Image Contrast Characteristics
 
Variations in the value of TR and TE have an important effect on the control of image contrast characteristics.
Short values of TR (less than e.g. 1000 ms) and TE (less than e.g. 25 ms) are common in images exhibiting T1 contrast.
Long values of TR (greater than e.g. 1500 ms) and TE (greater than e.g.60 ms) are common in images exhibiting T2 contrast.
Middle TR values (e.g. from 1000 to 1500 ms) and middle TE values (e.g. from 25 to 60 ms) are common for density weighted contrast.
The values are depending on the field strength. TR is also a major factor in total scan time.
 
Images, Movies, Sliders:
 Knee MRI Coronal Pd Spir 001  Open this link in a new window
    
 Brain MRI Transversal T1 002  Open this link in a new window
    
 MRI Pancreas T2  Open this link in a new window
    
 Knee MRI Sagittal T1 004  Open this link in a new window
 Knee MRI Transverse 001  Open this link in a new window
 
Radiology-tip.comContrast,  Low Contrast Detectability
spacer
Radiology-tip.comSonographic Features,  Echogenicity
spacer
 
• Share the entry 'Image Contrast Characteristics':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Proton Density Weighted Image
    • T1 Weighted Image
    • Contrast Agents
    • Image Quality
    • T2 Weighted Image
 
Further Reading:
  Basics:
Musculoskeletal MRI at 3.0 T: Relaxation Times and Image Contrast
Sunday, 1 August 2004   by www.ajronline.org    
IMAGE CONTRAST IN MRI(.pdf)
   by www.assaftal.com    
MRI Resources 
MRI Technician and Technologist Jobs - Portals - Software - Raman Spectroscopy - Shielding -
 
Repetition Time
 
(TR) The amount of time that exists between successive pulse sequences applied to the same slice.
It is delineated by initiating the first RF pulse of the sequence then repeating the same RF pulse at a time t. Variations in the value of TR have an important effect on the control of image contrast characteristics. TR is also a major factor in total scan time.
See also Scan Time and Image Contrast Characteristics.
spacer

• View the DATABASE results for 'Repetition Time' (33).Open this link in a new window

 
Further Reading:
  Basics:
Magnetic resonance imaging
   by www.scholarpedia.org    
MRI Resources 
MRI Centers - RIS - Artifacts - Intraoperative MRI - Resources - Functional MRI
 
ContrastForum -
related threads
 
Contrast is the relative difference of signal intensities in two adjacent regions of an image.
Due to the T1 and T2 relaxation properties in magnetic resonance imaging, differentiation between various tissues in the body is possible. Tissue contrast is affected by not only the T1 and T2 values of specific tissues, but also the differences in the magnetic field strength, temperature changes, and many other factors. Good tissue contrast relies on optimal selection of appropriate pulse sequences (spin echo, inversion recovery, gradient echo, turbo sequences and slice profile).
Important pulse sequence parameters are TR (repetition time), TE (time to echo or echo time), TI (time for inversion or inversion time) and flip angle. They are associated with such parameters as proton density and T1 or T2 relaxation times. The values of these parameters are influenced differently by different tissues and by healthy and diseased sections of the same tissue.
For the T1 weighting it is important to select a correct TR or TI. T2 weighted images depend on a correct choice of the TE. Tissues vary in their T1 and T2 times, which are manipulated in MRI by selection of TR, TI, and TE, respectively. Flip angles mainly affect the strength of the signal measured, but also affect the TR/TI/TE parameters.
Conditions necessary to produce different weighted images:
T1 Weighted Image: TR value equal or less than the tissue specific T1 time - TE value less than the tissue specific T2 time.
T2 Weighted Image: TR value much greater than the tissue specific T1 time - TE value greater or equal than the tissue specific T2 time.
Proton Density Weighted Image: TR value much greater than the tissue specific T1 time - TE value less than the tissue specific T2 time.
See also Image Contrast Characteristics, Contrast Reversal, Contrast Resolution, and Contrast to Noise Ratio.

