Spectra
 
Plural of spectrum. Spectra are components of the MR signal according to frequencies. Spectroscopy uses NMR spectral components to identify the metabolites, e.g. of the brain.
spacer
 
• Share the entry 'Spectra':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Functional Brain MR Spectroscopy
    • Spectrometer
    • Echo Offset
    • Spectrum
    • Spectral Presaturation Inversion Recovery
Searchterm 'Spectra' was also found in the following service: 
spacer
Ultrasound  (29) Open this link in a new window
Spectral Map
 
The mapping of a CSI spectral matrix to an anatomical image. It shows the regional changes in metabolites as superimposed contours.
spacer
MRI Resources 
Manufacturers - Liver Imaging - Process Analysis - Calculation - Shoulder MRI - Veterinary MRI
 
Spectral Line
 
A spectral line is a particular distinct frequency or narrow band of a frequency set. The resonance of this frequency occurs corresponding to a particular chemical shift. Theoretically, the frequency of a pure sine wave displays sharp spectral lines at the point of Larmor frequency. In reality, the spectral lines spread into a blurred peak, caused by field inhomogeneities and spin-spin effect.
spacer

• View the DATABASE results for 'Spectral Line' (21).Open this link in a new window

Searchterm 'Spectra' was also found in the following services: 
spacer
News  (6)  Resources  (28)  
 
Spectral Width
 
The overall width in hertz needed to observe a particular NMR spectrum. This width is generally set using the Nyquist limit; namely, that the temporal sampling rate must be equal to twice the maximum spread in frequencies.
spacer
Searchterm 'Spectra' was also found in the following service: 
spacer
Ultrasound  (29) Open this link in a new window
MAGNETOM Spectra
 
www.healthcare.siemens.com/magnetic-resonance-imaging/3t-mri-scanner/magnetom-spectra From Siemens Medical Systems;
Received FDA clearance in 2012.
The MAGNETOM Spectra is a cost-optimized high field MRI system with Tim 4G and Dot technologies. The system consumes less energy compared to other 3 Tesla scanners. The magnet-cooling helium is contained in a closed loop, which prevents the gas from escaping and reduces the need for refills. TimTX includes innovative techniques in the radio frequency excitation hardware as well as new application and processing features enabling uniform RF distribution in all body regions.
Device Information and Specification
CLINICAL APPLICATION
Whole Body
CONFIGURATION
Short bore
3 Tesla
Head, spine, torso/ body coil, neurovascular, neck and multi-purpose flex coils. Peripheral vascular, breast, shoulder, knee, wrist, foot//ankle, endorectal optional.
CHANNELS
24
Chemical shift imaging, single voxel spectroscopy
IMAGING TECHNIQUES
iPAT, mSENSE and GRAPPA (image, k-space), noncontrast angiography, radial motion compensation
FOV
0.5 - 45 cm
BORE DIAMETER
or W x H
At isocenter: L-R 60 cm
TABLE CAPACITY
200 kg
MAGNET WEIGHT
7200 kg
DIMENSION H*W*D (gantry included)
173 x 231 x 219 cm
5-GAUSS FRINGE FIELD
2.6 m / 4.6 m
CRYOGEN USE
Zero boil off rate, approx. 10 years
COOLING SYSTEM
Water; single cryogen, 2 stage refrigeration
up to 125 T/m/s
MAX. AMPLITUDE
33 mT/m
Passive, active; first order standard, second order optional
POWER REQUIREMENTS
380 / 400 / 420 / 440 / 460 / 480 V, 3-phase + ground; connection value with chiller 100 kvA /without chiller 60 kVA
spacer
MRI Resources 
MRI Physics - Image Quality - Mass Spectrometry - Case Studies - Knee MRI - Services and Supplies
 
     1 - 5 (of 49)     next
Result Pages : [1 2]  [3 4 5 6 7 8 9 10]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 20 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]