Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Echo' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Echo' found in 100 terms [] and 162 definitions []
previous     16 - 20 (of 262)     next
Result Pages : [1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20]  [ ... ]
Searchterm 'Echo' was also found in the following services: 
spacer
News  (7)  Resources  (10)  Forum  (35)  
 
Stimulated Echo
 
A form of a spin echo produced by three pulse RF sequences, consisting of two RF pulses following an initial exciting RF pulse. The stimulated echo appears at a time delay after the third pulse equal to the interval between the first two pulses. Although classically produced with 90° pulses, any RF pulses other than an ideal 180° can produce a stimulated echo. The intensity of the echo depends in part on the T1 relaxation time because the excitation is 'stored' as longitudinal magnetization between the second and third RF pulses. For example, use of stimulated echoes with spatially selective excitation with orthogonal magnetic field gradients permits volume-selective excitation for spectroscopic localization.
mri safety guidance
Image Guidance
Artifacts may appear as a series of fine lines. A narrow bandwidth causes a wide read window, which allows the stimulated echo to be incorporated into the image data. This can be supported by increasing the received bandwidth, which would narrow the read window, thus not incorporating the extraneous echo. Another help would be to change the first echo time, which may change the spacing of the stimulated echoes to outside that of the read window for the second echo.
spacer
 
• Related Searches:
    • Motion Probing Gradient
    • Point Resolved Spectroscopy
    • Cardiac MRI
    • Volume Selective Excitation
    • Radio Frequency Pulse
 
Further Reading:
  Basics:
Magnetic resonance imaging
   by www.scholarpedia.org    
Clinical evaluation of a speed optimized T2 weighted fast spin echo sequence at 3.0 T using variable flip angle refocusing, half-Fourier acquisition and parallel imaging
Wednesday, 25 October 2006
Searchterm 'Echo' was also found in the following services: 
spacer
Radiology  (4) Open this link in a new windowUltrasound  (159) Open this link in a new window
Turbo Spin EchoInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(TSE) A pulse sequence characterized by a series of rapidly applied 180° rephasing pulses and multiple echoes.

See Fast spin echo.
 
Images, Movies, Sliders:
 MRI Liver T2 TSE  Open this link in a new window
    
 Brain MRI Images Axial T2  Open this link in a new window
 Breast MRI 2 Transverse T1 001  Open this link in a new window
    
 
spacer

• View the DATABASE results for 'Turbo Spin Echo' (17).Open this link in a new window

 
Further Reading:
  Basics:
Fast Spin Echo(.pdf)
Tuesday, 24 January 2006   by www.81bones.net    
Advances in Magnetic Resonance Neuroimaging
Friday, 27 February 2009   by www.ncbi.nlm.nih.gov    
  News & More:
Spin echoes, CPMG and T2 relaxation - Introductory NMR & MRI from Magritek
2013   by www.azom.com    
Comparison of Fast Spin-Echo Versus Conventional Spin-Echo MRI forEvaluating Meniscal Tears
June 2005   by www.ajronline.org    
MRI Resources 
Directories - Patient Information - Safety pool - Collections - Coils - MR Myelography
 
Number of EchoesInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
The number of series of 180° RF rephasing pulses and their corresponding echoes for a fast or turbo spin echo pulse sequence is named echo train length (ETL) or turbo spin echo turbo factor (TSE TF).
spacer

• View the DATABASE results for 'Number of Echoes' (4).Open this link in a new window

Searchterm 'Echo' was also found in the following services: 
spacer
News  (7)  Resources  (10)  Forum  (35)  
 
Turbo Field EchoInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(TFE) Turbo field echo is a gradient echo pulse sequence with data acquisition after an initial 180° (similar to IR) preparation pulse for contrast enhancement. The difference between a FFE and TFE other than the speed of the sequence is that the image is acquired while approaching steady state (the echoes are collected during the time in which the tissues are experiencing T1 relaxation).
The contrast is prepared one time, which means the contrast is changing while the echoes are collected and can be manipulated by selecting the type and timing of the prepulse. A delay time is given before the actual image acquisition. To achieve T1 contrast the 180° prepulse is followed by an operator selected delay time, that results in no signal from the targeted tissue. So when the echoes are acquired, no signal is present, additional RF spoiling is performed to optimize for T1 contrast. The delay chosen corresponds to when T1 relaxation reaches and suppresses T1 signal or optimizes the difference between tissues. Contrast for these sequences are enhanced when K-space is filled using a centric or low-high ordering. A TFE can be acquired with a 2D or 3D technique and with or without T1, T2 weighting.
See Ultrafast Gradient Echo Sequence, TurboFLASH and Magnetization Prepared Rapid Gradient Echo (MPRAGE).
spacer

• View the DATABASE results for 'Turbo Field Echo' (6).Open this link in a new window

 
Further Reading:
  Basics:
Sequence for Philips(.pdf)
   by www.droid.cuhk.edu.hk    
Pediatric and Adult Cochlear Implantation1
2003   by radiographics.rsnajnls.org    
Searchterm 'Echo' was also found in the following services: 
spacer
Radiology  (4) Open this link in a new windowUltrasound  (159) Open this link in a new window
Echo Train Length
 
The number of series of 180° RF rephasing pulses and their corresponding echoes for a fast or turbo spin echo pulse sequence.

See Fast Spin Echo.
spacer

• View the DATABASE results for 'Echo Train Length' (4).Open this link in a new window

 
Further Reading:
  Basics:
Fast Spin Echo(.pdf)
Tuesday, 24 January 2006   by www.81bones.net    
  News & More:
Comparison of Fast Spin-Echo Versus Conventional Spin-Echo MRI forEvaluating Meniscal Tears
June 2005   by www.ajronline.org    
MRI Resources 
Patient Information - Used and Refurbished MRI Equipment - Cardiovascular Imaging - Resources - Examinations - MRCP
 
previous      16 - 20 (of 262)     next
Result Pages : [1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20]  [ ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 19 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]