Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Spin Echo Sequence' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Spin Echo Sequence' found in 2 terms [] and 27 definitions []
previous     21 - 25 (of 29)     next
Result Pages : [1]  [2 3 4 5 6]
Searchterm 'Spin Echo Sequence' was also found in the following service: 
spacer
Forum  (1)  
 
Radio Frequency Spin Echo
 
A spin echo produced by a RF pulse.

See also Spin Echo Sequence.
spacer
MRI Resources 
Mobile MRI Rental - MRI Centers - Resources - Sequences - MRCP - Cardiovascular Imaging
 
Refocusing
 
A rephasing or refocusing occurs when spins return to the same starting phase they had directly after the initial exciting RF pulse. E.g. spin echo sequences use 180° pulses to refocus the spins to generate signal echoes and gradient echo sequences use a refocusing gradient to maximize remaining transverse magnetization.
spacer

• View the DATABASE results for 'Refocusing' (22).Open this link in a new window

 
Further Reading:
  Basics:
Clinical evaluation of a speed optimized T2 weighted fast spin echo sequence at 3.0 T using variable flip angle refocusing, half-Fourier acquisition and parallel imaging
Wednesday, 25 October 2006
MRI Resources 
Online Books - Guidance - MRCP - Blood Flow Imaging - Crystallography - Bioinformatics
 
Relaxation Effect
 
The relaxation effect is the transition of an atom or molecule from a higher energy level to a lower one. The return of the excited proton from the high energy to the low energy level is associated with the loss of energy to the surrounding tissue. The T1 and T2 relaxation times define the way that the protons return to their resting levels after the initial radio frequency (RF) pulse. The T1 and T2 relaxation rates have an effect of the signal to noise ratio (SNR) of MR images.
The relaxation process is a result of both T1 and T2, and can be controlled by the dependency of one of the two biological parameters T1 and T2 in the recorded signal. A T1 weighted spin echo sequence is based on a short repetition time (TR) and a change of it will affect the acquisition time and the T1 weighting of the image. Increased TR results in improved SNR caused by longer recovering time for the longitudinal magnetization. Increased TE improves the T2 weighting, combined with a long TR (of several T1 times) to minimize the T1 effect.
spacer

• View the DATABASE results for 'Relaxation Effect' (2).Open this link in a new window

 
Further Reading:
  News & More:
MRI's inside story
Thursday, 4 December 2003   by www.economist.com    
Searchterm 'Spin Echo Sequence' was also found in the following service: 
spacer
Forum  (1)  
 
Rephasing
 
The process of returning out of phase magnetic moments back into phase coherence. Caused either by rapidly reversing a magnetic gradient (Field Echo) or by applying a 180° RF pulse (Spin Echo). In the spin echo pulse sequence this action effectively cancels out the spurious T2* information from the signal.

See also Spin Echo Sequence and Gradient Echo Sequence.
spacer

• View the DATABASE results for 'Rephasing' (21).Open this link in a new window

MRI Resources 
Pacemaker - Used and Refurbished MRI Equipment - Open Directory Project - Veterinary MRI - Portals - Shoulder MRI
 
Saturation
 
In MR, saturation is a nonequilibrium state with no net magnetization. The same amount of nuclear spins is aligned against and with the magnetic field. Saturation methods like FatSat, SPIR etc., work with a frequency selective saturation pulse for a specific chemical shift applied before the actual sequence starts. This saturation pulse adjusts the magnetization from tissue components to zero. The hydrogen nuclei of fat and water resonate at different frequencies, which makes it possible to excite just the fat with repeatedly applying RF pulses at the Larmor frequency with interpulse times compared to T1. The resulting signal is then destroyed with a gradient pulse (Spoiler Gradient Pulse). Fat is the chemical compound to be saturated at a fat saturation sequence. When the actual sequence follows, (e.g., a spin echo sequence) the unwanted suppressed component will not resonate.

See also Saturation Recovery.
spacer

• View the DATABASE results for 'Saturation' (54).Open this link in a new window

MRI Resources 
Stimulator pool - Blood Flow Imaging - MRI Technician and Technologist Schools - MRI Technician and Technologist Career - Resources - Non-English
 
previous      21 - 25 (of 29)     next
Result Pages : [1]  [2 3 4 5 6]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 20 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]