Segmented K-Space Data Acquisition
 
A set of k-space lines collected in a specified order but not constituting a complete coverage of k-space, thus can be used in conjunction with all ultrafast MRI techniques. Several segmental acquisitions may need to be run for complete coverage of k-space. If these lines are recorded for a single rather than multiple images, imaging time can be shortened considerably maintaining an acceptable temporal resolution.
For example, rapidly acquiring eight k-space lines per segment after each trigger until 128 lines of k-space are acquired in 16 triggers, thus makes image acquisition of multiple cardiac phases or anatomical slices possible in a breath-hold.
spacer
Selective Excitation
 
Controlling the frequency spectrum (bandwidth) of a RF pulse (via tailoring) while imposing a magnetic field gradient on spins, such that only a desired region will have an appropriate resonant frequency to be excited.
Originally used to excite all but a desired region; now often used to select only a desired region, such as a plane, for excitation. Used without simultaneous magnetic field gradients, tailored RF pulses can be used to selectively excite a particular spectral line or group of lines. RF and gradient pulse combinations can be designed to select both spatial regions and spectral frequencies.
spacer

• View the DATABASE results for 'Selective Excitation' (17).Open this link in a new window

MRI Resources 
Jobs - Spectroscopy - Calculation - Veterinary MRI - IR - Contrast Agents
 
Selective Irradiation
 
spacer
Self Induction
 
Magnetic field produced in a coil of wire that opposes the AC current being conducted.
spacer
Self Shielding
 
Magnetic shielding by attaching a high permeability yoke to the magnet (passive shielding) or by incorporating additional magnetic field-generating coils designed to reduce the external field (active shielding).

See also Magnetic shielding, Active Shielding and Passive Shielding.
spacer

• View the DATABASE results for 'Self Shielding' (3).Open this link in a new window

MRI Resources 
Directories - Resources - Brain MRI - General - Spectroscopy pool - Image Quality
 
previous       16 - 20 (of 167)      next 
S--Sa   Sa-Sc   Sc-Se    Se-Se    Se-Se   Se-Se   Sh-Sh   Sh-SH   Sh-Sh   SH-Si   Si-Si   Si-Si   Si-Si   Si-Si   Si-Si   Sk-Sl   Sl-Sl   Sl-So   So-Sp   Sp-Sp   Sp-Sp   Sp-Sp   SP-Sp   Sp-Sp   Sp-Sp   Sp-Sp   Sp-St   St-St   St-St   St-Su   Su-Su   Su-Sw   Sy-Sy   Sy-Sy   
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]