Quadrature CoilInfoSheet: - Coils - 
Intro, 
Overview, 
etc.MRI Resource Directory:
 - Coils -
 
A coil that produces an RF field with circular polarization. The RF power received from the RF power amplifier comes in two signals (quadrature detection), which have a phase difference of 90°. The RF transmit coil converts the power into a circularly polarized RF magnetic field.
Quadrature coils can be used as both, transmit and/or receive coil. When used as a transmitter coil a factor of two power reduction over a linear coil results; as a receiver an increase in SNR of up to a factor of √2, can be achieved.

See also the related poll result: '3rd party coils are better than the original manufacturer coils'
spacer
 
• Share the entry 'Quadrature Coil':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Sense Coil
    • Imaging Coil
    • Array Coil
    • Signal to Noise Ratio
    • Receiver Coil
 
Further Reading:
  News & More:
HEAD AND NECK MRI GRADIENT COIL DESIGN
Thursday, 26 August 1999   by www.imaging.robarts.ca    
MRI Resources 
Fluorescence - Open Directory Project - Distributors - Contrast Enhanced MRI - Shielding - Resources
 
Receiver CoilInfoSheet: - Coils - 
Intro, 
Overview, 
etc.MRI Resource Directory:
 - Coils -
 
A coil, or antenna, positioned within the imaging volume and connected to the receiver circuitry that is used to detect or receive the MR signal from the patient as the disturbed spins relax back into their equilibrium distribution. Also called receive-only coil.
Special-purpose coils are designed to optimize the SNR from a given region of the body. State-of-the-art coil systems include the use of four or more coils with four separate receivers. This method is often referred to as a phased array system. Receiver coil types include also solenoid, planar, volume and quadrature coils. The quality of the MR images depends on the SNR of the acquired signal from the patient. SNR is of the utmost importance in obtaining clear images of the interior of the human body.
spacer

• View the DATABASE results for 'Receiver Coil' (11).Open this link in a new window

MRI Resources 
Contrast Enhanced MRI - Jobs - Pregnancy - Movies - Corporations - Diffusion Weighted Imaging
 
OPART™InfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www3.toshiba.co.jp/medical/products/mr/opart/index.htm From Toshiba America Medical Systems Inc.;
OPART™ is a 0.35 T superconducting open MR system featuring cryogenless operation. Superconducting speed and performance are combined with the patient access advantages of open MRI design for unmatched clinical versatility. OPART™ features innovative technologies such as digital RF, high speed gradients and optimized RF coils, which support advanced MRI applications.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Quadrature, solenoid and multi-channel configurations
SYNCHRONIZATION
ECG/peripheral optional, respiratory gating
PULSE SEQUENCES
SE, FE, IR, FastSE, FastIR, FastFLAIR, Fast STIR, FastFE, FASE, Hybrid EPI, Multi Shot EPI; Angiography: 2D(gate/non-gate)/3D TOF, SORS-STC
IMAGING MODES
Single, multislice, volume study
55 cm, vertical opening
POWER REQUIREMENTS
380/400/415/440/480 V
COOLING SYSTEM TYPE
Cryogenless
STRENGTH
25 mT/m
Passive, active
spacer

• View the DATABASE results for 'OPART™' (2).Open this link in a new window

MRI Resources 
Anatomy - Case Studies - Health - Spectroscopy pool - Safety pool - Knee MRI
 
