Spoiler Gradient PulseInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
Magnetic field gradient pulse applied to effectively remove transverse magnetization by producing a rapid variation of its phase along the direction of the gradient. This is done after the echo so that transverse magnetization is destroyed prior to the next excitation pulse, to spoil any remaining xy-magnetization or to refocus the xy-magnetization.
For example, when used to remove the unwanted signal resulting from an imperfect 180° refocusing RF pulse, a corresponding compensating gradient pulse may be applied prior to the refocusing RF pulse in order to avoid spoiling the desired transverse magnetization resulting from the initial excitation. Also called homospoil pulse.
spacer
 
• Share the entry 'Spoiler Gradient Pulse':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Homospoil
    • Refocusing
    • Saturation
    • Spoiled Gradient Echo Sequence
    • Saturation Recovery
 
Further Reading:
  News & More:
Clinical evaluation of a speed optimized T2 weighted fast spin echo sequence at 3.0 T using variable flip angle refocusing, half-Fourier acquisition and parallel imaging
Wednesday, 25 October 2006
MRI Resources 
Research Labs - PACS - Cardiovascular Imaging - RIS - Absorption and Emission - Developers
 
Dephasing Gradient
 
Magnetic field gradient pulse used to create spatial variation of phase of transverse magnetization. For example, it may be applied prior to signal detection in the presence of a magnetic field gradient with opposite polarity (or of the same polarity if separated by a refocusing RF pulse) so that the resulting gradient echo signal will represent a more complete sampling of the Fourier transformation of the desired image.

See also Spoiler Gradient Pulse.
spacer

• View the DATABASE results for 'Dephasing Gradient' (6).Open this link in a new window

 
Further Reading:
  Basics:
RARE
Monday, 3 December 2012   by www2.warwick.ac.uk    
MRI Resources 
Calculation - Safety Training - Safety pool - Shoulder MRI - Corporations - MRI Accidents
 
Homospoil
 
Use of a magnetic field gradient to effectively eliminate residual transverse magnetization by producing a strong position dependence of phase within a resolution element. See Spoiler Gradient Pulse.
spacer

• View the DATABASE results for 'Homospoil' (2).Open this link in a new window

MRI Resources 
Supplies - MRI Technician and Technologist Jobs - Spectroscopy - Collections - Liver Imaging - Lung Imaging
 
Radio Frequency Spoiling
 
The use of varying phase or timing of the RF pulses to prevent setting up a condition of steady state free precession, e.g. in rapid excitation MR imaging.

See also Spoiler Gradient Pulse and Rapid Excitation Magnetic Resonance Imaging.
spacer
MRI Resources 
Image Quality - MRA - Liver Imaging - Sequences - Education pool - Shielding
 
Rapid Excitation Magnetic Resonance ImagingInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(RE MRI) There are several approaches to speeding up the MRI data acquisition process by repeating the excitation by RF pulses in times short compared to T1, typically using small flip angles and gradient echo refocusing. When TR is also on the order of or shorter than T2, the repeated RF pulses will tend to refocus transverse magnetization remaining from prior excitations, setting up a condition of steady state free precession, and a dependence of signal strength (and image contrast) on both T1 and T2.
This can be modified in various ways, particularly:
1) to spoil the tendency to build up a steady state by reducing coherence between excitations, e.g. by variation of the phase or timing of consecutive RF pulses or of the strength of spoiler gradient pulses, thus increasing the relative dependence of signal strength on T1 or
2) acquire the signal when it is refocusing immediately prior to the next RF pulse, thus increasing the relative dependence of signal strength on T2.

See also Ultrafast Gradient Echo Sequence.
spacer

• View the DATABASE results for 'Rapid Excitation Magnetic Resonance Imaging' (2).Open this link in a new window

MRI Resources 
Crystallography - Research Labs - MRI Reimbursement - Veterinary MRI - Corporations - Blood Flow Imaging
 
     1 - 5 (of 14)     next
Result Pages : [1]  [2]  [3]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Next big thing in MRI will be :
AI 
remote operator 
personalized protocols 
helium-free 
molecular MRI 
portable MRI 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 25 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]