Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 

Next big thing in MRI will be :
AI 
remote operator 
personalized protocols 
helium-free 
molecular MRI 
portable MRI 


 
MRI Artifacts
  
Ultrasound Imaging Artifacts Open this link in a new window
FID Signal Artifact 
Quick Overview
Artifact Information
NAME
FID (with spin echo) signal
DESCRIPTION
Line across the center of the image
REASON
Combination of problems
HELP
Call the service
A combination of B1 inhomogeneity, poor slice profile, and insufficient spoiler gradients between the refocusing pulse and the readout interval of a spin echo sequence results in a FID signal being detected along with the echo.
Since the FID is not phase encoded (normally the phase encoding occurs before the refocusing pulse), it is not dispersed along the phase encoding axis, but appears as a line across the center of the image.
mri safety guidance
Image Guidance
If the problem persists, it must be addressed by a service representative.

• View the DATABASE results for 'FID Signal Artifact' (2).Open this link in a new window

 
Further Reading:
  News & More:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
Similar Artifacts:
 • Receiver turn on Artifact
 • Central Point Artifact
 • Field Inhomogeneity Artifact
  Field Inhomogeneity Artifact top
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Look
      Ups




MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 19 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]