Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Repetition Time' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Repetition Time' found in 1 term [] and 32 definitions []
previous     26 - 30 (of 33)     next
Result Pages : [1]  [2 3 4 5 6 7]
Searchterm 'Repetition Time' was also found in the following services: 
spacer
News  (1)  Forum  (1)  
 
Specific Absorption Rate
 
(SAR) The Specific Absorption Rate is defined as the RF power absorbed per unit of mass of an object, and is measured in watts per kilogram (W/kg).
The SAR describes the potential for heating of the patient's tissue due to the application of the RF energy necessary to produce the MR signal. Inhomogeneity of the RF field leads to a local exposure where most of the absorbed energy is applied to one body region rather than the entire person, leading to the concept of a local SAR. Hot spots may occur in the exposed tissue, to avoid or at least minimize effects of such theoretical complications, the frequency and the power of the radio frequency irradiation should be kept at the lowest possible level. Averaging over the whole body leads to the global SAR.
It increases with field strength, radio frequency power and duty cycle, transmitter-coil type and body size. The doubling of the field strength from 1.5 Tesla (1.5T) to 3 Tesla (3T) leads to a quadrupling of SAR. In high and ultrahigh fields, some of the multiple echo, multiple-slice pulse sequences may create a higher SAR than recommended by the agencies. SAR can be reduced by lower flip angle and longer repetition times, which could potentially affect image contrast.
Normally no threatening increase in temperature could be shown. Even in high magnetic fields, the local temperature increases not more than 1°C. 2.1°C is the highest measured increase in skin temperature. Eddy currents may heat up implants and thus may cause local heating.

FDA SAR limits:
•
Whole body: 4W/kg/15-minute exposure averaged;
•
Head: 3W/kg/10-minute exposure averaged;
•
Head or torso: 8W/kg/5 minute exposure per gram of tissue;
•
Extremities: 12W/kg/5 minute exposure per gram of tissue.

IEC (International Electrotechnical Commission) SAR limits of some European countries:
All limits are averaged over 6 minutes.
•
Level 0 (normal operating mode): Whole body 2W/kg; Head 3.2W/kg; Head or Torso (local) 10W/kg; Extremities (local) 20W/kg;
•
Level I (first level controlled operating mode): Whole body 4W/kg; Head 3.2W/kg; Head or Torso (local) 10W/kg; Extremities (local) 20W/kg;
•
Level II (second level controlled operating mode): All values are over Level I values.
(For more details: IEC 60601-2-33 (2002))

In most countries standard MRI systems are limited to a maximum SAR of 4 W/kg, so most scanning in level II is impossible.
For Level I, in addition to routine monitoring, particular caution must be exercised for patients who are sensitive to temperature increases or to RF energy.
For Japan different SAR limits are valid.
spacer
 
• Related Searches:
    • MRI Safety
    • Magnetic Resonance Imaging MRI
    • MRI Risks
    • Low Field MRI
    • High Field MRI
 
Further Reading:
  Basics:
SED Guidance
Saturday, 1 January 2022   by www.mriphysics.scot.nhs.uk    
On the estimation of the worst-case implant-induced RF-heating in multi-channel MRI.
Thursday, 2 March 2017   by www.ncbi.nlm.nih.gov    
What MRI Sequences Produce the Highest Specific Absorption Rate (SAR), and Is There Something We Should Be Doing to Reduce the SAR During Standard Examinations?
Thursday, 16 April 2015   by www.ajronline.org    
Evaluation of Specific Absorption Rate as a Dosimeter of MRI-Related Implant Heating
2004   by www.imrser.org    
  News & More:
Specific Absorption Rate and Specific Energy Dose: Comparison of 1.5-T versus 3.0-T Fetal MRI
Tuesday, 7 April 2020   by pubs.rsna.org    
MRI in Patients with Implanted Devices: Current Controversies
Monday, 1 August 2016   by www.acc.org    
Commission delays electromagnetic fields legislation
Monday, 29 October 2007   by cordis.europa.eu:80    
Accounting for biological aggregation in heating and imaging of magnetic nanoparticles
Tuesday, 2 September 2014   by www.ecnmag.com    
Guidance for Industry and FDA Staff, Criteria for Significant Risk Investigations of Magnetic Resonance Diagnostic Devices
Monday, 14 July 2003   by www.fda.gov    
MRI Resources 
Knee MRI - Anatomy - Jobs - Open Directory Project - Directories - Used and Refurbished MRI Equipment
 
Spin Echo Imaging
 
MR imaging technique in which the spin echo is used rather than the FID. The proper selection of the echo and repetition time of the pulse sequence can help to control the amount of T1 or T2 contrast present in the image. Spin echo imaging can be used to create images that depend strongly on T2 if TE has a value on the order of or greater than T2 of the relevant image details. The spin echoes can be produced as a train of multiple echoes, e.g. using the CPMG pulse sequence.

