Hermitian Symmetry
 
The symmetry in k-space is a fundamental property of Fourier transformations. For a two-dimensional example, let g(x,y) be a complex function, i.e. the value of g at any (x,y) is a complex number. If nothing is known about the function g, data throughout all of k-space is needed to fully characterize it.
If the function g is 'real', meaning that at every (x,y) the imaginary component of g(x,y) is zero, then you only need half as much data to characterize g. The result is redundancy between the data on one half of k-space and the other. Specifically, if G(kx,ky) is the Fourier transformation of g(x,y), and g(x,y) is real, then G(kx,ky)=G*(- kx,- ky), where * indicates a complex conjugate. The data in mirrored positions in k-space, i.e. (kx,ky) versus (- kx,- ky), are conjugates of each other.

See Imaginary Numbers and Complex Conjugate.
spacer
 
Further Reading:
  Basics:
Echo-planar imaging (EPI) and functional MRI(.pdf)
1998   by www.uib.no    
Searchterm 'Real' was also found in the following services: 
spacer
Radiology  (9) Open this link in a new windowUltrasound  (52) Open this link in a new window
Apparent Diffusion CoefficientInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Diffusion Weighted Imaging -
 
(ADC) A diffusion coefficient to differentiate T2 shine through effects or artifacts from real ischemic lesions. In the human brain, water diffusion is a three-dimensional process that is not truly random because the diffusional motion of water is impeded by natural barriers. These barriers are cell membranes, myelin sheaths, white matter fiber tracts, and protein molecules.
The apparent water diffusion coefficients can be calculated by acquiring two or more images with a different gradient duration and amplitude (b-values). The contrast in the ADC map depends on the spatially distributed diffusion coefficient of the acquired tissues and does not contain T1 and T2* values.
The increased sensitivity of diffusion-weighted MRI in detecting acute ischemia is thought to be the result of the water shift intracellularly restricting motion of water protons (cytotoxic edema), whereas the conventional T2 weighted images show signal alteration mostly as a result of vasogenic edema.
The reduced ADC value also could be the result of decreased temperature in the nonperfused tissues, loss of brain pulsations leading to a decrease in apparent proton motion, increased tissue osmolality associated with ischemia, or a combination of these factors. The lower ADC measurements seen with early ischemia, have not been fully established, however, a lower apparent ADC is a sensitive indicator of early ischemic brain at a stage when ischemic tissue remains potentially salvageable.

See also Diffusion Weighted Imaging and Diffusion Tensor Tractography.
spacer

• View the DATABASE results for 'Apparent Diffusion Coefficient' (4).Open this link in a new window

 
Further Reading:
  Basics:
Implementation of Dual-Source RF Excitation in 3 T MR-Scanners Allows for Nearly Identical ADC Values Compared to 1.5 T MR Scanners in the Abdomen
Wednesday, 29 February 2012   by www.plosone.org    
EVALUATION OF HUMAN STROKE BY MR IMAGING
2000
  News & More:
The utility of texture analysis of kidney MRI for evaluating renal dysfunction with multiclass classification model
Tuesday, 30 August 2022   by www.nature.com    
Diffusion-weighted MRI in Advanced Epithelial Ovarian Cancer: Apparent Diffusion Coefficient as a Response Marker
Tuesday, 1 October 2019   by pubs.rsna.org    
EORTC study aims to qualify ADC as predictive imaging biomarker in preoperative regimens
Monday, 4 January 2016   by www.eurekalert.org    
Novel MRI Technique Could Reduce Breast Biopsies, University of Washington Study
Tuesday, 2 October 2012   by www.eurekalert.org    
Hopkins researchers use diffusion MRI technique to monitor ultrasound uterine fibroid treatment
Monday, 8 August 2005   by www.eurekalert.org    
MRI Resources 
Research Labs - Image Quality - Used and Refurbished MRI Equipment - Liver Imaging - Fluorescence - Open Directory Project
 
