Entry Slice Phenomenon (Artifact)InfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Entry slice phenomenon
DESCRIPTION
Bright signals in blood vessels at the first slice
REASON
Unsaturated spins
The entry slice phenomenon arise in MRI when blood with unsaturated spins flows in the observed slice(s). These spins will emit a strong signal, because of their unsaturated status (flow related enhancement). The number of slices affected depends on the flow velocity and the slice thickness; the direction of flow determines which slices are affected. Time of Flight MRA is based on this entry slice phenomenon.

See also Flow Compensation, Flow Related Enhancement, Artifact Overview and Artifacts Reduction Index.
 
Images, Movies, Sliders:
 TOF-MRA Circle of Willis Inverted MIP  Open this link in a new window
    

 
spacer
 
Further Reading:
  News & More:
Troubleshooting the ACR MRI Accreditation Phantom Tests
   by www.aapm.org    
Searchterm 'Saturation' was also found in the following service: 
spacer
Ultrasound  (5) Open this link in a new window
Frequency Selective RF Pulse
 
A RF pulse containing energy only within a specified frequency range. Usually used for slice excitation or for selective saturation pulses.
spacer
MRI Resources 
Diffusion Weighted Imaging - Image Quality - Mass Spectrometry - Coils - Bioinformatics - Mobile MRI Rental
 
Ghosting ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Ghosting, ghost
DESCRIPTION
Displaced reduplications of image in phase encoding direction
REASON
Motion, heartbeat, respiration
HELP
Triggering, breath hold, pharmaceuticals to reduce bowel motion
Ghosting artifacts are in the most cases caused by movements (e.g., respiratory motion, bowel motion, arterial pulsations, swallowing, and heartbeat) and appear in the phase encoding direction.
mri safety guidance
Image Guidance
Ghosting artifacts can be reduced by respiratory and cardiac triggering, the use of breath holding pulse sequences, flow compensation or presaturation pulses, depending on their origin. To reduce bowel motion also pharmaceuticals, such as glucagon or scopolamine are useful. This will decrease artifacts from both peristalsis and breathing.

See also Motion Artifact, Phase Encoded Motion Artifact, Cardiac Motion Artifact, and Artifact Reduction - Motion.
spacer

• View the DATABASE results for 'Ghosting Artifact' (5).Open this link in a new window

 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
Searchterm 'Saturation' was also found in the following services: 
spacer
News  (2)  Forum  (6)  
 
Inflow Magnetic Resonance AngiographyMRI Resource Directory:
 - MRA -
 
(I MRA) In MR imaging, inflowing non-saturated fluid gives a higher signal intensity than stationary tissue. This effect makes it especially useful for imaging of flowing blood. Other factors such as susceptibility and spin saturation, can affect the signal of the blood within the vessels. Furthermore turbulence is part of normal blood flow and can decrease signal intensity.

See also Time of Flight Angiography.
 
Images, Movies, Sliders:
 TOF-MRA Circle of Willis Inverted MIP  Open this link in a new window
    

 Circle of Willis, Time of Flight, MIP  Open this link in a new window
    
SlidersSliders Overview

 
spacer
Searchterm 'Saturation' was also found in the following service: 
spacer
Ultrasound  (5) Open this link in a new window
Inversion Transfer
 
spacer
MRI Resources 
Education - Corporations - Intraoperative MRI - Chemistry - Abdominal Imaging - Movies
 
previous      31 - 35 (of 53)     next
Result Pages : [1 2 3]  [4 5 6 7 8 9 10 11]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 19 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]