Short T1 Inversion RecoveryInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(STIR) Also called Short Tau (t) (inversion time) Inversion Recovery. STIR is a fat suppression technique with an inversion time t = T1 ln2 where the signal of fat is zero (T1 is the spin lattice relaxation time of the component that should be suppressed). To distinguish two tissue components with this technique, the T1 values must be different. Fluid Attenuation Inversion Recovery (FLAIR) is a similar technique to suppress water.
Inversion recovery doubles the distance spins will recover, allowing more time for T1 differences. A 180° preparation pulse inverts the net magnetization to the negative longitudinal magnetization prior to the 90° excitation pulse. This specialized application of the inversion recovery sequence set the inversion time (t) of the sequence at 0.69 times the T1 of fat. The T1 of fat at 1.5 Tesla is approximately 250 with a null point of 170 ms while at 0.5 Tesla its 215 with a 148 ms null point. At the moment of excitation, about 120 to 170 ms after the 180° inversion pulse (depending of the magnetic field) the magnetization of the fat signal has just risen to zero from its original, negative, value and no fat signal is available to be flipped into the transverse plane.
When deciding on the optimal T1 time, factors to be considered include not only the main field strength, but also the tissue to be suppressed and the anatomy. In comparison to a conventional spin echo where tissues with a short T1 are bright due to faster recovery, fat signal is reversed or darkened. Because body fluids have both a long T1 and a long T2, it is evident that STIR offers the possibility of extremely sensitive detection of body fluid. This is of course, only true for stationary fluid such as edema, as the MRI signal of flowing fluids is governed by other factors.

See also Fat Suppression and Inversion Recovery Sequence.
 
Images, Movies, Sliders:
 Sagittal Knee MRI Images STIR  Open this link in a new window
      

 
spacer
 
• Related Searches:
    • Black Blood MRA
    • Flow
    • Pulse, 180°
    • Chemical Shift Imaging
    • Image Contrast Characteristics
 
Further Reading:
  Basics:
Can Short Tau Inversion Recovery (STIR) Imaging Be Used as a Stand-Alone Sequence To Assess a Perianal Fistulous Tract on MRI? A Retrospective Cohort Study Comparing STIR and T1-Post Contrast Imaging
Wednesday, 17 January 2024   by www.cureus.com    
  News & More:
Generating Virtual Short Tau Inversion Recovery (STIR) Images from T1- and T2-Weighted Images Using a Conditional Generative Adversarial Network in Spine Imaging
Wednesday, 25 August 2021
Short tau inversion recovery (STIR) after intravenous contrast agent administration obscures bone marrow edema-like signal on forefoot MRI
Tuesday, 13 July 2021   by www.springermedizin.de    
Searchterm 'Spin' was also found in the following services: 
spacer
Radiology  (12) Open this link in a new windowUltrasound  (2) Open this link in a new window
Slice Overlap ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Slice overlap
DESCRIPTION
Loss of signal
REASON
HELP
Overlapping prevention
The slice overlap artifact is another name for crosstalk artifact. If slices of multislice acquisitions are overlapping, the spinning nuclei belonging to more than one slice getting multiple times saturated, which leads to signal loss in this areas.
mri safety guidance
Image Guidance
This problem occurs often in cervical or lumbar spine MRI, when scanning each disc with multi angle oblique technique. If prevention of overlapping is not possible, try to position the saturated region posterior to the spinal canal, outside the region of interest.

See also Crosstalk (Crosstalk), and Multiple Slice Imaging.
spacer
 
Further Reading:
  Basics:
Slice-overlap Artifacts
   by www.mritutor.org    
MRI Resources 
Databases - Spine MRI - Guidance - Education pool - Coils - Stimulator pool
 
Spectral Editing
 
Methods of selectively enhancing or suppressing the signal from a particular molecular substance by using its spin properties, typically through spin spin coupling, e.g. J-modulation.
spacer
Searchterm 'Spin' was also found in the following services: 
spacer
News  (64)  Resources  (27)  Forum  (47)  
 
Spectroscopy
 
Edward Purcell and Felix Bloch discovered the basic of spectroscopy in 1946 (see MRI History). Nuclear magnetic resonance spectroscopy (NMR Spectroscopy or MRS) is an analytical tool, based on nuclei that have a spin (nuclei with an odd number of neutrons and/or protons) like 1H, 13C, 17O, 19F, 31P etc.
Through nuclear magnetic principles as precession, chemical shift, spin spin coupling etc., the analysis of the content, purity, and molecular structure of a sample is possible. The spectrum produced by this process contains a number of peaks; the highs and the positions of these peaks allow the exact analysis. Unknown compounds can be matched against spectral libraries. Even very complex organic compounds as enzymes and proteins can be determined. For the wide uses of NMR spectroscopy (from mineralogy to medicine) there is a variety of different techniques available.
See Spectroscopic Imaging Techniques.
spacer

• View the DATABASE results for 'Spectroscopy' (90).Open this link in a new window


• View the NEWS results for 'Spectroscopy' (3).Open this link in a new window.
 
Further Reading:
  Basics:
MR Spectroscopy May Help Avoid Invasive Procedures And Treatments For Recurrent Brain Lesions
Wednesday, 11 February 2009   by www.sciencedaily.com    
MRI spectroscopy is highly sensitive for lipid-soluble metabolites from UC-MSCs
Monday, 9 September 2013   by phys.org    
  News & More:
Pioneering MRI imaging method captures brain glucose metabolism without the need for administration of radioactive substances
Friday, 28 April 2023   by www.eurekalert.org    
New quantum sensing technique allows high-resolution nuclear magnetic resonance spectroscopy
Wednesday, 17 June 2020   by phys.org    
MR Spectroscopy Detects Biochemical Alterations in Pre-Invasive Breast Cancer Patients
Wednesday, 11 March 2015   by radiationtherapynews.com    
MR Spectroscopy Shows Differences in Brains of Preterm Infants
Monday, 25 November 2013   by www.digitaljournal.com    
Proton Magnetic Resonance Spectroscopy and MRI Reveal No Evidence for Brain Mitochondrial Dysfunction in Children with Autism Spectrum Disorder.
Wednesday, 16 March 2011   by leftbrainrightbrain.co.uk    
Magnetic resonance spectroscopy for breast cancer
Wednesday, 11 July 2007   by www.news-medical.net    
Searchterm 'Spin' was also found in the following services: 
spacer
Radiology  (12) Open this link in a new windowUltrasound  (2) Open this link in a new window
Time Echo
 
(TE) Time between middle of exciting (e.g., 90°) RF pulse and middle of spin echo production. For multiple echoes, use TE1, TE2, etc. When the RF spin echo and gradient echo are not coincident in time, TE refers to the time of the gradient spin echo. See also Echo Time.
spacer
MRI Resources 
Lung Imaging - Breast Implant - MR Myelography - Education pool - Brain MRI - Chemistry
 
previous      86 - 90 (of 317)     next
Result Pages : [1 2 3 4 5 6 7 8 9 10]  [11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]