Spin Echo Multi SliceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(SEMS) This pulse sequence is composed of a 90° RF pulse followed by a 180° refocusing pulse. Both RF pulses are applied in the presence of a slice select gradient.
By choosing of different TR and TE, depending on the T1 and T2 values of the tissues, proton density, T1 weighted and T2 weighted images can be acquired.
The inversion recovery option enlarge the RF pulses with a 180° inverting pulse, applied a TI time before the beginning of the pulse sequence in order to manipulate image contrast.
See also Spin Echo Sequence.
spacer
 
Further Reading:
  Basics:
Fast Spin Echo(.pdf)
Tuesday, 24 January 2006   by www.81bones.net    
MRI Resources 
Non-English - Implant and Prosthesis pool - Fluorescence - MRI Technician and Technologist Career - PACS - Quality Advice
 
Spin Echo SequenceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
Spin Echo Timing Diagram (SE) The most common pulse sequence used in MR imaging is based of the detection of a spin or Hahn echo. It uses 90° radio frequency pulses to excite the magnetization and one or more 180° pulses to refocus the spins to generate signal echoes named spin echoes (SE).
In the pulse sequence timing diagram, the simplest form of a spin echo sequence is illustrated.
The 90° excitation pulse rotates the longitudinal magnetization (Mz) into the xy-plane and the dephasing of the transverse magnetization (Mxy) starts.
The following application of a 180° refocusing pulse (rotates the magnetization in the x-plane) generates signal echoes. The purpose of the 180° pulse is to rephase the spins, causing them to regain coherence and thereby to recover transverse magnetization, producing a spin echo.
The recovery of the z-magnetization occurs with the T1 relaxation time and typically at a much slower rate than the T2-decay, because in general T1 is greater than T2 for living tissues and is in the range of 100-2000 ms.
The SE pulse sequence was devised in the early days of NMR days by Carr and Purcell and exists now in many forms: the multi echo pulse sequence using single or multislice acquisition, the fast spin echo (FSE/TSE) pulse sequence, echo planar imaging (EPI) pulse sequence and the gradient and spin echo (GRASE) pulse sequence;; all are basically spin echo sequences.
In the simplest form of SE imaging, the pulse sequence has to be repeated as many times as the image has lines.
Contrast values:
PD weighted: Short TE (20 ms) and long TR.
T1 weighted: Short TE (10-20 ms) and short TR (300-600 ms)
T2 weighted: Long TE (greater than 60 ms) and long TR (greater than 1600 ms)
With spin echo imaging no T2* occurs, caused by the 180° refocusing pulse. For this reason, spin echo sequences are more robust against e.g., susceptibility artifacts than gradient echo sequences.

See also Pulse Sequence Timing Diagram to find a description of the components.
 
Images, Movies, Sliders:
 Shoulder Coronal T1 SE  Open this link in a new window
    

Courtesy of  Robert R. Edelman
 Shoulder Axial T1 SE  Open this link in a new window
 MRI Orbita T1  Open this link in a new window
    
 
spacer

• View the DATABASE results for 'Spin Echo Sequence' (24).Open this link in a new window

 
Further Reading:
  Basics:
Fast Spin Echo(.pdf)
Tuesday, 24 January 2006   by www.81bones.net    
Magnetic resonance imaging
   by www.scholarpedia.org    
FUNDAMENTALS OF MRI: Part I
   by www.e-radiography.net    
  News & More:
New MR sequence helps radiologists more accurately evaluate abnormalities of the uterus and ovaries
Thursday, 23 April 2009   by www.eurekalert.org    
MRI techniques improve pulmonary embolism detection
Monday, 19 March 2012   by medicalxpress.com    
MRI Resources 
Implant and Prosthesis - Equipment - MRA - Directories - Musculoskeletal and Joint MRI - Open Directory Project
 
Spine MRIMRI Resource Directory:
 - Spine MRI -
 
Magnetic resonance imaging (MRI) of the spine is a noninvasive procedure to evaluate different types of tissue, including the spinal cord, vertebral disks and spaces between the vertebrae through which the nerves travel, as well as distinguish healthy tissue from diseased tissue.
The cervical, thoracic and lumbar spine MRI should be scanned in individual sections. The scan protocol parameter like e.g. the field of view (FOV), slice thickness and matrix are usually different for cervical, thoracic and lumbar spine MRI, but the method is similar. The standard views in the basic spinal MRI scan to create detailed slices (cross sections) are sagittal T1 weighted and T2 weighted images over the whole body part, and transverse (e.g. multi angle oblique) over the region of interest with different pulse sequences according to the result of the sagittal slices. Additional views or different types of pulse sequences like fat suppression, fluid attenuation inversion recovery (FLAIR) or diffusion weighted imaging are created dependent on the indication.