See also Spin Echo Sequence.
 
Images, Movies, Sliders:
 Lumbar Spine T1 SE Sagittal  Open this link in a new window
    

Courtesy of  Robert R. Edelman
 MRI Orbita T1  Open this link in a new window
    
 
spacer

• View the DATABASE results for 'Spin Echo Imaging' (4).Open this link in a new window

MRI Resources 
Chemistry - Colonography - IR - Jobs pool - MRI Reimbursement - Breast Implant
 
Spin Phase EffectInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.
 
The spins flow with the blood through a slice and experience a RF pulse. If they flow out of the slice by the time the signal is recorded (because the repetition time (TR) is asynchronous with the pulsatile flow), the flowing blood produces intravascular signal void by 'time of flight' effects, turbulent dephasing and first echo dephasing. The liquid flow occasionally produces an intravascular high signal intensity due to flow related enhancement, even echo rephasing and diastolic pseudogating.

See also Flow Artifact and Flow Effects.
 
Images, Movies, Sliders:
 TOF-MRA Circle of Willis Inverted MIP  Open this link in a new window
    

 
spacer

• View the DATABASE results for 'Spin Phase Effect' (3).Open this link in a new window

Searchterm 'Repetition Time' was also found in the following services: 
spacer
News  (1)  Forum  (1)  
 
Spoiled Gradient Echo SequenceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
Spoiled gradient echo sequences use a spoiler gradient on the slice select axis during the end module to destroy any remaining transverse magnetization after the readout gradient, which is the case for short repetition times.
As a result, only z-magnetization remains during a subsequent excitation. This types of sequences use semi-random changes in the phase of radio frequency pulses to produce a spatially independent phase shift.
Companies use different acronyms to describe certain techniques.

Different terms for these gradient echo pulse sequences:
CE-FFE-T1 Contrast Enhanced Fast Field Echo with T1 Weighting,
GFE Gradient Field Echo,
FLASH Fast Low Angle Shot,
PS Partial Saturation,
RF spoiled FAST RF Spoiled Fourier Acquired Steady State Technique,
RSSARGE Radio Frequency Spoiled Steady State Acquisition Rewound Gradient Echo
S-GRE Spoiled Gradient Echo,
SHORT Short Repetition Techniques,
SPGR Spoiled Gradient Recalled (spoiled GRASS),
STAGE T1W T1 weighted Small Tip Angle Gradient Echo,
T1-FAST T1 weighted Fourier Acquired Steady State Technique,
T1-FFE T1 weighted Fast Field Echo.
In this context, 'contrast enhanced' refers to the pulse sequence, it does not mean enhancement with a contrast agent.
spacer

• View the DATABASE results for 'Spoiled Gradient Echo Sequence' (11).Open this link in a new window

 
Further Reading:
  News & More:
3-D VOLUMETRIC IMAGING FOR STEREOTACTIC LESIONAL AND DEEP BRAIN STIMULATION SURGERY
Cutting Edge Imaging of THE Spine
February 2007   by www.pubmedcentral.nih.gov    
MRI Resources 
Crystallography - Raman Spectroscopy - MRI Physics - - Open Directory Project - Blood Flow Imaging
 
Spoiled Steady State Acquisition Rewinded Gradient EchoInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(SARGE) Spoiled GRE sequences use a spoiler gradient on the slice select axis to destroy any remaining transverse magnetization after the readout gradient, with the result of short repetition times. This type of sequences use semi-random changes in the phase of RF pulses to produce a spatially independent phase shift.

See also Gradient Echo Sequence.
spacer
MRI Resources 
Corporations - Online Books - Mobile MRI - Cochlear Implant - Blood Flow Imaging - Contrast Enhanced MRI
 
previous      26 - 30 (of 33)     next
Result Pages : [1]  [2 3 4 5 6 7]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 20 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]