Array CoilInfoSheet: - Coils - 
Intro, 
Overview, 
etc.MRI Resource Directory:
 - Coils -
 
An array coil combines the advantages of smaller coils (high SNR) with those of larger coils (large measurement field). This type of RF coil is composed of separate multiple smaller coils, which can be used individually (switchable coil) or combined.
When used simultaneously, the elements can either be:
•
coupled array coils - electrically coupled to each other through common transmission lines or mutual inductance
•
isolated array coils - electrically isolated from each other with separate transmission lines and receivers and minimum effective mutual inductance, and with the signals from each transmission line processed independently or at different frequencies
•
phased array coils - multiple small coils arranged to efficiently cover a specific anatomic region and obtain high-resolution, high-SNR images of a larger volume. The data from the individual coils is integrated by special software to produce the high-resolution images.

See also the related poll result: '3rd party coils are better than the original manufacturer coils'
 
Images, Movies, Sliders:
 Anatomic Imaging of the Lumbar Spine  Open this link in a new window
      

Courtesy of  Robert R. Edelman

 
spacer
 
• Related Searches:
    • Quadrature Coil
    • Surface Coil
    • Element
    • Coil Diameter
    • Coupled Array Coils
 
Further Reading:
  News & More:
3T MRI-Compatible Small Body Array Gets FDA Nod
Monday, 15 January 2024   by www.diagnosticimaging.com    
Endorectal Coil Improves Prostate Cancer Detection
Friday, 25 November 2005   by www.sciencedaily.com    
Searchterm 'Receiver' was also found in the following services: 
spacer
Radiology  (2) Open this link in a new windowUltrasound  (10) Open this link in a new window
CHORUS 1.5T™InfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www.isoltech.co.kr/english/product/default.htm 'Next generation MRI system 1.5T CHORUS developed by ISOL Technology is optimized for both clinical diagnostic imaging and for research development.
CHORUS offers the complete range of feature oriented advanced imaging techniques- for both clinical routine and research. The compact short bore magnet, the patient friendly design and the gradient technology make the innovation to new degree of perfection in magnetic resonance.'
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Short bore
Head, C-spine, L-spine, TMJ, Knee, Shoulder, General purpose, Phased Array System: 4 digital receiver channels (Up to 12 channels)
Optional
SYNCHRONIZATION
ECG/peripheral: Optional/yes, respiratory gating
PULSE SEQUENCES
Spin Echo, Gradient Echo, Fast Spin Echo, Inversion Recovery (STIR, Fluid Attenuated Inversion Recovery), FLASH, FISP, PSIF, Turbo Flash ( MPRAGE ),TOF MR Angiography, Standard echo planar imaging package (SE-EPI, GE-EPI), Optional: Advanced P.A. Imaging Package (up to 4 ch.), Advanced echo planar imaging package, Single Shot and Diffusion Weighted EPI, IR/FLAIR EPI
IMAGING MODES
2D/3D, Travelling Sat, Multi-Slab 3D, MTC and TONE Pulse Sequence, Fat/Water Suppression, Presaturation (up to 6 bands), Flow Compensation using GMR pulse, Multi-Slice, Multi-Group Imaging
SINGLE/MULTI SLICE
Image reconstruction time (2562 ) : 0.02 s
FOV
40 cm
BORE DIAMETER
or W x H
58 cm diameter (patient)
MAGNET WEIGHT
4050 kg
H*W*D
233 x 206 x 160 cm
COOLING SYSTEM TYPE
Water-cooled coil and air-cooled amplifier
CRYOGEN USE
0.07 L/hr helium
STRENGTH
20 mT/m (Upto 27 mT/m)
5-GAUSS FRINGE FIELD
2.5 m / 3.8 m
Passive and active
spacer

• View the DATABASE results for 'CHORUS 1.5T™' (2).Open this link in a new window

MRI Resources 
Breast MRI - Coils - Contrast Enhanced MRI - MRI Technician and Technologist Jobs - Resources - Pregnancy
 
Cartesian Sampling
 
Cartesian sampling is used to refer to data collection with a fixed value of the phase encoding gradient. In 2D Fourier imaging with common Cartesian sampling of k-space sensitivity encoding by means of a receiver array enables to reduce the number of Fourier encoding steps. This is achieved by increasing the distance of sampling positions in k-space while maintaining the maximum k-values.
The Cartesian coordinates are obtained from the polar coordinates by the operations
x = r sin f
y = r cos f
using the trigonometric functions sine and cosine.
spacer

• View the DATABASE results for 'Cartesian Sampling' (4).Open this link in a new window

Searchterm 'Receiver' was also found in the following services: 
spacer
News  (2)  Resources  (1)  Forum  (5)  
 
Circularly Polarized CoilInfoSheet: - Coils - 
Intro, 
Overview, 
etc.MRI Resource Directory:
 - Coils -
 
A coil designed to excite or detect spins using two orthogonal transmit and/or receive channels. As a transmitter coil, there is a factor of 2 reductions in power required. The receiver coil has a better SNR than a linearly polarized coil.
spacer
Searchterm 'Receiver' was also found in the following services: 
spacer
Radiology  (2) Open this link in a new windowUltrasound  (10) Open this link in a new window
Data Clipping (Artifact)InfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Data clipping artifacts give images a washed-out, non-uniform appearance. If the receiver gain increase, so that the signal level is larger than the maximum ADC value, the reconstructed image brightness was scaled to compensate this. The overall intensity loss as well as the extensive signal is reconstructed outside of the object. This effect is called 'clipping' because on a plot of signal amplitude vs. time, it looks like the top and bottom of the echo has been 'clipped off' with scissors.

See Radio Frequency Overflow Artifact.
spacer
MRI Resources 
Abdominal Imaging - Liver Imaging - MRCP - Brain MRI - MRI Technician and Technologist Jobs - Distributors
 
previous      16 - 20 (of 45)     next
Result Pages : [1]  [2 3 4 5 6 7 8 9]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 28 March 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]