MRI Sequences
 
 
 
ball_redDriven Equilibrium 
In fast imaging sequences driven equilibrium sensitizes the sequence to variations in T2. This MRI technique turns transverse magnetization Mxy to the longitudinal axis using a pulse rather than waiting for T1 relaxation.
The first two pulses form a spin echo and, at the peak of the echo, a second 90° pulse returns the magnetization to the z-axis in preparation for a fresh sequence. In the absence of T2 relaxation, then all the magnetization can be returned to the z-axis. Otherwise, T2 signal loss during the sequence will reduce the final z-magnetization.
The advantage of this sequence type is, that both longitudinal and also transverse magnetization are back to equilibrium in a shorter amount of time. Therefore, contrast and signal can be increased while using a shorter TR. This pulse type can be applied to other sequences like FSE, GE or IR.

Sequences with driven equilibrium:
Driven Equilibrium Fast Gradient Recalled acquisition in the steady state - DE FGR,
Driven Equilibrium Fourier Transformation - DEFT,
Driven Equilibrium magnetization preparation - DE prep,
Driven Equilibrium Fast Spin Echo - DE FSE.

• View the DATABASE results for 'Driven Equilibrium' (8).Open this link in a new window

 
Further Reading:
  Basics:
3D Turbo Spin-Echo Sequence with Motion-Sensitized Driven-Equilibrium Preparation for Detection of Brain Metastases on 3T MR Imaging
Saturday, 3 December 2011   by www.ajnr.org    
  News & More:
Advances in high-field MR imaging of the spine
Wednesday, 5 August 2009   by www.appliedradiology.com    
Comparison of New Methods for Magnetic Resonance Imaging of Articular Cartilage(.pdf)
2002
Driven Equilibrium Fast Gradient Recalled Acquisition in the Steady State 
(DE FGR) A gradient echo sequence using a pulse, which sensitizes the sequence to variations in T2, rather than waiting for T1 relaxation.
See Driven Equilibrium, Gradient Recalled Echo Sequence and Steady State Free Precession.
Driven Equilibrium Fast Spin Echo 
(DE FSE) A fast spin echo sequence with application of a pulse, which sensitizes the sequence to variations in T2.
 
Further Reading:
  News & More:
Metal Artefact Reduction
Thursday, 9 June 2011   by www.revisemri.com    
Driven Equilibrium Fourier Transformation 
(DEFT) This sequence enhances fluid signal by using a 'tip-up' pulse following a spin echo train.
See Driven Equilibrium and Fourier Transformation Imaging.
Driven Equilibrium Magnetization Preparation 
(DE prep) See Driven Equilibrium.
  Refocused Gradient Echo Sequence top
In those days spirits were brave, the stakes were high, men were real men, women were real women and small furry creatures from Alpha Centauri were real small furry creatures from Alpha Centauri.
- Douglas Adams
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Look
      Ups






MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 20 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]