The Mx and My data (digital MR signal) sampled and stored during data acquisition as a function of time and phase from an MRI sequence are called raw data. The horizontal axis is the frequency code, the vertical axis the phase code (of the protons).
Every point in the raw data matrix contains part of the information for the complete image. A point in the raw data matrix does not correspond to a point in the image matrix. The outer rows of the raw data matrix provide information regarding the borders and contours of the image, detailed structures, and also determine the resolution. This is also called k-space data.
spacer
 
• Share the entry 'Raw Data':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Echo Sharing
    • Interleaved K-Space Coverage
    • Sinc Interpolation
    • Interferogram
    • Matrix Size
 
Further Reading:
  Basics:
MR Image Reconstruction from Raw Data
   by dukemil.egr.duke.edu    
Searchterm 'Raw Data' was also found in the following services: 
spacer
Radiology  (9) Open this link in a new windowUltrasound  (7) Open this link in a new window
Single Shot Technique
 
In single shot techniques (used for EPI, TSE, FSE, RARE, HASTE), the entire raw data set is acquired with a single excitation pulse. The magnetization of a fully relaxed spin system is used. Each of the subsequent echoes is given a different phase encoding. For improved SNR, spatial resolution or FOV, the needed raw data are acquired over a number of sequence repetitions. Each repetition then collects a fraction of the complete raw data set. Only slightly more than a half of the raw data is acquired. The image is obtained through half Fourier reconstruction.
A single shot sequence is useful in cases where movement is to expect e.g. in abdominal Imaging or fetal MRI.

See also Half Fourier Acquisition Single Shot Turbo Spin Echo.
spacer

• View the DATABASE results for 'Single Shot Technique' (7).Open this link in a new window

 
Further Reading:
  Basics:
Clinical evaluation of a speed optimized T2 weighted fast spin echo sequence at 3.0 T using variable flip angle refocusing, half-Fourier acquisition and parallel imaging
Wednesday, 25 October 2006
MRI Resources 
Spine MRI - Cochlear Implant - Used and Refurbished MRI Equipment - Spectroscopy - Pregnancy - Open Directory Project
 
Interferogram
 
In some cases raw data are displayable. These are typically just the real part of the complex raw data, and have been scaled to emphasize the interesting details. Therefore, they cannot be successfully transformed to reproduce the displayed images, but may help to find artifacts or the reason for bad image quality.
spacer
Searchterm 'Raw Data' was also found in the following services: 
spacer
Resources  (4)  Forum  (3)  
 
DC Offset ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
This artifact looks like a pixel at the exact center, which is brighter or darker than the surroundings by the Fourier transformation of a constant raw data offset.

See Central Point Artifact.
spacer
 
Further Reading:
  Basics:
Image Artefacts
   by www.fmrib.ox.ac.uk    
Searchterm 'Raw Data' was also found in the following services: 
spacer
Radiology  (9) Open this link in a new windowUltrasound  (7) Open this link in a new window
Double Contrast SequenceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
This sequence type is a TSE counterpart to double echo sequences. To keep the echo train as short as possible, only echoes for PD and T2 weighted images, where the phase encoding gradient has a small amplitude, are measured separately. The echoes that determine resolution are used in both raw data matrices. This reduces the number of echoes required. Also the SAR drops and more slices can be acquired in the same repetition time.
spacer
 
Further Reading:
  Basics:
BASIC PRINCIPLES OF MR IMAGING
   by spinwarp.ucsd.edu    
MRI Resources 
Education - Bioinformatics - Software - Homepages - Blood Flow Imaging - Knee MRI
 
     1 - 5 (of 23)     next
Result Pages : [1]  [2 3 4 5]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 20 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]