Indications:
•
Neurological deficit, evidence of radiculopathy, cauda equina compression
•
Primary tumors or drop metastases
•
Infection/inflammatory disease, multiple sclerosis
•
Postoperative evaluation of lumbar spine: disk vs. scar
•
Evaluation of syrinx
•
Localized back pain with no radiculopathy (leg pain)

Contrast enhanced MRI techniques delineate infections vs. malignancies, show a syrinx cavity and support to differentiate the postoperative conditions. After surgery for disk disease, significant fibrosis can occur in the spine. This scarring can mimic residual disk herniation. Magnetic resonance myelography evaluates spinal stenosis and various intervertebral discs can be imaged with multi angle oblique techniques. Cine series can be used to show true range of motion studies of parts of the spine. Advanced open MRI devices are developed to perform positional scans in the position of pain or symptom (e.g. Upright™ MRI formerly Stand-Up MRI).
 
Images, Movies, Sliders:
 Anatomic Imaging of the Lumbar Spine  Open this link in a new window
      

Courtesy of  Robert R. Edelman

 
spacer

• View the DATABASE results for 'Spine MRI' (11).Open this link in a new window


• View the NEWS results for 'Spine MRI' (4).Open this link in a new window.
 
Further Reading:
  Basics:
Newer Sequences for Spinal MR Imaging: Smorgasbord or Succotash of Acronyms?
   by www.ajnr.org    
Cutting Edge Imaging of THE Spine
February 2007   by www.pubmedcentral.nih.gov    
Landmark Independent Study by UCLA School of Medicine Reports Comparison of Dynamic™ Upright® MRI With Static Upright MRI in More Than 1,000 Patients (1,302):
Thursday, 15 November 2007   by www.fonar.com    
  News & More:
Recommendations for MRI Assessment in Managing Axial Spondyloarthritis
Wednesday, 8 January 2020   by www.rheumatologyadvisor.com    
MRI Of The Spine Identifies Smoldering Myeloma Patients At High Risk Of Progressing To Multiple Myeloma
Tuesday, 26 August 2014   by www.myelomabeacon.com    
Intensive training of young tennis players causes spinal damage
Wednesday, 18 July 2007   by www.eurekalert.org    
Searchterm 'T1 Weighted' was also found in the following services: 
spacer
News  (2)  Resources  (1)  Forum  (4)  
 
T1W High Resolution Isotropic Volume ExaminationInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(THRIVE) THRIVE is an optimized fast T1 weighted 3 dimensional imaging technique combining sensitivity encoding, large volume coverage and uniform fat suppression. THRIVE improves for example dynamic liver, small bowel, breast, prostate and pancreas MRI, providing isotropic images with high resolution in short breath-hold times.
spacer

• View the DATABASE results for 'T1W High Resolution Isotropic Volume Examination' (3).Open this link in a new window

MRI Resources 
Knee MRI - Breast Implant - MRCP - Contrast Enhanced MRI - RIS - MRI Reimbursement
 
Volumetric Imaging
 
Volumetric imaging is a 3D technique where all the MRI signals are collected from the entire tissue sample and imaged as a whole entity, therefore providing a high signal to noise ratio. The acquisition of isotropic voxels or thin slices with high spatial resolution allows to create multiplanar reconstructions in all planes; a compensation for the usually longer scan time. The acquisition time can be reduced by parallel imaging technique.
New T2 weighted variants of 3D sequences (FSE-XETA, T2-SPACE, VISTA) have been introduced that differ from conventional FSE sequences. An echo train containing up to 200 echoes obtained at a minimum echo spacing allows very fast acquisition. A flip angle modulation (flip angle sweep - FAS) during the FSE readout carries magnetization as long as possible to avoid blurring and provide optimal signal at the effective echo time. This type of imaging is well suited for brain and spine MRI procedures.
Newer T1 weighted variants include Liver Acquisition with Volume Acquisition (LAVA) and T1W High Resolution Isotropic Volume Examination (THRIVE), which have advantages for dynamic breath hold imaging in liver and abdominal examinations.

See also Volume Imaging, 3 Dimensional Imaging.
 
Images, Movies, Sliders:
 Brain MRI Sagittal T1 001  Open this link in a new window
    
 Circle of Willis, Time of Flight, MIP  Open this link in a new window
    
SlidersSliders Overview

 MRI of the Skull Base  Open this link in a new window
    
SlidersSliders Overview

 
spacer

• View the DATABASE results for 'Volumetric Imaging' (4).Open this link in a new window


• View the NEWS results for 'Volumetric Imaging' (1).Open this link in a new window.
 
Further Reading:
  Basics:
Cutting Edge Imaging of THE Spine
February 2007   by www.pubmedcentral.nih.gov    
3-D VOLUMETRIC IMAGING FOR STEREOTACTIC LESIONAL AND DEEP BRAIN STIMULATION SURGERY
MRI Resources 
Diffusion Weighted Imaging - Lung Imaging - Breast MRI - Brain MRI - Mass Spectrometry - Devices
 
previous      51 - 55 (of 56)     next
Result Pages : [1]  [2 3 4 5 6 7 8 9 10 11 12]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 25 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]