Audio Frequency ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Audio frequency
DESCRIPTION
Ghosting, lines or spots
REASON
Wrong modulation at audio rate, wrong audio signal
HELP
AC-line synchronization
Two types of audio-frequency problems are possible:
1. Modulation of the MR signal at an audio rate
2. Audio signal component at digitizer input
Problem 1 looks like ghosts, weak copies of the real image, displaced along the phase encoding direction. The number and intensity of the ghosts depends upon the relationship between the period of the audio modulation and the repetition time.
Problem 2 shows up as lines or spots at the appropriate points along the frequency direction. If there is no correlation between the audio period and TR, lines are generated or discrete spots occur.
mri safety guidance
Image Guidance
Both problems can be lessened by use of AC-line synchronization (line trigger).
spacer
Searchterm 'Real' was also found in the following services: 
spacer
News  (47)  Resources  (9)  Forum  (34)  
 
Beta TestMRI Resource Directory:
 - General -
 
The phase of software testing where a program or system is turned over to a select group of users outside the development organization for use in real-life situations.
spacer

• View the DATABASE results for 'Beta Test' (2).Open this link in a new window

Searchterm 'Real' was also found in the following services: 
spacer
Radiology  (9) Open this link in a new windowUltrasound  (52) Open this link in a new window
Cardiac Gating
 
This method synchronize the heartbeat with the beginning of the TR, whereat the r wave is used as the trigger. Cardiac gating times the acquisition of MR data to physiological motion in order to minimize motion artifacts. ECG gating techniques are useful whenever data acquisition is too slow to occur during a short fraction of the cardiac cycle.
Image blurring due to cardiac-induced motion occurs for imaging times of above approximately 50 ms in systole, while for imaging during diastole the critical time is of the order of 200-300 ms. The acquisition of an entire image in this time is only possible with using ultrafast MR imaging techniques. If a series of images using cardiac gating or real-time echo planar imaging EPI are acquired over the entire cardiac cycle, pixel-wise velocity and vascular flow can be obtained.
In simple cardiac gating, a single image line is acquired in each cardiac cycle. Lines for multiple images can then be acquired successively in consecutive gate intervals. By using the standard multiple slice imaging and a spin echo pulse sequence, a number of slices at different anatomical levels is obtained. The repetition time (TR) during a ECG-gated acquisition equals the RR interval, and the RR interval defines the minimum possible repetition time (TR). If longer TRs are required, multiple integers of the RR interval can be selected. When using a gradient echo pulse sequence, multiple phases of a single anatomical level or multiple slices at different anatomical levels can be acquired over the cardiac cycle.
Also called cardiac triggering.
 
Images, Movies, Sliders:
 Cardiac Infarct Short Axis Cine Overview  Open this link in a new window
    

Courtesy of  Robert R. Edelman
 Infarct 4 Chamber Cine  Open this link in a new window
    

Courtesy of  Robert R. Edelman
 
spacer

• View the DATABASE results for 'Cardiac Gating' (15).Open this link in a new window

 
Further Reading:
  Basics:
Cardiac MRI - Technical Aspects Primer
Wednesday, 7 August 2002
Electrocardiogram in an MRI Environment: Clinical Needs, Practical Considerations, Safety Implications, Technical Solutions and Future Directions
Wednesday, 25 January 2012   by cdn.intechopen.com    
Motion-compensation of Cardiac Perfusion MRI using a Statistical Texture Ensemble(.pdf)
June 2003   by www.imm.dtu.dk    
MRI Resources 
Intraoperative MRI - Services and Supplies - Absorption and Emission - Patient Information - Distributors - NMR
 
previous      11 - 15 (of 47)     next
Result Pages : [1]  [2 3 4 5 6 7 8 9 10]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Next big thing in MRI will be :
AI 
remote operator 
personalized protocols 
helium-free 
molecular MRI 
portable MRI 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 20 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]