Magnetic Shielding
 
Means to confine the region of strong magnetic field surrounding a magnet; most commonly the use of material with high permeability (passive shielding) or by employing secondary counteracting coils outside of the primary coils (active shielding). The high permeability material can be employed in the form of a yoke immediately surrounding the magnet (self-shielding) or installed in the walls of a room as full or partial room-shielding. Unlike shielding ionizing radiation, for example, magnetic shielding can only be accomplished by forcing the unavoidable magnetic return flux through more confined areas or structures, not by absorbing it.

See also Radio Frequency Shielding Radio Frequency Shielding, and Faraday cage.

See also the related poll result: 'Most outages of your scanning system are caused by failure of'
spacer
 
• Share the entry 'Magnetic Shielding':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Faraday Shield
    • Permanent Magnet
    • Superconducting Magnet
    • Room Shielding
    • Open MRI
 
Further Reading:
  Basics:
Faraday's Law
   by hyperphysics.phy-astr.gsu.edu    
  News & More:
Magnetic Sensitivity of MRI Systems to External Iron: The Design Process
   by www.integratedsoft.com    
MRI Resources 
Shielding - - Societies - Abdominal Imaging - Resources - Research Labs
 
Active Shielding
 
Magnetic shielding through the use of secondary shielding coils designed to produce a magnetic field that cancels the field from primary coils in regions where it is not desired. These coils may be inside the magnet cryostat. Active shielding can be applied to the main magnet or to the gradient magnetic fields.

See also Magnetic Shielding, Self Shielding and Room Shielding.
spacer

• View the DATABASE results for 'Active Shielding' (6).Open this link in a new window

MRI Resources 
Pathology - Universities - Raman Spectroscopy - Open Directory Project - Image Quality - Calculation
 
Passive Shielding
 
Magnetic shielding through the use of high permeability material. The iron provides a return path for the stray field lines of magnetic flux and so significantly decreases the flux away from the magnet.
Passive shielding (see also Faraday cage) significantly eases the problems of siting a MR imager in a confined space. Ferromagnetic objects are less prone to being attracted to the magnet, ancillary electronic equipment, credit cards and computer disks can be brought closer to the magnet and the MRI safety limit for pacemaker wearers (the 5 gauss line = 0.5 mT) is reduced from, typically, 10 m to 2 m from the magnet. A passive shield for a whole-body MRI magnet weights many tons. An alternative method of controlling stray field is active shielding.

See also Active Shielding, Magnetic Shielding, Self Shielding and Room Shielding.
spacer

• View the DATABASE results for 'Passive Shielding' (3).Open this link in a new window

 
Further Reading:
  Basics:
Faraday's Law
   by hyperphysics.phy-astr.gsu.edu    
  News & More:
Magnetic Sensitivity of MRI Systems to External Iron: The Design Process
   by www.integratedsoft.com    
Searchterm 'Magnetic Shielding' was also found in the following services: 
spacer
News  (1)  Resources  (2)  
 
Room Shielding
 
Magnetic shielding through the use of high permeability material in the walls (plus floor and ceiling) of the magnet room. Room shielding can be complete (e.g., six sides of a box, Faraday cage), or partial if the fringe field is to be reduced only in certain areas (see also Magnetic Shielding).
spacer

• View the DATABASE results for 'Room Shielding' (5).Open this link in a new window

 
Further Reading:
  Basics:
Faraday's Law
   by hyperphysics.phy-astr.gsu.edu    
  News & More:
Magnetic Sensitivity of MRI Systems to External Iron: The Design Process
   by www.integratedsoft.com    
MRI Resources 
Research Labs - Most Wanted - Lung Imaging - Developers - NMR - Resources
 
Self Shielding
 
Magnetic shielding by attaching a high permeability yoke to the magnet (passive shielding) or by incorporating additional magnetic field-generating coils designed to reduce the external field (active shielding).

See also Magnetic shielding, Active Shielding and Passive Shielding.
spacer

• View the DATABASE results for 'Self Shielding' (3).Open this link in a new window

MRI Resources 
Bioinformatics - Databases - MRI Physics - Coils - Stent - Societies
 
     1 - 5 (of 19)     next
Result Pages : [1]  [2 3]  [4]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 29 March 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]