Sequential Plane Imaging
 
MR imaging technique in which the image of an object is built up from successive planes in the object. In various schemes, the planes are selected by oscillating magnetic field gradients or selective excitation.
spacer
Searchterm 'Excitation' was also found in the following services: 
spacer
Radiology  (5) Open this link in a new windowUltrasound  (6) Open this link in a new window
Single Shot Technique
 
In single shot techniques (used for EPI, TSE, FSE, RARE, HASTE), the entire raw data set is acquired with a single excitation pulse. The magnetization of a fully relaxed spin system is used. Each of the subsequent echoes is given a different phase encoding. For improved SNR, spatial resolution or FOV, the needed raw data are acquired over a number of sequence repetitions. Each repetition then collects a fraction of the complete raw data set. Only slightly more than a half of the raw data is acquired. The image is obtained through half Fourier reconstruction.
A single shot sequence is useful in cases where movement is to expect e.g. in abdominal Imaging or fetal MRI.

See also Half Fourier Acquisition Single Shot Turbo Spin Echo.
spacer

• View the DATABASE results for 'Single Shot Technique' (7).Open this link in a new window

 
Further Reading:
  Basics:
Clinical evaluation of a speed optimized T2 weighted fast spin echo sequence at 3.0 T using variable flip angle refocusing, half-Fourier acquisition and parallel imaging
Wednesday, 25 October 2006
MRI Resources 
Chemistry - Movies - Jobs pool - Cardiovascular Imaging - Artifacts - Pathology
 
Slice SelectionForum -
related threads
 
Exclusive excitation of spins in one slice performed by the coincident combination of a gradient magnetic field and a narrow bandwidth or slice selective RF pulse at a specific Larmor frequency.
spacer

• View the DATABASE results for 'Slice Selection' (8).Open this link in a new window

 
Further Reading:
  Basics:
RARE
Monday, 3 December 2012   by www2.warwick.ac.uk    
  News & More:
Magnetic resonance imaging
   by www.scholarpedia.org    
Searchterm 'Excitation' was also found in the following services: 
spacer
News  (1)  Forum  (11)  
 
Spatially Localized Spectroscopy
 
Process by which regions of tissue are selectively sampled to produce spectra from defined volumes in space. These methods may be employed to sample a single region in space (single voxel method) or multiple regions simultaneously (multivoxel methods). The spatial selectivity can be achieved by a variety of methods including surface coils, surface coils in conjunction with RF gradient methods, or RF pulses in combination with switched magnetic field gradients, for example, volume-selective excitation. An indirect method of achieving spatial selectivity is the destruction of coherence of the magnetization in regions that lie outside the region of interest. A variety of spatial encoding schemes have been employed for multivoxel localization. See Chemical shift imaging.
spacer
Searchterm 'Excitation' was also found in the following services: 
spacer
Radiology  (5) Open this link in a new windowUltrasound  (6) Open this link in a new window
Spin Echo SequenceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
Spin Echo Timing Diagram (SE) The most common pulse sequence used in MR imaging is based of the detection of a spin or Hahn echo. It uses 90° radio frequency pulses to excite the magnetization and one or more 180° pulses to refocus the spins to generate signal echoes named spin echoes (SE).
In the pulse sequence timing diagram, the simplest form of a spin echo sequence is illustrated.
The 90° excitation pulse rotates the longitudinal magnetization (Mz) into the xy-plane and the dephasing of the transverse magnetization (Mxy) starts.
The following application of a 180° refocusing pulse (rotates the magnetization in the x-plane) generates signal echoes. The purpose of the 180° pulse is to rephase the spins, causing them to regain coherence and thereby to recover transverse magnetization, producing a spin echo.
The recovery of the z-magnetization occurs with the T1 relaxation time and typically at a much slower rate than the T2-decay, because in general T1 is greater than T2 for living tissues and is in the range of 100-2000 ms.
The SE pulse sequence was devised in the early days of NMR days by Carr and Purcell and exists now in many forms: the multi echo pulse sequence using single or multislice acquisition, the fast spin echo (FSE/TSE) pulse sequence, echo planar imaging (EPI) pulse sequence and the gradient and spin echo (GRASE) pulse sequence;; all are basically spin echo sequences.
In the simplest form of SE imaging, the pulse sequence has to be repeated as many times as the image has lines.
Contrast values:
PD weighted: Short TE (20 ms) and long TR.
T1 weighted: Short TE (10-20 ms) and short TR (300-600 ms)
T2 weighted: Long TE (greater than 60 ms) and long TR (greater than 1600 ms)
With spin echo imaging no T2* occurs, caused by the 180° refocusing pulse. For this reason, spin echo sequences are more robust against e.g., susceptibility artifacts than gradient echo sequences.

See also Pulse Sequence Timing Diagram to find a description of the components.
 
Images, Movies, Sliders:
 Shoulder Coronal T1 SE  Open this link in a new window
    

Courtesy of  Robert R. Edelman
 Shoulder Axial T1 SE  Open this link in a new window
 MRI Orbita T1  Open this link in a new window
    
 
spacer

• View the DATABASE results for 'Spin Echo Sequence' (24).Open this link in a new window

 
Further Reading:
  Basics:
Fast Spin Echo(.pdf)
Tuesday, 24 January 2006   by www.81bones.net    
Magnetic resonance imaging
   by www.scholarpedia.org    
FUNDAMENTALS OF MRI: Part I
   by www.e-radiography.net    
  News & More:
New MR sequence helps radiologists more accurately evaluate abnormalities of the uterus and ovaries
Thursday, 23 April 2009   by www.eurekalert.org    
MRI techniques improve pulmonary embolism detection
Monday, 19 March 2012   by medicalxpress.com    
MRI Resources 
Spine MRI - MRI Technician and Technologist Schools - Stimulator pool - Fluorescence - Used and Refurbished MRI Equipment - MRI Physics
 
previous      81 - 85 (of 96)     next
Result Pages : [1 2]  [3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 24 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]