Signa 3.0T™InfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www.gehealthcare.com/usen/mr/s_excite3/index.html (Signa VH/i 3.0T)
With GE Healthcare leading-edge technology in ultra-high-field imaging. The 3 T VH/i provides a platform for advanced applications in radiology, cardiology, psychology and psychiatry. Real-time image processing lets you acquire multislice whole brain images and map brain functions for research or surgical planning. And the 3 T Signa VH/i is flexible enough to provide clinicians with high performance they require. It can provide not only outstanding features in brain scanning and neuro-system research, but also a wide range of use in scanning breasts, extremities, the spine and the cardiovascular systems.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Cylindrical - high homogeneity
T/R quadrature head, T/R quadrature body, T/R phased array extremity (opt)
SYNCHRONIZATION
ECG/peripheral, respiratory gating
PULSE SEQUENCES
SE, IR, 2D/3D GRE, FGRE, RF-spoiled GRE, FSE, Angiography: 2D/3D TOF, 2D/3D phase contrast vascular
IMAGING MODES
Single, multislice, volume study, fast scan, multi slab, cine, localizer
SINGLE SLICE
100 Images/sec with Reflex100
MULTISLICE
100 Images/sec with Reflex100
1 cm to 40 cm continuous
2D 0.5-100mm in 0.1mm incremental
1280 x 1024
MEASURING MATRIX
128x512 steps 32 phase encode
PIXEL INTENSITY
256 gray levels
55cm
MAGNET WEIGHT
15102 kg incl. cryogen's
H*W*D
260cm x 238cm x 265cm
POWER REQUIREMENTS
480 or 380/415, 3 phase ||
COOLING SYSTEM TYPE
Closed-loop water-cooled grad.
Less than 0.14 L/hr liquid He
STRENGTH
40mT/m
5-GAUSS FRINGE FIELD, radial/axial
5.4 m x 3.2 m
Superconductive + hi order active
spacer

• View the DATABASE results for 'Signa 3.0T™' (2).Open this link in a new window

MRI Resources 
MRI Reimbursement - Jobs pool - RIS - Artifacts - Spine MRI - Musculoskeletal and Joint MRI
 
Signa Profile™InfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www.gehealthcare.com/usen/mr/profilei/index.html From GE Healthcare;
the New Signa Profile/i is a patient friendly open MRI system that virtually eliminates patient anxiety and claustrophobia, without compromising diagnostic utility.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Integrated transmit body coil, body flex sizes: M, L, XL, quadrature, head coil quadrature, 4 channel neurovascular array, 8 channel CTL array, quad. c-spine, 2 channel shoulder array, extremity coil, 3 channel wrist array, 4 channel breast array, 6, 9, 11 inch general purpose loop coils
SYNCHRONIZATION
Standard cardiac gating, ECG/peripheral, respiratory gating
PULSE SEQUENCES
Standard: SE, IR, 2D/3D GRE and SPGR, Angiography: 2D/3D TOF, 2D/3D phase contrast; 2D/3D FSE, 2D/3D FRFSE, FGRE and FSPGR, SSFP, FLAIR, EPI, optional: 2D/3D Fiesta, fat/water separation, T1 FLAIR
IMAGING MODES
Localizer, single slice, multislice, volume, fast, POMP, multi slab, cine, slice and frequency zip, extended dynamic range, tailored RF
TR
6 to 12000 msec in increments of 1 msec
TE
1.3 to 2000 msec in increments of 1 msec
SINGLE/MULTI SLICE
Simultaneous scan and reconstruction;; 80 images/second reconstruction
3cm to 40 cm continuous
2D: 2.7mm - 20mm 3D: 0.2mm - 5mm
1280 x 1024
MEASURING MATRIX
128x512 steps 32 phase//freq.
PIXEL INTENSITY
256 gray levels
0.08 mm; 0.02 mm optional
MAGNET TYPE
Permanent
120 x 44 cm
MAGNET WEIGHT
10,000 kg w/gradient enclosure
H*W*D
147 x 214 x 193 cm
POWER REQUIREMENTS
200 - 480, 3-phase
COOLING SYSTEM TYPE
None required
STRENGTH
15 mT/m
5-GAUSS FRINGE FIELD
1.7 m/1.7 m
spacer

• View the DATABASE results for 'Signa Profile™' (2).Open this link in a new window

MRI Resources 
Services and Supplies - Diffusion Weighted Imaging - Musculoskeletal and Joint MRI - Pacemaker - Sequences - Artifacts
 
     1 - 5 (of 11)     next
Result Pages : [1]  [2 3]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 29 March 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]