIONS Broadband digital beamformer, cineloop review, extensive measurement tools and analysis packages
OPTIONAL PACKAGE DICOM, etc.
STORAGE, CONNECTIVITY, OS HDD, CD, USB, optional MOD and DICOM 3.0
DATA PROCESSING 256-digitally processed channels
H*W*D m (inch.) - * - * - (58 * 20 * 32)
WEIGHT - kg (135 lbs.)
spacer

US Resources  
Hospitals - Modes - History of UltraSound - RIS - Distributors - Software
 
History of UltrasoundMRI Resource Directory:<br> - History of UltraSound -
 
point In 1880 the Curie brothers discovered the piezoelectric effect in quartz. Converse piezoelectricity was mathematically deduced from fundamental thermodynamic principles by Lippmann in 1881.
point In 1917, Paul Langevin (France) and his coworkers developed an underwater sonar system (called hydrophone) that uses the piezoelectric effect to detect submarines through echo location.
point In 1935, the first RADAR system was produced by the British physicist Robert Watson-Wat. Also about 1935, developments began with the objective to use ultrasonic power therapeutically, utilizing its heating and disruptive effects on living tissues. In 1936, Siemens markets the first ultrasonic therapeutic machine, the Sonostat.
point Shortly after the World War II, researchers began to explore medical diagnostic capabilities of ultrasound. Karl Theo Dussik (Austria) attempted to locate the cerebral ventricles by measuring the transmission of ultrasound beam through the skull. Other researchers try to use ultrasound to detect gallstones, breast masses, and tumors. These first investigations were performed with A-mode.
point Shortly after the World War II, researchers in Europe, the United States and Japan began to explore medical diagnostic capabilities of ultrasound. Karl Theo Dussik (Austria) attempted to locate the cerebral ventricles by measuring the transmission of ultrasound beam through the skull. Other researchers, e.g. George Ludwig (United States) tried to use ultrasound to detect gallstones, breast masses, and tumors. This first experimentally investigations were performed with A-mode. Ultrasound pioneers contributed innovations and important discoveries, for example the velocity of sound transmission in animal soft tissues with a mean value of 1540 m/sec (still in use today), and determined values of the optimal scanning frequency of the ultrasound transducer.
point In the early 50`s the first B-mode images were obtained. Images were static, without gray-scale information in simple black and white and compound technique. Carl Hellmuth Hertz and Inge Edler (Sweden) made in 1953 the first scan of heart activity. Ian Donald and Colleagues (Scotland) were specialized on obstetric and gynecologic ultrasound research. By continuous development it was possible to study pregnancy and diagnose possible complications.
point After about 1960 two-dimensional compound procedures were developed. The applications in obstetric and gynecologic ultrasound boomed worldwide from the mid 60’s with both, A-scan and B-scan equipment. In the late 60’s B-mode ultrasonography replaced A-mode in wide parts.
point In the 70’s gray scale imaging became available and with progress of computer technique ultrasonic imaging gets better and faster.
point After continuous work, in the 80’s fast realtime B-mode gray-scale imaging was developed. Electronic focusing and duplex flow measurements became popular. A wider range of applications were possible.
point In the 90’s, high resolution scanners with digital beamforming, high transducer frequencies, multi-channel focus and broad-band transducer technology became state of the art. Optimized tissue contrast and improved diagnostic accuracy lead to an important role in breast imaging and cancer detection. Color Doppler and Duplex became available and sensitivity for low flow was continuously improved.
point Actually, machines with advanced ultrasound system performance are equipped with realtime compound imaging, tissue harmonic imaging, contrast harmonic imaging, vascular assessment, matrix array transducers, pulse inversion imaging, 3D and 4D ultrasound with panoramic view.
read more

Radiology-tip.comDiagnostic Imaging
spacer
Radiology-tip.comMRI History
spacer

 Further Reading:
  News & More:
Physics Tutorial: Ultrasound PhysicsOpen this link in a new window
   by www.physics247.com    
A-Mode Area RatioOpen this link in a new window
   by www.wildultrasound.com    
US Resources  
UltraSound Reimbursement - Artifacts - Image Libraries - Renal - Examinations -
 
Related Searches:
 • Echo
 • Near Field
 • Partial Volume Artifact
 • Beam Steering
 • Far Field
SEARCH FOR   
 
  12345ABCDEFGHIJKLMNOPQRSTUVWZ
     1 - 5 (of 20)     next
Result Pages : [1]  [2 3 4]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

US-TIP    
Community   
User
Pass
Forgot your UserID/Password ?  


Look
      Ups



UltraSound - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • US-TIP • The-Medical-Market
Copyright © 2006 - 2016 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertising
 [last update: 2015-03-04 09:17:02]