Multi Slice Imaging
 
An imaging technique in which the repetition period (TR) is utilized for acquiring additional slices in other layers or planes, in differentiation for 2D techniques where every repetition period is used for single slice. The maximum number of slices of a pulse sequence depends on the repetition time.
spacer
 
Further Reading:
  Basics:
Magnetic resonance imaging
   by www.scholarpedia.org    
The Basics of MRI
   by www.cis.rit.edu    
  News & More:
Magnetic resonance-guided motorized transcranial ultrasound system for blood-brain barrier permeabilization along arbitrary trajectories in rodents
Thursday, 24 December 2015   by www.ncbi.nlm.nih.gov    
MRI Resources 
Societies - Homepages - Bioinformatics - PACS - Case Studies - MRI Training Courses
 
Multiple Slice ImagingInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
A usual variation of sequential plane imaging technique that can be used with selective excitation technique that does not affect adjacent slices. Since, in SE imaging TR longer than TE, the machine would be idling most of the time, if a single slice would be acquired; multiple slice imaging was introduced early on. Adjacent slices are imaged while waiting for relaxation of the first slice toward equilibrium, resulting in decreased image acquisition time for the set of slices. The maximum number of slices of a pulse sequence depends on the repetition time.
spacer

• View the DATABASE results for 'Multiple Slice Imaging' (4).Open this link in a new window

MRI Resources 
Pediatric and Fetal MRI - Collections - Jobs - Movies - Case Studies - Pathology
 
Partial Flip Angle
 
(PFI) A flip angle of less than 90° only partially converts the z-magnetization, leaving a fraction cos a along the longitudinal direction. A flip angle of 90° converts all the z-magnetization into xy-magnetization.
When the repetition time is shorter than T1, the use of a partial flip angle can lead to higher signal intensity. The maximum signal intensity is given by the Ernst angle. For spin echo pulse sequences using an odd number of 180° pulses, an effect similar to the use of a partial flip angle is obtained by using a flip angle greater than 90° to offset the inversion of the remaining longitudinal magnetization by the 180° pulse.
spacer
Searchterm 'Repetition Time' was also found in the following services: 
spacer
News  (1)  Forum  (1)  
 
Pulse SequenceForum -
related threadsMRI Resource Directory:
 - Sequences -
 
A pulse sequence is a preselected set of defined RF and gradient pulses, usually repeated many times during a scan, wherein the time interval between pulses and the amplitude and shape of the gradient waveforms will control NMR signal reception and affect the characteristics of the MR images. Pulse sequences are computer programs that control all hardware aspects of the MRI measurement process.
Usual to describe pulse sequences, is to list the repetition time (TR), the echo time (TE), if using inversion recovery, the inversion time (TI) with all times given in milliseconds, and in case of a gradient echo sequence, the flip angle. For example, 3000/30/1000 would indicate an inversion recovery pulse sequence with TR of 3000 msec., TE of 30 msec., and TI of 1000 msec.
Specific pulse sequence weightings are dependent on the field strength, the manufacturer and the pathology.

See also Interpulse Times.
spacer

• View the DATABASE results for 'Pulse Sequence' (96).Open this link in a new window


• View the NEWS results for 'Pulse Sequence' (1).Open this link in a new window.
 
Further Reading:
  Basics:
MYELIN-SELECTIVE MRI: PULSE SEQUENCE DESIGN AND OPTIMIZATION
   by www.imaging.robarts.ca    
Faster speed, better spatial resolution lead 3T benefits
   by sipi.usc.edu    
  News & More:
New MR sequence helps radiologists more accurately evaluate abnormalities of the uterus and ovaries
Thursday, 23 April 2009   by www.eurekalert.org    
MRI Resources 
Used and Refurbished MRI Equipment - Portals - MRI Accidents - Blood Flow Imaging - Fluorescence - Distributors
 
Relaxation Effect
 
The relaxation effect is the transition of an atom or molecule from a higher energy level to a lower one. The return of the excited proton from the high energy to the low energy level is associated with the loss of energy to the surrounding tissue. The T1 and T2 relaxation times define the way that the protons return to their resting levels after the initial radio frequency (RF) pulse. The T1 and T2 relaxation rates have an effect of the signal to noise ratio (SNR) of MR images.
The relaxation process is a result of both T1 and T2, and can be controlled by the dependency of one of the two biological parameters T1 and T2 in the recorded signal. A T1 weighted spin echo sequence is based on a short repetition time (TR) and a change of it will affect the acquisition time and the T1 weighting of the image. Increased TR results in improved SNR caused by longer recovering time for the longitudinal magnetization. Increased TE improves the T2 weighting, combined with a long TR (of several T1 times) to minimize the T1 effect.
spacer

• View the DATABASE results for 'Relaxation Effect' (2).Open this link in a new window

 
Further Reading:
  News & More:
MRI's inside story
Thursday, 4 December 2003   by www.economist.com    
MRI Resources 
Diffusion Weighted Imaging - MRI Centers - Breast MRI - Corporations - Services and Supplies - Colonography
 
previous      21 - 25 (of 33)     next
Result Pages : [1]  [2 3 4 5 6 7]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 23 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]