 | Info Sheets |
| | | | | | | | | | | | | | | | | | | | | | |
 | Out- side |
| | | | |
|
| | | | |
Result : Searchterm 'Intera 1.0T™' found in 1 term [ ] and 1 definition [ ]
| 1 - 2 (of 2) Result Pages : [1] | |  | | |  |  |
| Intera 1.0T™ |   |
| |
|
| From Philips Medical Systems;
|
the Intera-family offers with this mid field system maximum efficiency, flexibility and performance. All Philips MR products contain SENSE (coil SENSitivity Encoding) technology - that increases scanning speed.
|
|
Device Information and Specification |
| CLINICAL APPLICATION |
Whole body |
| CONFIGURATION |
Short bore compact |
| SURFACE COILS |
Standard: head, body, C1, C3; Optional: Small joint, flex-E, flex-R, endocavitary (L and S), dual TMJ, knee, neck, T/L spine, breast; Optional phased array: Spine, pediatric, 3rd party connector, Optional SENSE Coils: Flex-S-M-L, Flex Body, Flex Cardiac, |
| SPECTROSCOPY |
Optional |
| SYNCHRONIZATION |
ECG/peripheral: Optional/yes, respiratory gating |
| PULSE SEQUENCES |
SE, Modified-SE, IR (T1, T2, PD), STIR, FLAIR, SPIR, FFE, T1-FFE, T2-FFE, Balanced FFE, TFE, Balanced TFE, Dynamic, Keyhole, 3D, Multi Chunk 3D, Multi Stack 3D, K Space Shutter, MTC, TSE, Dual IR, DRIVE, EPI, Cine, 2DMSS, DAVE, Mixed Mode; Angiography: Inflow MRA, TONE, PCA, CE MRA |
| IMAGING MODES |
Single Slice 2D , Multi Single Slice 2D, Multi Slice 2D, 3D, Multi Chunk 3D, Multi Stack 3D |
| TR |
Min. 2.9 (Omni) msec, 1.6 (Power) msec |
| TE |
Min. 1.0 (Omni) msec, 0.7 (Power) msec |
| SINGLE/MULTI SLICE |
RapidView Recon. greater than 500 @ 256 Matrix |
| FOV |
Max. 53 cm |
| SLICE THICKNESS |
0.1 mm(Omni), 0.05 mm (Power) |
| DISPLAY MATRIX |
128 x 128, 256 x 256,512 x 512,1024 x 1024 (64 for Bold img) |
| MEASURING MATRIX |
Variable in 1% increments |
| PIXEL INTENSITY |
Lum.: 120 cd/m2; contrast: 150:1 |
| SPATIAL RESOLUTION |
Variable (op. param. depend.) |
| MAGNET TYPE |
Superconducting |
BORE DIAMETER or W x H |
60 x 60 cm |
| MAGNET WEIGHT |
2700 kg |
| H*W*D |
240 x 188 x 157 cm |
| POWER REQUIREMENTS |
380/400 V |
| COOLING SYSTEM TYPE |
|
| CRYOGEN USE |
0.03 L/hr helium |
| FIELD STRENGTH |
1.0 T |
| STRENGTH |
23 mT/m (Omni), 30 (Power) mT/m |
| 5-GAUSS FRINGE FIELD |
2.3 m / 3.3 m |
| SHIMMING |
Passive and dynamic |
|
| |  | | | |
|  |  |
| Philips Medical Systems |  |
| |
|
Philips Medical System is the diagnostics business of Royal Philips Electronics of the Netherlands, one of the world's biggest electronics companies and Europe's largest. Philips is quoted on the NYSE (symbol: PHG), London, Frankfurt, Amsterdam and other stock exchanges.
On October 19, 2001, Philips Medical Systems completed a 3-year acquisition strategy through its purchase of Marconi Medical Systems. Marconi Medical Systems offered leading multislice CT, MRI, and Nuclear Gamma Camera systems to medical institutions around the world. As well as new 3.0T developments, Philips is also in collaboration with researchers at the University of Nottingham, with the intention of developing an ultrahigh field strength clinical 7.0T whole body MR system.
| |