 
Images, Movies, Sliders:
 Fetus (Brain) and Dermoid in Mother  Open this link in a new window
      

Courtesy of  Robert R. Edelman

 Circle of Willis, Time of Flight, MIP  Open this link in a new window
    
SlidersSliders Overview

 Anatomic MRI of the Knee 1  Open this link in a new window
    
SlidersSliders Overview

 Anatomic Imaging of the Liver  Open this link in a new window
      

 Brain MRI Inversion Recovery  Open this link in a new window
    
 
spacer

• View the DATABASE results for 'Contrast' (373).Open this link in a new window


• View the NEWS results for 'Contrast' (77).Open this link in a new window.
 
Further Reading:
  Basics:
Magnetic resonance imaging
   by www.scholarpedia.org    
MRI's inside story
Thursday, 4 December 2003   by www.economist.com    
Image Characteristics and Quality
   by www.sprawls.org    
  News & More:
Sharpening MRI Images
Sunday, 1 December 2013   by www.onlinetmd.com    
A natural boost for MRI scans
Monday, 21 October 2013   by www.eurekalert.org    
A groundbreaking new graphene-based MRI contrast agent
Friday, 8 June 2012   by www.nanowerk.com    
New MRI Chemical Offers Amazing Contrast
Friday, 22 January 2010   by news.softpedia.com    
MRI Resources 
Spectroscopy - General - Equipment - Mobile MRI Rental - NMR - PACS
 
Image Quality
 
The quality of magnetic resonance imaging is particularly dependent on image resolution (matrix, field of view, slice thickness), contrast (TE, TR), signal to noise ratio (bandwidth, signal averaging) and lack of artifacts. These MRI parameters are affected by the field homogeneity, the field strength, the coil, the pulse sequence type and imaging techniques like parallel imaging.
See also Image Contrast Characteristics.
spacer

• View the DATABASE results for 'Image Quality' (42).Open this link in a new window


• View the NEWS results for 'Image Quality' (4).Open this link in a new window.
 
Further Reading:
  Basics:
Guidelines for Developing Automated Quality Control Procedures for Brain Magnetic Resonance Images Acquired in Multi-Centre Clinical Trials
Tuesday, 26 April 2011   by www.intechopen.com    
Image Characteristics and Quality
   by www.sprawls.org    
  News & More:
Patient movement during MRI: Additional points to ponder
Tuesday, 5 January 2016   by www.healthimaging.com    
Novel Imaging Technique Improves Prostate Cancer Detection
Tuesday, 6 January 2015   by health.ucsd.edu    
NanoMRI could image single biomolecules
Monday, 29 December 2014   by nanotechweb.org    
Breast MRI more widely used now that equipment, images are better
Sunday, 21 October 2012   by www.kentucky.com    
MRI Resources 
NMR - Research Labs - Service and Support - Homepages - Implant and Prosthesis pool - Equipment
 
Proton Density Weighted Image
 
An image produced by controlling the selection of scan parameters to minimize the effects of T1 and T2, resulting in an image dependent primarily on the density of protons in the imaging volume. Proton density contrast is a quantitative summary of the number of protons per unit tissue. The higher the number of protons in a given unit of tissue, the greater the transverse component of magnetization, and the brighter the signal on the proton density contrast image. Conversely the lower the number of protons in a given unit of tissue, the less the transverse magnetization and the darker the signal on the proton density image. Also called (Rho) r-weighted.
See also Density Weighted Imaging and Image Contrast Characteristics.
 
Images, Movies, Sliders:
 Knee MRI Coronal Pd Spir 001  Open this link in a new window
    
 
spacer

• View the DATABASE results for 'Proton Density Weighted Image' (4).Open this link in a new window

 
Further Reading:
  Basics:
Musculoskeletal MRI at 3.0 T: Relaxation Times and Image Contrast
Sunday, 1 August 2004   by www.ajronline.org    
Dogs Are People, Too
Saturday, 5 October 2013   by www.nytimes.com    
MRI Resources 
Cardiovascular Imaging - Jobs - Service and Support - RIS - Abdominal Imaging - DICOM
 
     1 - 5 (of 17)     next
Result Pages : [1]  [2]  [3 4]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?  



The impact of TTIP/TPP on the MRI scanner market will bring :
more variety 
better scanners 
more risk 
less regulation 
cheaper scanners 
Lost in Translation 
no change at all 

Look
      Ups





Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • US-TIP • The-Medical-Market
Copyright © 2003 - 2016 SoftWays. All rights reserved. [ 24 February 2017]
Terms of Use | Privacy Policy | Advertising
 [last update: 2017-02-21 00:45:00]